Method | Data Stream | Content | Type | |
---|---|---|---|---|
recovered_wfp | vel3d_k_wfp_instrument | Data Products | Report M2M Stats | Science |
recovered_wfp | vel3d_k_wfp_metadata | Metadata | Report M2M | Engineering |
recovered_wfp | vel3d_k_wfp_string | String Data | Report M2M | Engineering |
telemetered | vel3d_k_wfp_stc_instrument | Data Products | Report M2M Stats | Science |
telemetered | vel3d_k_wfp_stc_metadata | Metadata | Report M2M | Engineering |
Deployment | Cruise | Start Date | Stop Date | Mooring Asset | Node Asset | Sensor Asset | Latitude | Longitude | Deployment Depth | Water Depth |
---|---|---|---|---|---|---|---|---|---|---|
1 Review | AR24 | 11/12/2017 | 04/11/2018 | CGMCP-03ISPM-00001 | CGVEH-CWFPCP-30001 | CGINS-VEL3DK-00468 | 40.3647 | -70.8882 | 67 | 91 |
2 Review | AR31 | 10/23/2018 | 04/17/2019 | CGMCP-03ISPM-00002 | CGVEH-CWFPCP-30001 | CGINS-VEL3DK-00468 | 40.3626 | -70.8787 | 68 | 92 |
3 Review | AR39 | 10/06/2019 | 06/13/2020 | CGMCP-03ISPM-00003 | CGVEH-CWFPCP-30001 | CGINS-VEL3DK-00468 | 40.3648 | -70.8887 | 68 | 92 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP03ISPM-WFP01-01-VEL3DK000 |
Annotation ID 1406 needs to be updated when the code is corrected. By Lori Garzio, on 8/22/19 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP03ISPM-WFP01-01-VEL3DK000 Method: telemetered Stream: vel3d_k_wfp_stc_instrument |
11/11/17, 10:26 PM |
During descending profiles, the vertical velocity estimates may be incorrect due to interference with beam 3 measurements arising from the nearby CTD sensor. Users should cross-compare these vertical velocity estimates with those derived from the CTD pressure record to determine profiles of possible concern. Good vertical velocity estimates will closely align with the CTD derived estimates. Regardless, there should be minimal impact on the estimates of the horizontal velocities. However, care should be taken with these estimates to confirm this conclusion.
Id: 1657
By: swhite
|
|
CP03ISPM-WFP01-01-VEL3DK000 Method: recovered_wfp Stream: vel3d_k_wfp_instrument |
11/11/17, 10:26 PM |
During descending profiles, the vertical velocity estimates may be incorrect due to interference with beam 3 measurements arising from the nearby CTD sensor. Users should cross-compare these vertical velocity estimates with those derived from the CTD pressure record to determine profiles of possible concern. Good vertical velocity estimates will closely align with the CTD derived estimates. Regardless, there should be minimal impact on the estimates of the horizontal velocities. However, care should be taken with these estimates to confirm this conclusion.
Id: 1658
By: swhite
|
|
CP03ISPM-WFP01-01-VEL3DK000 Method: telemetered Stream: vel3d_k_wfp_stc_instrument |
11/11/17, 10:26 PM | 4/11/18, 3:06 PM |
Deployment 1; During the times indicated by the start and end times of this annotation, the VEL3D instrument on the Profiler contained a firmware bug which caused beam 3 (downward-facing) to be on during ascents, and beam 1 (upward-facing) to be on during descents, instead of vice versa. Therefore the issue of incorrect vertical velocity estimates due to interference with the CTD noted in a previous annotation will affect the ascending profiles rather than the descending profiles.
Id: 1729
By: areed
|
CP03ISPM-WFP01-01-VEL3DK000 Method: recovered_wfp Stream: vel3d_k_wfp_instrument |
11/11/17, 10:26 PM | 4/11/18, 3:06 PM |
Deployment 1; During the times indicated by the start and end times of this annotation, the VEL3D instrument on the Profiler contained a firmware bug which caused beam 3 (downward-facing) to be on during ascents, and beam 1 (upward-facing) to be on during descents, instead of vice versa. Therefore the issue of incorrect vertical velocity estimates due to interference with the CTD noted in a previous annotation will affect the ascending profiles rather than the descending profiles.
Id: 1730
By: areed
|
CP03ISPM-WFP01-01-VEL3DK000 |
11/12/17, 12:00 AM | 2/14/19, 7:00 PM |
UPDATED, 2020-04-15: Corrected algorithm has been uploaded to the system and is active. * The algorithm supplied by the vendor to calculate water velocities is incorrect. Corrected code is currently in review. Id: 1406 By: swhite |
CP03ISPM-WFP01 |
12/31/19, 7:00 PM | 6/13/20, 3:24 PM |
Deployment 3: Due to a clock rollover issue, all WFPs have ceased sampling on 1/1/2020. No WFP data are being collected after this date. A resolution is being researched by the vendor. Id: 1880 By: cdobson |