3-Wavelength Fluorometer

Reference Designator
CP03ISPM-WFP01-04-FLORTK000
Start Depth
15
End Depth
70
Location
Wire Following Profiler Body
Current Status
Review Complete
Class
FLORT (3-Wavelength Fluorometer)
Series
FLORT-K
Science Discipline
Biological
Make
WET Labs
Model
ECO Triplet
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/CP03ISPM/WFP01/04-FLORTK000/metadata
Method Data Stream Content Type
recovered_wfp flort_sample Data Products Report M2M Stats Science
telemetered flort_sample Data Products Report M2M Stats Science
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review AR24 11/12/2017 04/11/2018 CGMCP-03ISPM-00001 CGVEH-CWFPCP-30001 CGINS-FLORTK-01537 40.3647 -70.8882 67 91
2 Review AR31 10/23/2018 04/17/2019 CGMCP-03ISPM-00002 CGVEH-CWFPCP-30001 CGINS-FLORTK-01537 40.3626 -70.8787 68 92
3 Review AR39 10/06/2019 06/13/2020 CGMCP-03ISPM-00003 CGVEH-CWFPCP-30001 CGINS-FLORTK-01537 40.3648 -70.8887 68 92
Metadata Start Date End Date Comment
CP03ISPM-WFP01-04-FLORTK000

Comparison with the shipboard CTD cast was done for D0001 and R0001. For both comparisons, the chl-a recorded from the profiler was higher compared to chl-a from the shipboard CTD cast. Further analysis using the shipboard bottle data should be conducted. This is outside the scope of this project.

By Lori Garzio, on 3/12/19

CP03ISPM-WFP01-04-FLORTK000
Deployment: 1

The pressure coordinate (pressure_depth) is an array of all fill values. This variable should contain valid pressure data. There is another variable in the files called in_ctd_pressure that contains what appears to be valid pressure data. This is the case for every FLORTK on every WFP. https://opendap.oceanobservatories.org/thredds/dodsC/ooi/lgarzio@marine.rutgers.edu/20181213T005359-CP03ISPM-WFP01-04-FLORTK000-recovered_wfp-flort_sample/deployment0001_CP03ISPM-WFP01-04-FLORTK000-recovered_wfp-flort_sample_20171112T050202-20180411T180529.nc.ascii?pressure_depth[0:1:1000],int_ctd_pressure[0:1:1000]

By Lori Garzio, on 3/12/19

CP03ISPM-WFP01-04-FLORTK000
Deployment: 1
1/1/18, 12:00 AM 2/15/18, 12:00 AM

Suspect optical backscatter in the middle of the deployment.

By Lori Garzio, on 8/22/19

New Note

Metadata Start Date End Date Comment
CP03ISPM-WFP01
12/31/19, 7:00 PM 6/13/20, 3:24 PM

Deployment 3: Due to a clock rollover issue, all WFPs have ceased sampling on 1/1/2020. No WFP data are being collected after this date. A resolution is being researched by the vendor.

Id: 1880 By: cdobson