Single Point Velocity Meter

Reference Designator
CP03ISSM-RID26-04-VELPTA000
Start Depth
7
End Depth
7
Location
Near Surface Instrument Frame
Current Status
Review Complete
Class
VELPT (Single Point Velocity Meter)
Series
VELPT-A
Science Discipline
Physical
Make
Nortek
Model
Aquadopp 300m
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/CP03ISSM/RID26/04-VELPTA000/metadata
Method Data Stream Content Type
recovered_host velpt_ab_dcl_diagnostics_metadata_recovered Diagnostics Metadata Report M2M Engineering
recovered_host velpt_ab_dcl_diagnostics_recovered Diagnostics Data Report M2M Engineering
recovered_host velpt_ab_dcl_instrument_recovered Data Products Report M2M Stats Science
recovered_inst velpt_ab_diagnostics_metadata_recovered Diagnostics Metadata Report M2M Engineering
recovered_inst velpt_ab_diagnostics_recovered Diagnostics Data Report M2M Engineering
recovered_inst velpt_ab_instrument_metadata_recovered Metadata Report M2M Engineering
recovered_inst velpt_ab_instrument_recovered Data Products Report M2M Stats Science
telemetered velpt_ab_dcl_diagnostics Diagnostics Data Report M2M Engineering
telemetered velpt_ab_dcl_diagnostics_metadata Diagnostics Metadata Report M2M Engineering
telemetered velpt_ab_dcl_instrument Data Products Report M2M Stats Science
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review KN-224 12/14/2014 12/29/2014 CGMCP-03ISSM-00001 CGINS-VELPTA-11397 40.3619 -70.878 7 94
2 Review AT-27 05/09/2015 10/21/2015 CGMCP-03ISSM-00002 CGINS-VELPTA-12203 40.3595 -70.8853 7 97
3 Review AT-31 10/21/2015 05/14/2016 CGMCP-03ISSM-00003 CGINS-VELPTA-12747 40.3647 -70.8884 7 96
4 Review AR-04 05/28/2016 10/15/2016 CGMCP-03ISSM-00004 CGINS-VELPTA-11377 40.3595 -70.885 7 95
5 Review AR-08 10/11/2016 06/15/2017 CGMCP-03ISSM-00005 CGINS-VELPTA-12234 40.3619 -70.8783 7 92
6 Review AR-18 06/15/2017 11/01/2017 CGMCP-03ISSM-00006 CGINS-VELPTA-12224 40.3671 -70.8816 7 93
7 Review AR-24 11/01/2017 03/28/2018 CGMCP-03ISSM-00007 CGINS-VELPTA-11377 40.359 -70.8855 7 93
8 Review AR-28 03/25/2018 10/26/2018 CGMCP-03ISSM-00008 CGINS-VELPTA-11412 40.3672 -70.8821 7 94
9 Review AR-31 10/29/2018 04/06/2019 CGMCP-03ISSM-00009 CGINS-VELPTA-11386 40.3673 -70.8811 7 92
10 Review AR-34 04/07/2019 10/01/2019 CGMCP-03ISSM-00010 CGINS-VELPTA-12198 40.3672 -70.8815 7 93
11 Review AR-39 09/26/2019 CGMCP-03ISSM-00011 CGINS-VELPTA-12234 40.3594 -70.8851 7 93
Metadata Start Date End Date Comment
CP03ISSM-RID26-04-VELPTA000

recovered_inst data are not available for download from deployments 1 and 4 - the raw data files are available according to the ingest csvs. The data should be ingested or annotated to indicate why they are not available.

By Lori Garzio, on 8/14/19

CP03ISSM-RID26-04-VELPTA000

3-31% of the data points collected during all deployments are when the instrument is tilted too far ( has pitch or roll >20 degrees). Example plot attached

By Lori Garzio, on 8/14/19

CP03ISSM-RID26-04-VELPTA000
Deployment: 2
5/9/15, 12:57 AM 10/21/15, 6:50 PM

Suspect upward_velocity values for deployments 2 (middle panel in attached plot) - values periodically exceed northward and eastward values.

By Lori Garzio, on 8/14/19

New Note

Metadata Start Date End Date Comment
CP03ISSM-RID26-04-VELPTA000
12/14/14, 1:52 PM

According to the manufacturer, data are suspect when the instrument is tilted more than 20 degrees (measured by pitch and roll).

Id: 817 By: lgarzio

CP03ISSM-RID26-04-VELPTA000
12/14/14, 1:52 PM 12/29/14, 7:00 PM

No recovered host data available.

Id: 452 By: leila

CP03ISSM-RID26-04-VELPTA000
4/12/16, 2:00 AM 5/14/16, 9:45 AM

Failed.

Id: 611 By: leila
Flag: not_operational Exclude: No

CP03ISSM-RID26-04-VELPTA000
10/11/16, 9:39 AM 6/15/17, 11:56 AM

No recovered instrument data available.

Id: 456 By: leila

CP03ISSM-RID26
Method: telemetered
6/17/18, 4:04 PM 10/26/18, 3:52 PM

Data telemetry disabled due to chronic issues with power generation and management. The instruments could be revived to send data via telemetry when sufficient power is available. Note: Instruments may continue to collect data on internal battery power.

Id: 1510 By: leila

CP03ISSM
1/7/20, 7:00 PM

Deployment 11: UPDATED 2020-04-27: As of 2020-03-12 telemetered data for most instruments are being sent home via Iridium. A gap in telemetered data is expected between 2020-01-08 and 2020-03-12 and will not be backfilled but these data will be available upon recovery.* As of 2020-01-08 both fleet broadband telemetry pathways have failed and data can no longer be telemetered to shore from this mooring. Instruments will continue to log data which will be available upon recovery.

Id: 1889 By: cdobson
Flag: not_operational Exclude: No