Velocity Profiler (150kHz)

Reference Designator
CP03ISPM-RII01-02-ADCPTG010
Start Depth
70
End Depth
70
Location
Instrument Frame
Current Status
Review Complete
Class
ADCPT (Velocity Profiler (short range))
Series
ADCPT-G
Science Discipline
Physical
Make
Teledyne RDI
Model
WorkHorse Sentinel 150khz - inductive
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/CP03ISPM/RII01/02-ADCPTG010/metadata
Method Data Stream Content Type
recovered_host adcps_jln_stc_instrument_recovered Data Products Report M2M Stats Science
recovered_host adcps_jln_stc_metadata_recovered Metadata Report M2M Engineering
recovered_inst adcp_config Configuration Data Report M2M Engineering
recovered_inst adcp_engineering Engineering Data Report M2M Engineering
recovered_inst adcp_velocity_earth Earth Coordinates Data Products Report M2M Stats Science
telemetered adcps_jln_stc_instrument Data Products Report M2M Stats Science
telemetered adcps_jln_stc_metadata Metadata Report M2M Engineering
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review AR24 11/12/2017 04/11/2018 CGMCP-03ISPM-00001 CGINS-ADCPTG-20496 40.3647 -70.8882 68 91
2 Review AR31 10/23/2018 04/17/2019 CGMCP-03ISPM-00002 CGINS-ADCPTG-18660 40.3626 -70.8787 69 92
3 Review AR39 10/06/2019 06/13/2020 CGMCP-03ISPM-00003 CGINS-ADCPTG-18660 40.3648 -70.8887 69 92
Metadata Start Date End Date Comment
CP03ISPM-RII01-02-ADCPTG010
Deployment: 1

There are no recovered_inst data available for download. According to the ingest csv, the recovered_inst raw data files are available. These files should be ingested or the dataset should be annotated to indicate why the data aren't available.

By Lori Garzio, on 8/16/19

CP03ISPM-RII01-02-ADCPTG010
Deployment: 1

The .nc files are missing the pressure coordinate. Also missing two key variables: percent_bad_beams and bin_depths

By Lori Garzio, on 8/16/19

New Note

Metadata Start Date End Date Comment
CP03ISPM-RII01-02-ADCPTG010
Method: telemetered
Stream: adcps_jln_stc_instrument
11/11/17, 10:26 PM

Telemetered inductive ADCP data is in pd12 format which does not include the depth-to-first-bin value which is needed for proper calculation of the data products. A nominal value is entered into the system so that telemetered data can be used when needed. For the best data the recovered data in pd0 format should be used.

Id: 1918 By: cdobson