3-Wavelength Fluorometer

Reference Designator
GP05MOAS-PG514-04-FLORTO000
Start Depth
0
End Depth
200
Location
Current Status
Will Not Review
Class
FLORT (3-Wavelength Fluorometer)
Series
FLORT-O
Science Discipline
Biological
Make
WET Labs
Model
ECO BB3
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/GP05MOAS/PG514/04-FLORTO000/metadata
Method Data Stream Content Type
recovered_host flort_o_glider_data Data Products Report M2M Stats Science
telemetered flort_o_glider_data Data Products Report M2M Stats Science

No deployments found

Metadata Start Date End Date Comment
GP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/31/19

New Note

Metadata Start Date End Date Comment
GP05MOAS-PG514
6/2/15, 6:00 AM 7/25/15, 2:12 PM

Deployment 2: Recovered data are not available for the beginning of the deployment due to a logging issue with the recovered high-resolution files.

Id: 593 By: lgarzio

GP05MOAS-PG514
6/2/15, 6:00 AM 11/15/15, 7:00 PM

Deployment 2: Glider was found damaged upon recovery, and the NUTNR sensor was missing.

Id: 175 By: lgarzio