Data Stream Report

Instrument Name
3-Axis Motion Pack
Reference Designator
CE04OSSM-SBD11-01-MOPAK0000
Method
telemetered
Stream
mopak_o_dcl_accel
Stream Type
Engineering
Stream Content
Accelerometer Data
Uframe Route
Driver
mopak_o_dcl_telemetered_driver
Parser
cg_stc_eng_stc

Annotations

Metadata Start Date End Date Comment
CE04OSSM-SBD11-01-MOPAK0000
Method: telemetered
Stream: mopak_o_dcl_accel
10/23/18, 2:20 PM

Due to the presence of Class II SD cards in the data logger, data acquisition from the MOPAK has been disabled for the duration of the deployment. The high data rate of the MOPAK (10 Hz for 20 minutes) is consider potentially damaging to the SD card due to their slower read/write speeds. This issue has been addressed internally and steps have been taken to ensure systems are not deployed with Class II SD cards in the future.

Id: 1541 By: wingardc
Flag: not_evaluated Exclude: No

Parameters

Id Parameter Display Name Standard Name Unit Fill Value Precision Identifier Type Level
PD7 time Time, UTC time seconds since 1900-01-01 -9999999
PD10 port_timestamp Port Timestamp, UTC seconds since 1900-01-01 -9999999
PD11 driver_timestamp Driver Timestamp, UTC seconds since 1900-01-01 -9999999
PD12 internal_timestamp Internal Timestamp, UTC seconds since 1900-01-01 -9999999
PD16 preferred_timestamp Preferred Timestamp 1 empty
PD863 ingestion_timestamp Ingestion Timestamp, UTC seconds since 1900-01-01 -9999
PD1595 mopak_accelx Acceleration X g0 -9999999
PD1596 mopak_accely Acceleration Y g0 -9999999
PD1597 mopak_accelz Acceleration Z g0 -9999999
PD1598 mopak_ang_ratex Angular Rate X rad s-1 -9999999
PD1599 mopak_ang_ratey Angular Rate Y rad s-1 -9999999
PD1600 mopak_ang_ratez Angular Rate Z rad s-1 -9999999
PD1601 mopak_magx Magnetometer Vector X Gs -9999999
PD1602 mopak_magy Magnetometer Vector Y Gs -9999999
PD1603 mopak_magz Magnetometer Vector Z Gs -9999999
PD1604 mopak_timer Timer counts 0