CTD (180 m)

Reference Designator
GA01SUMO-RII11-02-CTDMOQ014
Start Depth
180
End Depth
180
Location
Fixed on Inductive Wire
Current Status
Review Complete
Class
CTDMO (CTD Mooring (Inductive))
Series
CTDMO-Q
Science Discipline
Physical
Make
Sea-Bird
Model
SBE 37IM
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/GA01SUMO/RII11/02-CTDMOQ014/metadata
Method Data Stream Content Type
recovered_host ctdmo_ghqr_imodem_instrument_recovered Data Products Report M2M Stats Science
recovered_host ctdmo_ghqr_imodem_metadata_recovered Metadata Report M2M Engineering
recovered_inst ctdmo_ghqr_instrument_recovered Data Products Report M2M Stats Science
telemetered ctdmo_ghqr_imodem_instrument Data Products Report M2M Stats Science
telemetered ctdmo_ghqr_imodem_metadata Metadata Report M2M Engineering
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review AT-26-30 03/15/2015 11/26/2015 CGMGA-01SUMO-00001 CGINS-CTDMOQ-12635 -42.9228 -42.4393 180 5161
2 Review NBP-15-10 11/14/2015 11/08/2016 CGMGA-01SUMO-00002 CGINS-CTDMOQ-13381 -42.9246 -42.5867 180 5188
3 Review NBP16-09 10/27/2016 01/14/2018 CGMGA-01SUMO-00003 CGINS-CTDMOQ-12635 -42.9204 -42.4409 180 5175
Metadata Start Date End Date Comment
GA01SUMO-RII11
12/4/15, 12:00 AM 3/14/16, 12:00 AM

Deployment 2: The recovered_host data from the following instruments have the same 4 data gaps [['2015-12-04', '2015-12-14'], ['2016-01-06', '2016-01-11'], ['2016-02-05', '2016-02-10'], ['2016-03-11', '2016-03-14']]. These gaps should be annotated to explain why the data are missing, or gaps should be filled in if the data are available. [02-FLORDG031, 02-FLORDG032, 02-FLORDG033, 02-DOSTAD031, 02-DOSTAD032, 02-DOSTAD033]

By Lori Garzio, on 2/11/19

GA01SUMO-RII11-02-CTDMOQ014
Deployment: 2

Annotation ID 666 should be updated to indicate which sensor was bent.

By Lori Garzio, on 1/30/19

GA01SUMO-RII11-02-CTDMOQ014
Deployment: 3
10/27/16, 1:47 AM 1/14/18, 10:24 AM

Density and salinity gradually drifted higher towards the end of the deployment. A similar pattern isn't observed in surrounding CTDs and annotation ID 391 says that the temperature probe was bent. It's unclear when the issue started, so excluding the entire deployment from the final stats.

By Lori Garzio, on 1/29/19

GA01SUMO-RII11-02-CTDMOQ014
Deployment: 3
12/31/17, 10:52 PM 1/1/18, 1:00 AM

There are 14 data points at the end of the deployment that are available in the telemetered and recovered_host streams that aren't available in the recovered_inst stream.

By Lori Garzio, on 1/30/19

New Note

Metadata Start Date End Date Comment
GA01SUMO
3/15/15, 5:25 PM 11/26/15, 5:08 AM

Deployment 1: Buoy log files showed that seawater was first detected inside the buoy well on 3/26/2015. Both leak detectors showed increasing presence of water over the period of deployment. Upon recovery of the mooring, no external damage was visible and the buoy did not appear to be floating lower than normal.

Id: 653 By: lgarzio

GA01SUMO-RII11-02-CTDMOQ014
3/15/15, 5:25 PM 11/26/15, 5:08 AM

Deployment 1: Missing a conductivity cell shroud at recovery.

Id: 660 By: lgarzio

GA01SUMO-RII11-02-CTDMOQ014
11/14/15, 4:06 PM 11/8/16, 5:10 AM

Deployment 2: At recovery, instrument was missing its casing and a sensor was bent.

Id: 666 By: lgarzio

GA01SUMO-RII11-02-CTDMOQ014
10/26/16, 9:47 PM 12/31/17, 7:00 PM

Deployment 3: Upon recovery, the conductivity cell guard was missing and the temperature probe was bent.

Id: 391 By: lgarzio