3-Wavelength Fluorometer

Reference Designator
GI05MOAS-PG564-04-FLORTO000
Start Depth
0
End Depth
200
Location
Current Status
Will Not Review
Class
FLORT (3-Wavelength Fluorometer)
Series
FLORT-O
Science Discipline
Biological
Make
WET Labs
Model
ECO BB3
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/GI05MOAS/PG564/04-FLORTO000/metadata
Method Data Stream Content Type
recovered_host flort_o_glider_data Data Products Report M2M Stats Science
telemetered flort_o_glider_data Data Products Report M2M Stats Science
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review AR-07-01 07/09/2016 07/11/2016 CGVGI-05MOAS-00000 CGVEH-GLDRGP-00564 CGINS-FLORTO-01331 59.8639 -39.124 200 200
Metadata Start Date End Date Comment
GI05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/31/19

New Note

No annotations yet.