3-Wavelength Fluorometer

Reference Designator
GA05MOAS-PG563-04-FLORTO000
Start Depth
0
End Depth
200
Location
Current Status
Review Complete
Class
FLORT (3-Wavelength Fluorometer)
Series
FLORT-O
Science Discipline
Biological
Make
WET Labs
Model
ECO BB3
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/GA05MOAS/PG563/04-FLORTO000/metadata
Method Data Stream Content Type
recovered_host flort_o_glider_data Data Products Report M2M Stats Science
telemetered flort_o_glider_data Data Products Report M2M Stats Science
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review NBP-15-10 11/24/2015 05/02/2016 CGVGA-05MOAS-00000 CGVEH-GLDRGP-00563 CGINS-FLORTO-01330 -42.5883 -42.645 200 200
Metadata Start Date End Date Comment
GA05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/31/19

GA05MOAS-PG563-04-FLORTO000

There are no global ranges in the system for sci_bb3slo_b470_units, flort_o_bback_b470_total, flort_o_bback_b532_total, sci_bb3slo_b660_units, and flort_o_bback_b660_total.

By Lori Garzio, on 9/17/19

GA05MOAS-PG563-04-FLORTO000
Deployment: 1

The arrays for the following variables are all NaNs: sci_bb3slo_b470_units, flort_o_bback_b470_total, flort_o_bback_b532_total, sci_bb3slo_b660_units, flort_o_bback_b660_total. According to the provenance.json (attached), the following cal values are missing: CC_1_dark_counts_volume_scatter, CC_1_scale_factor_volume_scatter, CC_1_measurement_wavelength, CC_2_dark_counts_volume_scatter, CC_2_scale_factor_volume_scatter, CC_2_measurement_wavelength, CC_3_dark_counts_volume_scatter, CC_3_scale_factor_volume_scatter, and CC_3_measurement_wavelength.

By Lori Garzio, on 9/17/19

New Note

Metadata Start Date End Date Comment
GA05MOAS-PG563
11/24/15, 4:56 PM 1/29/16, 7:00 PM

Deployment 1: The ballast pump developed erratic behavior which progressed to complete pump failure. Early signs of issue seen in late Dec 2015. Glider went adrift, the science sensors were turned off on 1/30/2016 and the glider was recovered on 05/02/2016.

Id: 415 By: lgarzio