3-Wavelength Fluorometer

Reference Designator
CE02SHSM-RID27-02-FLORTD000
Review Status
Review Complete
Note
Depth
7m
Class
FLORT (3-Wavelength Fluorometer)
Make / Model
WET Labs / ECO Triplet-w

Dataset Reviews Last processed: 7/2/19, 12:16 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host flort_sample 174 174 0 0 0 0 2,461,437 1 7 / Complete
2 recovered_host flort_sample 223 223 0 0 0 0 3,169,994 1 7 / Complete
3 recovered_host flort_sample 135 135 0 0 1 1 1,874,826 1 7 / Complete
4 recovered_host flort_sample 209 209 0 0 0 0 2,950,737 1 7 / Complete
5 recovered_host flort_sample 179 14 0 153 3 13 133,141 1 7 / Complete
6 recovered_host flort_sample 177 166 1 0 2 11 2,483,807 1 7 / 1 Complete
7 recovered_host flort_sample 176 147 0 0 3 32 835,441 1 7 / Complete
Data Ranges Review Images

Test Notes

  1. fail: data found in another stream (gaps: [1] days: [2])

Data Coverage

Deployment: 1234567
flort_sample 100%100%100%100%8%94%84%

Lat/Lon Differences (km)

Deployment: 1234567
1 0.00
2 0.330.00
3 0.290.090.00
4 0.110.440.400.00
5 0.280.090.010.390.00
6 0.110.440.400.000.390.00
7 0.290.090.000.400.010.400.00

Example Composite Data Plot

System Annotations

Metadata Start Date End Date Comment
CE02SHSM-RID27
Method: telemetered
11/22/19, 3:45 PM 1/7/20, 1:30 PM

Due to reduced power on the mooring (limited recharging of the batteries with light winds and sea lions blocking the solar panels), we have had to reduce the operating schedule of the 7 meter platform. Normally we run the system continuously, but have reduced that schedule to hourly.

Id: 1836 By: wingardc

CE02SHSM-RID27
Method: telemetered
1/13/20, 1:45 PM

During a large storm event on 2020-01-11, power from the wind turbines was lost. This is the primary means of recharging the battery system on the mooring. With the loss of recharging power, we have had to implement a conservative power model. These systems and instruments have been disabled for the foreseeable future, and no further data will collected from them.

Id: 1903 By: wingardc
Flag: not_operational Exclude: No

Review Notes

Metadata Start Date End Date Comment
CE02SHSM-RID27-02-FLORTD000

The variable pressure_depth (dbar) is an array of all fill values for every deployment. This should contain valid pressure data.

By Lori Garzio, on 7/3/19

CE02SHSM-RID27-02-FLORTD000
Deployment: 2
3/1/16, 12:00 AM 5/16/16, 9:09 PM

Suspect data at the end of the deployment - negative chl-a and CDOM values, total_volume_scattering_coefficient maxes out.

By Lori Garzio, on 7/3/19

CE02SHSM-RID27-02-FLORTD000
Deployment: 5

The missing data for deployment 5 should be annotated.

By Lori Garzio, on 3/26/19

CE02SHSM-RID27-02-FLORTD000
Deployment: 6
12/1/17, 12:00 AM 4/4/18, 1:57 PM

total_volume_scattering_coefficient data show a repetitive pattern of increasing to the upper detection limit of the instrument and dropping to zero. A similar issue has been seen on CE06ISSM-RID16-02-FLORTD000 (see annotation ID 97). chl-a and CDOM data are also suspect. This dataset should be annotated.

By Lori Garzio, on 7/3/19

CE02SHSM-RID27-02-FLORTD000
Deployment: 7

Data gaps should be annotated.

By Lori Garzio, on 7/3/19

New Note