Method | Data Stream | Content | Type | |
---|---|---|---|---|
streamed | ctdpf_optode_calibration_coefficients | Optode Calibration Data | Report M2M | Engineering |
streamed | ctdpf_optode_configuration | Optode Configuration Data | Report M2M | Engineering |
streamed | ctdpf_optode_hardware | Optode Hardware Data | Report M2M | Engineering |
streamed | ctdpf_optode_sample | Optode Data Products | Report M2M Stats | Science |
streamed | ctdpf_optode_settings | Optode Settings Data | Report M2M | Engineering |
streamed | ctdpf_optode_status | Optode Status Data | Report M2M | Engineering |
Deployment | Cruise | Start Date | Stop Date | Mooring Asset | Node Asset | Sensor Asset | Latitude | Longitude | Deployment Depth | Water Depth |
---|---|---|---|---|---|---|---|---|---|---|
1 Review | TN-313 | 08/07/2014 | 06/13/2015 | ATAPL-65310-050-0007 | ATAPL-67627-00003 | 45.8168 | -129.754 | 2606 | 2606 | |
2 Review | TN-326 | 07/13/2015 | 07/11/2016 | ATAPL-65310-840-0011 | ATAPL-67627-00005 | 45.8169 | -129.754 | 2606 | 2606 | |
3 Review | SKQ201610S | 07/12/2016 | 07/31/2017 | ATAPL-65310-050-0007 | ATAPL-67627-00003 | 45.8169 | -129.754 | 2606 | 2606 | |
4 Review | RR1713 | 07/31/2017 | 07/09/2018 | ATAPL-65310-840-0011 | ATAPL-67627-00005 | 45.8169 | -129.754 | 2606 | 2606 | |
5 Review | RR1810 | 07/09/2018 | 06/17/2019 | ATAPL-65310-840-0011 | ATAPL-67627-00003 | 45.8169 | -129.754 | 2606 | 2606 | |
6 Review | AT-42-12 | 06/17/2019 | ATAPL-65310-840-0011 | ATAPL-67627-00004 | 45.8169 | -129.754 | 2606 | 2606 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
RS03AXBS-LJ03A-12-CTDPFB301 |
The deployment depth in asset management is 2606 for each deployment, however the actual pressure data from the CTD is around 2645 dbar. By Lori Garzio, on 3/28/19 |
||
RS03AXBS-LJ03A-12-CTDPFB301 Deployment: 2 |
7/13/15, 12:00 AM | 7/12/16, 12:00 AM |
Annotation ID 213: "Due to errors during vendor calibration, a calibration offset (CC_offset) of -111.19 is being applied during the on-demand creation of the L1 Seawater Pressure (dbar) data product from this instrument." - it doesn't look like this offset is actually being applied as the pressure is still 120 dbar deeper than the other deployments and the deployment depth from asset management. This issue needs to be investigated. Excluding from summary data ranges. Annotation ID 25 should be deleted (it's a duplicate of ID 213, but 213 has more info) By Lori Garzio, on 3/28/19 |
RS03AXBS-LJ03A-12-CTDPFB301 Deployment: 3 |
7/12/16, 12:00 AM | 7/31/17, 1:00 AM | |
RS03AXBS-LJ03A-12-CTDPFB301 Deployment: 3 |
9/10/16, 12:00 AM | 9/15/16, 12:00 AM | |
RS03AXBS-LJ03A-12-CTDPFB301 Deployment: 4 |
10/15/17, 12:00 AM | 7/9/18, 12:30 AM |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
RS03AXBS-LJ03A-12-CTDPFB301 Method: streamed Stream: ctdpf_optode_sample |
8/23/14, 8:34 PM | 8/25/14, 12:41 PM |
Data interruption.
Id: 1209
By: knuth
|
RS03AXBS-LJ03A-12-CTDPFB301 Method: streamed Stream: ctdpf_optode_sample |
11/4/14, 11:05 AM | 11/5/14, 2:08 PM |
Data interruption.
Id: 1210
By: knuth
|
RS03AXBS-LJ03A-12-CTDPFB301 Method: streamed Stream: ctdpf_optode_sample |
11/14/14, 1:36 PM | 11/17/14, 1:09 PM |
Data interruption.
Id: 1211
By: knuth
|
RS03AXBS |
12/7/14, 2:45 PM | 12/15/14, 7:00 PM |
Note: PFE down. HVPS1 MOV explosion, 800A breaker tripped, investigation and restoration Id: 1212 By: knuth |
RS03AXBS-LJ03A-12-CTDPFB301 Method: streamed Stream: ctdpf_optode_sample |
12/7/14, 3:59 PM | 12/11/14, 6:20 PM |
Data interruption.
Id: 1213
By: knuth
|
RS03AXBS |
1/7/15, 2:32 AM | 1/7/15, 3:06 AM |
Note: PNWGP Portland <-> Seattle outage Id: 1214 By: knuth |
RS03AXBS |
1/30/15, 7:00 PM | 2/3/15, 7:00 PM |
Note: Intermittent partial data loss due to storage drive problems at OTB Id: 1215 By: knuth |
RS03AXBS-LJ03A-12-CTDPFB301 Method: streamed Stream: ctdpf_optode_sample |
2/1/15, 3:36 PM | 2/3/15, 4:57 AM |
Data interruption.
Id: 1216
By: knuth
|
RS03AXBS-LJ03A-12-CTDPFB301 Method: streamed Stream: ctdpf_optode_sample |
3/7/15, 6:01 PM | 3/9/15, 3:39 AM |
Data interruption.
Id: 1217
By: knuth
|
RS03AXBS |
3/21/15, 10:10 AM | 3/22/15, 12:20 AM |
Note: PNWGP outage due to City of Seattle fiber cable work Id: 1218 By: knuth |
RS03AXBS-LJ03A-12-CTDPFB301 Method: streamed Stream: ctdpf_optode_sample |
3/30/15, 1:49 PM | 4/2/15, 7:06 PM |
Data interruption.
Id: 1219
By: knuth
|
RS03AXBS |
6/12/15, 8:00 PM | 6/15/15, 12:30 PM |
Note: Network issues due to fire that damaged fibers between Portland and Seattle Id: 1220 By: knuth |
RS03AXBS-LJ03A-12-CTDPFB301 Method: streamed Stream: ctdpf-optode-sample |
7/12/15, 8:00 PM | 7/11/16, 8:00 PM |
An issue with the vendor’s calibration of the strain gauge pressure sensor is under investigation. The offset for pressure from 12-CTDPFB301 to 2016-07-12 is approximately 111 dbar. Id: 25 |
RS03AXBS-LJ03A-12-CTDPFB301 Method: streamed Stream: ctdpf_optode_sample |
7/13/15, 1:43 AM | 7/14/15, 7:18 PM |
Data interruption.
Id: 1221
By: knuth
|
RS03AXBS-LJ03A-12-CTDPFB301 Method: streamed Stream: ctdpf_optode_sample |
7/13/15, 1:43 AM | 7/10/16, 8:00 PM |
Due to errors during vendor calibration, a calibration offset (CC_offset) of -111.19 is being applied during the on-demand creation of the L1 Seawater Pressure (dbar) data product from this instrument. For more information please contact help@oceanobservatories.org. Id: 213 By: friedrich.knuth |
RS03AXBS |
8/28/15, 8:00 PM | 8/28/15, 8:30 PM |
Note: Outage during major utility power failure in Seattle Id: 1222 By: knuth |
RS03AXBS |
1/7/16, 1:10 AM | 1/7/16, 1:52 AM |
Note: Four 1-minute outages between Portland and Seattle due to maintenance Id: 1223 By: knuth |
RS03AXBS |
3/10/16, 6:06 PM | 3/11/16, 4:30 AM |
Note: Fiber break between Portland and Seattle Id: 1224 By: knuth |
RS03AXBS |
5/20/16, 12:33 PM | 5/20/16, 2:04 PM |
Note: Fiber break between Portland and Pacific City Id: 1226 By: knuth |
RS03AXBS-LJ03A-12-CTDPFB301 Method: streamed Stream: ctdpf_optode_sample |
6/15/16, 8:57 PM | 6/19/16, 2:14 PM |
Data interruption.
Id: 1227
By: knuth
|
RS03AXBS-LJ03A-12-CTDPFB301 Method: streamed Stream: ctdpf_optode_sample |
7/11/16, 8:00 PM | 8/1/17, 8:41 AM |
The oxygen products from this instrument are suspect, due to errors during vendor calibration. Until a correction can be applied, the oxygen data during this time range should not be used.
Id: 3
By: friedrich.knuth
|
RS03AXBS-LJ03A-12-CTDPFB301 Method: streamed Stream: ctdpf-optode-sample |
7/11/16, 8:00 PM | 7/11/17, 8:00 PM |
The derived O2 products (L1 Temperature Compensated and L2 Salinity Corrected Dissolved Oxygen) are suspect, due to errors during vendor calibration. Until a correction can be applied, they should not be used from 7/12/2016 onwards. Id: 14 |
RS03AXBS |
7/11/16, 10:53 PM | 7/11/16, 11:51 PM |
Note: Unexplained loss of power at Pittock Building in Portland Id: 1228 By: knuth |
RS03AXBS |
12/17/16, 1:00 PM | 12/17/16, 2:00 PM |
Note: Corvalis data center lost power Id: 1229 By: knuth |
RS03AXBS |
12/21/16, 8:50 PM | 12/23/16, 7:44 AM |
Note: Fiber break in Portland due to train crash Id: 1230 By: knuth |
RS03AXBS |
1/8/17, 2:58 PM | 1/8/17, 4:41 PM |
Note: Network outage during major Seattle utility power failure Id: 1231 By: knuth |
RS03AXBS |
1/9/17, 1:32 PM | 1/9/17, 4:30 PM |
Note: Lightning strike in Pacific City led to data interruption through isolation of both cable lines from shore station equipment. Id: 1232 By: knuth |
RS03AXBS |
2/7/17, 8:00 AM | 2/7/17, 10:00 AM |
Note: Outage during PNWGP 1-hour router-maintenance Id: 1233 By: knuth |
RS03AXBS-LJ03A-12-CTDPFB301 Method: streamed Stream: ctdpf_optode_sample |
2/15/17, 9:43 AM | 2/16/17, 5:09 PM |
Data interruption.
Id: 1234
By: knuth
|
RS03AXBS |
2/15/17, 9:43 AM | 2/15/17, 12:16 PM |
Note: On Wednesday, February 15, power to the North and South cable lines of the Cabled Array was interrupted and all deployed infrastructure went offline. Power was restored on February 16. No hardware issue subsea was identified. Id: 1235 By: knuth |
RS03AXBS |
1/18/18, 10:00 PM | 1/19/18, 2:00 PM |
A brief violent storm caused loss of power to both the North and South cable lines and the deployed infrastructure during this time. For more information please contact help@oceanobservatories.org. (Redmine # 13109)
Id: 326
By: friedrich.knuth
|
RS03AXBS |
8/22/18, 1:00 PM | 8/24/18, 4:00 PM |
On Wednesday Aug 22 2018, an automatic protective shutdown of primary power to the OOI Cabled Array occurred during installation of a new chiller system. Power was restored on Aug 24. Due to loss of power to the Cabled Array marine infrastructure, data were not collected by the instruments.
Id: 1491
By: lgarzio
|