Asset: CGINS-FLORTK-01119

Type
Sensor
Mobile
0
Description of Equipment
3-Wavelength Fluorometer: FLORT Series K
Manufacturer
WET Labs
Model
ECO Triplet
Manufacturer Serial No
1119
Firmware Version
Acquisition Date
1/28/14
Original Cost
$0.00
Comments
Deployment Cruise Reference Designator Deployment Number Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
KN217 CP02PMCI-WFP01-04-FLORTK000 1 04/16/2014 10/10/2014 CGMCP-02PMCI-00001 CGVEH-CWFPCP-10302 CGINS-FLORTK-01119 40.2267 -70.8886 104 127
AT27 CP02PMCI-WFP01-04-FLORTK000 3 05/01/2015 10/15/2015 CGMCP-02PMCI-00003 CGVEH-CWFPCP-10302 CGINS-FLORTK-01119 40.2267 -70.8888 105 128
AR4 CP02PMUI-WFP01-04-FLORTK000 7 05/21/2016 10/04/2016 CGMCP-02PMUI-00007 CGVEH-CWFPCP-10302 CGINS-FLORTK-01119 40.3653 -70.7809 68 92
AR18 CP02PMCI-WFP01-04-FLORTK000 8 06/16/2017 11/07/2017 CGMCP-02PMCI-00008 CGVEH-CWFPCP-10302 CGINS-FLORTK-01119 40.227 -70.8889 104 127
AR28 CP02PMCI-WFP01-04-FLORTK000 10 04/11/2018 11/05/2018 CGMCP-02PMCI-00010 CGVEH-CWFPCP-10302 CGINS-FLORTK-01119 40.2265 -70.8889 104 127
AR34 CP02PMCI-WFP01-04-FLORTK000 12 04/17/2019 10/07/2019 CGMCP-02PMCI-00012 CGVEH-CWFPCP-10302 CGINS-FLORTK-01119 40.2267 -70.8886 103 126
AR44 CP02PMCI-WFP01-04-FLORTK000 14 06/12/2020 CGMCP-02PMCI-00014 CGVEH-CWFPCP-10302 CGINS-FLORTK-01119 40.2267 -70.889 104 127

No notes yet.

Class Asset UID Start Date Serial Name Value Notes
FLORTK CGINS-FLORTK-01119 09/27/2013 1119 CC_angular_resolution 1.076
FLORTK CGINS-FLORTK-01119 09/27/2013 1119 CC_dark_counts_cdom 60.0
FLORTK CGINS-FLORTK-01119 09/27/2013 1119 CC_dark_counts_chlorophyll_a 50.0
FLORTK CGINS-FLORTK-01119 09/27/2013 1119 CC_dark_counts_volume_scatter 49.0
FLORTK CGINS-FLORTK-01119 09/27/2013 1119 CC_depolarization_ratio 0.039
FLORTK CGINS-FLORTK-01119 09/27/2013 1119 CC_measurement_wavelength 700.0
FLORTK CGINS-FLORTK-01119 09/27/2013 1119 CC_scale_factor_cdom 0.0904
FLORTK CGINS-FLORTK-01119 09/27/2013 1119 CC_scale_factor_chlorophyll_a 0.0125
FLORTK CGINS-FLORTK-01119 09/27/2013 1119 CC_scale_factor_volume_scatter 3.095e-06
FLORTK CGINS-FLORTK-01119 09/27/2013 1119 CC_scattering_angle 124.0
FLORTK CGINS-FLORTK-01119 02/12/2015 1119 CC_angular_resolution 1.076
FLORTK CGINS-FLORTK-01119 02/12/2015 1119 CC_dark_counts_cdom 47.0
FLORTK CGINS-FLORTK-01119 02/12/2015 1119 CC_dark_counts_chlorophyll_a 50.0
FLORTK CGINS-FLORTK-01119 02/12/2015 1119 CC_dark_counts_volume_scatter 48.0
FLORTK CGINS-FLORTK-01119 02/12/2015 1119 CC_depolarization_ratio 0.039
FLORTK CGINS-FLORTK-01119 02/12/2015 1119 CC_measurement_wavelength 700.0
FLORTK CGINS-FLORTK-01119 02/12/2015 1119 CC_scale_factor_cdom 0.0805
FLORTK CGINS-FLORTK-01119 02/12/2015 1119 CC_scale_factor_chlorophyll_a 0.012
FLORTK CGINS-FLORTK-01119 02/12/2015 1119 CC_scale_factor_volume_scatter 3.133e-06
FLORTK CGINS-FLORTK-01119 02/12/2015 1119 CC_scattering_angle 124.0
FLORTK CGINS-FLORTK-01119 12/30/2015 1119 CC_angular_resolution 1.076
FLORTK CGINS-FLORTK-01119 12/30/2015 1119 CC_dark_counts_cdom 47.0
FLORTK CGINS-FLORTK-01119 12/30/2015 1119 CC_dark_counts_chlorophyll_a 49.0
FLORTK CGINS-FLORTK-01119 12/30/2015 1119 CC_dark_counts_volume_scatter 48.0
FLORTK CGINS-FLORTK-01119 12/30/2015 1119 CC_depolarization_ratio 0.039
FLORTK CGINS-FLORTK-01119 12/30/2015 1119 CC_measurement_wavelength 700.0
FLORTK CGINS-FLORTK-01119 12/30/2015 1119 CC_scale_factor_cdom 0.072
FLORTK CGINS-FLORTK-01119 12/30/2015 1119 CC_scale_factor_chlorophyll_a 0.0114
FLORTK CGINS-FLORTK-01119 12/30/2015 1119 CC_scale_factor_volume_scatter 3.069e-06
FLORTK CGINS-FLORTK-01119 12/30/2015 1119 CC_scattering_angle 124.0
FLORTK CGINS-FLORTK-01119 12/12/2016 1119 CC_angular_resolution 1.076
FLORTK CGINS-FLORTK-01119 12/12/2016 1119 CC_dark_counts_cdom 47
FLORTK CGINS-FLORTK-01119 12/12/2016 1119 CC_dark_counts_chlorophyll_a 50
FLORTK CGINS-FLORTK-01119 12/12/2016 1119 CC_dark_counts_volume_scatter 48
FLORTK CGINS-FLORTK-01119 12/12/2016 1119 CC_depolarization_ratio 0.039
FLORTK CGINS-FLORTK-01119 12/12/2016 1119 CC_measurement_wavelength 700
FLORTK CGINS-FLORTK-01119 12/12/2016 1119 CC_scale_factor_cdom 0.0661
FLORTK CGINS-FLORTK-01119 12/12/2016 1119 CC_scale_factor_chlorophyll_a 0.0115
FLORTK CGINS-FLORTK-01119 12/12/2016 1119 CC_scale_factor_volume_scatter 3.164E-06
FLORTK CGINS-FLORTK-01119 12/12/2016 1119 CC_scattering_angle 124
FLORTK CGINS-FLORTK-01119 01/26/2018 1119 CC_angular_resolution 1.076
FLORTK CGINS-FLORTK-01119 01/26/2018 1119 CC_dark_counts_cdom 50
FLORTK CGINS-FLORTK-01119 01/26/2018 1119 CC_dark_counts_chlorophyll_a 51
FLORTK CGINS-FLORTK-01119 01/26/2018 1119 CC_dark_counts_volume_scatter 49
FLORTK CGINS-FLORTK-01119 01/26/2018 1119 CC_depolarization_ratio 0.039
FLORTK CGINS-FLORTK-01119 01/26/2018 1119 CC_measurement_wavelength 700
FLORTK CGINS-FLORTK-01119 01/26/2018 1119 CC_scale_factor_cdom 0.0624
FLORTK CGINS-FLORTK-01119 01/26/2018 1119 CC_scale_factor_chlorophyll_a 0.0057
FLORTK CGINS-FLORTK-01119 01/26/2018 1119 CC_scale_factor_volume_scatter 3.426E-06
FLORTK CGINS-FLORTK-01119 01/26/2018 1119 CC_scattering_angle 124
FLORTK CGINS-FLORTK-01119 01/11/2019 1119 CC_angular_resolution 1.076
FLORTK CGINS-FLORTK-01119 01/11/2019 1119 CC_dark_counts_cdom 51
FLORTK CGINS-FLORTK-01119 01/11/2019 1119 CC_dark_counts_chlorophyll_a 52
FLORTK CGINS-FLORTK-01119 01/11/2019 1119 CC_dark_counts_volume_scatter 49
FLORTK CGINS-FLORTK-01119 01/11/2019 1119 CC_depolarization_ratio 0.039
FLORTK CGINS-FLORTK-01119 01/11/2019 1119 CC_measurement_wavelength 700
FLORTK CGINS-FLORTK-01119 01/11/2019 1119 CC_scale_factor_cdom 0.0562
FLORTK CGINS-FLORTK-01119 01/11/2019 1119 CC_scale_factor_chlorophyll_a 0.0114
FLORTK CGINS-FLORTK-01119 01/11/2019 1119 CC_scale_factor_volume_scatter 3.410E-06
FLORTK CGINS-FLORTK-01119 01/11/2019 1119 CC_scattering_angle 124
FLORTK CGINS-FLORTK-01119 12/10/2019 1119 CC_angular_resolution 1.076 source_file FLORT-K_BBFL2_SN_1119_Calibration_2019-12-10.zip > BBFL2-1119 Charsheets.pdf
FLORTK CGINS-FLORTK-01119 12/10/2019 1119 CC_dark_counts_cdom 51
FLORTK CGINS-FLORTK-01119 12/10/2019 1119 CC_dark_counts_chlorophyll_a 51
FLORTK CGINS-FLORTK-01119 12/10/2019 1119 CC_dark_counts_volume_scatter 48
FLORTK CGINS-FLORTK-01119 12/10/2019 1119 CC_depolarization_ratio 0.039
FLORTK CGINS-FLORTK-01119 12/10/2019 1119 CC_measurement_wavelength 700
FLORTK CGINS-FLORTK-01119 12/10/2019 1119 CC_scale_factor_cdom 0.0550
FLORTK CGINS-FLORTK-01119 12/10/2019 1119 CC_scale_factor_chlorophyll_a 0.0111
FLORTK CGINS-FLORTK-01119 12/10/2019 1119 CC_scale_factor_volume_scatter 3.471E-06
FLORTK CGINS-FLORTK-01119 12/10/2019 1119 CC_scattering_angle 124