pCO2 Air-Sea

Reference Designator
CP04OSSM-SBD12-04-PCO2AA000
Start Depth
0
End Depth
0
Location
Buoy Well (1 probe in air, 1 probe in water)
Current Status
Review Complete
Class
PCO2A (pCO2 Air-Sea)
Series
PCO2A-A
Science Discipline
Chemical
Make
Pro-Oceanus
Model
pCO2-pro
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/CP04OSSM/SBD12/04-PCO2AA000/metadata
Method Data Stream Content Type
recovered_host pco2a_a_dcl_instrument_air_recovered In-Air Data Products Report M2M Stats Science
recovered_host pco2a_a_dcl_instrument_water_recovered In-Water Data Products Report M2M Stats Science
telemetered pco2a_a_dcl_instrument_air In-Air Data Products Report M2M Stats Science
telemetered pco2a_a_dcl_instrument_water In-Water Data Products Report M2M Stats Science
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review KN-224 12/12/2014 05/10/2015 CGMCP-04OSSM-00001 CGINS-PCO2AA-33184 39.9362 -70.8803 0 455
2 Review AT-27 05/10/2015 10/23/2015 CGMCP-04OSSM-00002 CGINS-PCO2AA-34229 39.9406 -70.8785 0 442
3 Review AT-31 10/22/2015 05/13/2016 CGMCP-04OSSM-00003 CGINS-PCO2AA-35261 39.9373 -70.8867 0 454
4 Review AR-04 05/27/2016 10/12/2016 CGMCP-04OSSM-00004 CGINS-PCO2AA-34222 39.94 -70.8774 0 446
5 Review AR-08 10/12/2016 06/08/2017 CGMCP-04OSSM-00005 CGINS-PCO2AA-34223 39.9371 -70.887 0 450
6 Review AR-18 06/06/2017 11/02/2017 CGMCP-04OSSM-00006 CGINS-PCO2AA-33183 39.9394 -70.8768 0 443
7 Review AR-24 10/28/2017 03/29/2018 CGMCP-04OSSM-00007 CGINS-PCO2AA-31124 39.9376 -70.8863 0 452
8 Review AR-28 03/24/2018 10/30/2018 CGMCP-04OSSM-00008 CGINS-PCO2AA-35262 39.9391 -70.8764 0 445
9 Review AR-31 10/25/2018 02/04/2019 CGMCP-04OSSM-00009 CGINS-PCO2AA-35260 39.9361 -70.885 0 449
10 Review AR-34 04/05/2019 09/28/2019 CGMCP-04OSSM-00010 CGINS-PCO2AA-34231 39.9399 -70.8774 0 451
11 Review AR-39 09/29/2019 CGMCP-04OSSM-00011 CGINS-PCO2AA-35264 39.9368 -70.8865 0 447
Metadata Start Date End Date Comment
CP04OSSM-SBD12-04-PCO2AA000

pco2_co2flux is an array of fill values for every deployment. See CP01CNSM-SBD12-04-PCO2AA000 for additional notes on potentially the same issue.

By Lori Garzio, on 7/31/19

CP04OSSM-SBD12-04-PCO2AA000
Deployment: 4

measured_air_co2 values are higher at the beginning of deployment 4 (~420-430 ppm) compared to the end of deployment 3 (~400-410 ppm)

By Lori Garzio, on 7/31/19

CP04OSSM-SBD12-04-PCO2AA000
Deployment: 5
10/12/16, 2:26 PM 6/8/17, 12:05 PM

Suspect data: measured_water_co2 values drop off regularly.

By Lori Garzio, on 7/31/19

CP04OSSM-SBD12-04-PCO2AA000
Deployment: 6
6/6/17, 2:51 PM 11/2/17, 2:15 PM

Suspect data: measured_water_co2 values drop off regularly.

By Lori Garzio, on 7/31/19

CP04OSSM-SBD12-04-PCO2AA000
Deployment: 7
10/28/17, 2:24 PM 3/29/18, 12:42 PM

Suspect data - measured_air_co2 values are too high.

By Lori Garzio, on 7/31/19

New Note

Metadata Start Date End Date Comment
CP04OSSM
Method: telemetered
10/22/15, 3:38 PM 5/13/16, 1:18 PM

Fuel cell impacted available instrument power, so telemetered data are not available or reliable. Recovered data are available for instruments that could record internally.

Id: 1362 By: leila

CP04OSSM-SBD12-04-PCO2AA000
1/30/17, 1:00 AM 6/8/17, 8:00 AM

Instrument stopped at the end of January.

Id: 692 By: leila
Flag: not_operational Exclude: No

CP04OSSM
Method: telemetered
1/30/19, 10:00 AM 4/8/19, 7:39 AM

CP04OSSM D00009 buoy broke free and went adrift on 1/30/2019, 1500UTC. All instruments on the MFN ceased telemetry on 1/30/2019 and instruments on the buoy/NSIF continued to sample through 2/1/2019. No more telemetered data expected for this deployment.

Id: 1557 By: cdobson

CP04OSSM-SBD12-04-PCO2AA000
4/5/19, 10:55 AM 9/28/19, 8:38 AM

Due to a firmware bug, the sensor zeroing function for correcting the baseline drift over time is not currently saving the new reference value to compare the raw CO2 value outputs during the measurements. The sensors are zeroing at the end of the sample cycle when scheduled, but the firmware is not storing the new value. This results in the original zero reference value being used for all measurements and an observable and consistent negative drift in both the air and water CO2 values. The vendor has indicated that the data may be correctable post-recovery, however at this time no such correction algorithm has been provided

Id: 1716 By: swhite
Flag: fail Exclude: Yes

CP04OSSM-SBD12-04-PCO2AA000
10/31/19, 8:00 PM 11/14/19, 7:00 AM

Deployment 11: *UPDATED 2019-11-14: Instrument is powered back on and functional as of 2019-11-14. A data gap is expected 2019-11-04 through 2019-11-14.* PCO2A2 was disabled 2019-11-04 after issues were observed starting 2019-11-01. No data expected after this point.

Id: 1812 By: cdobson