3-Wavelength Fluorometer

Reference Designator
CP05MOAS-PG583-03-FLORTM000
Start Depth
0
End Depth
200
Location
Current Status
Review Complete
Class
FLORT (3-Wavelength Fluorometer)
Series
FLORT-M
Science Discipline
Biological
Make
WET Labs
Model
ECO Puck FLBBCD-SLK
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/CP05MOAS/PG583/03-FLORTM000/metadata
Method Data Stream Content Type
recovered_host flort_m_sample Data Products Report M2M Stats Science
telemetered flort_m_sample Data Products Report M2M Stats Science
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review SI20170418 04/19/2017 07/27/2017 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00583 CGINS-FLORTM-04093 40.3704 -70.9544 200 200
2 Review TS20200613 06/13/2020 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00583 CGINS-FLORTM-04093 40.0852 -70.3662 200 200
Metadata Start Date End Date Comment
CP05MOAS

Bad altimeter pings are apparent in many deployments of the coastal gliders. A few example plots are attached - m_water_depth plotted as semi-transparent bathymetry.

By Lori Garzio, on 5/10/19

CP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/15/19

CP05MOAS-PG583

Global ranges need to be added to the system for every instrument on this glider.

By Lori Garzio, on 9/10/19

New Note

No annotations yet.