CTD

Reference Designator
CP01CNPM-WFP01-03-CTDPFK000
Start Depth
15
End Depth
70
Location
Wire Following Profiler Body
Current Status
Review Complete
Class
CTDPF (CTD Profiler)
Series
CTDPF-K
Science Discipline
Physical
Make
Sea-Bird
Model
SBE 52MP
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/CP01CNPM/WFP01/03-CTDPFK000/metadata
Method Data Stream Content Type
recovered_wfp ctdpf_ckl_wfp_instrument_recovered Data Products Report M2M Stats Science
recovered_wfp ctdpf_ckl_wfp_metadata_recovered Metadata Report M2M Engineering
telemetered ctdpf_ckl_wfp_instrument Data Products Report M2M Stats Science
telemetered ctdpf_ckl_wfp_metadata Metadata Report M2M Engineering
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review AR24 11/06/2017 04/12/2018 CGMCP-01CNPM-00001 CGVEH-CWFPCP-30002 CGINS-CTDPFK-00185 40.1392 -70.7795 113 134
2 Review AR31 11/02/2018 04/09/2019 CGMCP-01CNPM-00002 CGVEH-CWFPCP-30002 CGINS-CTDPFK-00185 40.1346 -70.7704 112 133
3 Review AR39 10/07/2019 06/14/2020 CGMCP-01CNPM-00003 CGVEH-CWFPCP-30002 CGINS-CTDPFK-00185 40.1392 -70.7794 111 132
Metadata Start Date End Date Comment
CP01CNPM-WFP01-03-CTDPFK000

Comparisons with the shipboard CTD casts were done for D0001 and R0001. The profiles do not match the CTD cast for D0001, and more closely align with the CTD cast for R0001. This pattern is consistent with the profiler's DOFST (data do not match the shipboard CTD cast at deployment and match at recovery). Comparison to the shipboard bottle data is outside the scope of this project.

By Lori Garzio, on 3/12/19

CP01CNPM

system missing global ranges for all parameters because the platform is not listed in the data_qc_global_range_values.csv

By Leila Belabbassi, on 3/15/19

New Note

Metadata Start Date End Date Comment
CP01CNPM-WFP01
12/31/18, 7:00 PM 2/6/19, 7:00 PM

Deployment 2: The WFP slid up the wire between profiles on various dates including 2019-01-01, 2019-01-29, and from 2019-02-05 through 2019-02-07.

Id: 1623 By: cdobson

CP01CNPM-WFP01
2/10/19, 7:00 PM 2/17/19, 7:00 PM

Deployment 2: The WFP becamestuck at 25m from 2019-02-11 through 2019-02-18.

Id: 1622 By: cdobson

CP01CNPM-WFP01
2/20/19, 7:00 PM 4/4/19, 8:07 AM

Deployment 2: Telemetry of data from the WFP ceased on 2019-02-21.

Id: 1621 By: cdobson

CP01CNPM-WFP01
12/2/19, 7:00 PM 6/14/20, 8:23 AM

Deployment 3: Starting 12/03/2019 WFP01 was unable to complete full range descending profiles.

Id: 1842 By: cdobson
Flag: not_operational Exclude: No

CP01CNPM-WFP01
12/31/19, 7:00 PM

Deployment 3: Due to a clock rollover issue, all WFPs have ceased sampling on 1/1/2020. No WFP data are being collected after this date. A resolution is being researched by the vendor.

Id: 1879 By: cdobson