pCO2 Air-Sea

Reference Designator
CE02SHSM-SBD12-04-PCO2AA000
Review Status
Review Complete
Note
Depth
0m
Class
PCO2A (pCO2 Air-Sea)
Make / Model
Pro-Oceanus / pCO2-pro

Dataset Reviews Last processed: 7/23/19, 7:27 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host 174 158 0 15 0 0 33,939 1 / 1 2 2 Complete
1 recovered_host 174 158 0 15 0 0 33,939 1 1 / 1 2 2 Complete
2 recovered_host 223 223 0 0 0 0 47,952 1 / 2 2 Complete
2 recovered_host 223 223 0 0 0 0 47,814 1 1 / 2 2 Complete
3 recovered_host 135 135 0 0 0 0 28,890 1 / 2 2 Complete
3 recovered_host 135 135 0 0 0 0 28,890 1 1 / 2 2 Complete
4 recovered_host 209 207 0 0 1 2 44,244 1 / 2 2 Complete
4 recovered_host 209 207 0 0 1 2 44,231 1 1 / 2 2 Complete
5 recovered_host 179 178 0 1 0 0 38,214 1 / 2 2 Complete
5 recovered_host 179 178 0 1 0 0 38,211 1 1 / 2 2 Complete
6 recovered_host 177 172 5 0 0 0 36,891 1 / 2 2 Complete
6 recovered_host 177 172 5 0 0 0 36,891 1 1 / 2 2 Complete
7 recovered_host 176 140 1 35 1 1 28,269 1 / 2 2 Complete
7 recovered_host 176 140 1 35 1 1 28,269 1 1 / 2 2 Complete
Data Ranges Review Images

Test Notes

  1. pass unique test - ascending not tested
  2. no other streams for comparison

Data Coverage

Deployment: 1234567
91%100%100%99%99%97%80%
91%100%100%99%99%97%80%

Lat/Lon Differences (km)

Deployment: 1234567
1 0.00
2 0.330.00
3 0.290.090.00
4 0.110.440.400.00
5 0.280.090.010.390.00
6 0.110.440.400.000.390.00
7 0.290.090.000.400.010.400.00

Example Composite Data Plot

System Annotations

Metadata Start Date End Date Comment
CE02SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_air
4/2/15, 4:30 PM 9/22/15, 3:30 AM

PCO2A air measurement for the Spring 2015 deployment is considered suspect. Recent investigations have indicated a vendor error in assembly may have pinched closed an internal tube carrying the surface air stream into the sample volume resulting in an erroneous measurement.

Id: 1932 By: wingardc
Flag: suspect Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_air
4/2/15, 4:30 PM 9/22/15, 3:30 AM

PCO2A air measurement for the Spring 2015 deployment is considered suspect. Recent investigations have indicated a vendor error in assembly may have pinched closed an internal tube carrying the surface air stream into the sample volume resulting in an erroneous measurement.

Id: 1938 By: wingardc
Flag: suspect Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_air_recovered
4/2/18, 9:30 PM 9/25/18, 12:30 PM

A vendor error in assembly pinched closed an internal tube carrying the surface air stream into the sample volume, resulting in an erroneous measurement. Users should not use the PCO2A air measurement for the Spring 2018 deployment (deployment 07), nor the associated sea-air pCO2 flux estimate.

Id: 1939 By: wingardc
Flag: fail Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_air
4/2/18, 9:30 PM 9/25/18, 12:30 PM

UPDATED, 2020-01-28: The QC flag for the air measurement has been reset to fail. Users should not use the PCO2A air measurement for the Spring 2018 deployment (deployment 07), nor the associated sea-air pCO2 flux estimate. *PCO2A air measurement for the Spring 2018 deployment is considered suspect. Recent investigations have indicated a vendor error in assembly may have pinched closed an internal tube carrying the surface air stream into the sample volume resulting in an erroneous measurement.

Id: 1446 By: wingardc
Flag: fail Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_water
4/2/18, 9:30 PM 9/25/18, 12:30 PM

A vendor error in assembly pinched closed an internal tube carrying the surface air stream into the sample volume, resulting in an erroneous measurement. Users should not use the PCO2A air measurement for the Spring 2018 deployment (deployment 07), nor the associated sea-air pCO2 flux estimate.

Id: 1988 By: wingardc
Flag: fail Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_water_recovered
4/2/18, 9:30 PM 9/25/18, 12:30 PM

A vendor error in assembly pinched closed an internal tube carrying the surface air stream into the sample volume, resulting in an erroneous measurement. Users should not use the PCO2A air measurement for the Spring 2018 deployment (deployment 07), nor the associated sea-air pCO2 flux estimate.

Id: 1989 By: wingardc
Flag: fail Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_air
9/27/18, 11:30 PM 9/28/18, 2:30 PM

A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.

Id: 1933 By: wingardc
Flag: fail Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_air_recovered
9/27/18, 11:30 PM 9/28/18, 2:30 PM

A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.

Id: 1940 By: wingardc
Flag: fail Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_air
10/29/18, 7:30 PM 10/30/18, 10:30 AM

A HITL review of the data has found unusual, suspect data during this time range. The data during this time range is marked as suspect, and users should exercise caution using this data in their analysis.

Id: 1934 By: wingardc
Flag: suspect Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_air_recovered
10/29/18, 7:30 PM 10/30/18, 10:30 AM

A HITL review of the data has found unusual, suspect data during this time range. The data during this time range is marked as suspect, and users should exercise caution using this data in their analysis.

Id: 1941 By: wingardc
Flag: suspect Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_air
11/6/18, 11:30 AM 11/29/18, 8:30 PM

A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.

Id: 1935 By: wingardc
Flag: fail Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_air_recovered
11/6/18, 11:30 AM 11/29/18, 8:30 PM

A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.

Id: 1942 By: wingardc
Flag: fail Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_water
11/6/18, 11:30 AM 11/29/18, 8:30 PM

A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.

Id: 1946 By: wingardc
Flag: fail Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_water_recovered
11/6/18, 11:30 AM 11/29/18, 8:30 PM

A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.

Id: 1947 By: wingardc
Flag: fail Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_water
11/11/18, 2:30 AM 4/21/19, 11:30 AM

CO2 sensor internal to the instrument has failed. Sensor started reporting erroneous values for the CO2 measurements and then flat-lined to 0.

Id: 1539 By: wingardc
Flag: fail Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_air_recovered
11/11/18, 2:30 AM 4/21/19, 11:30 AM

CO2 sensor internal to the instrument has failed. Sensor started reporting erroneous values for the CO2 measurements and then flat-lined to 0.

Id: 1943 By: wingardc
Flag: fail Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_water_recovered
11/11/18, 2:30 AM 4/21/19, 11:30 AM

CO2 sensor internal to the instrument has failed. Sensor started reporting erroneous values for the CO2 measurements and then flat-lined to 0.

Id: 1948 By: wingardc
Flag: fail Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_air
11/11/18, 3:00 AM 4/21/19, 11:54 AM

Suspect CO2 sensor internal to the instrument has failed. Sensor started reporting erroneous values for the CO2 measurements and then flat-lined to 0.

Id: 1538 By: wingardc
Flag: fail Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: telemetered
4/19/19, 7:50 PM 8/15/19, 5:30 PM

UPDATED, 2019-09-04: The vendor provided a protocol to reset the zero measurement on a daily basis. This protocol has been implemented on the mooring and the pCO2 air and water measurements have reset. They are still somewhat low, and should be considered as suspect. The vendor will work to provided a finalized and corrected dataset after the moorings are recovered and the sensors are returned for servicing. *Due to a firmware bug, the sensor zeroing function for correcting the baseline drift over time is not currently saving the updated reference value used to compute the raw CO2 value outputs during the measurements. The sensors are zeroing at the end of the sample cycle when scheduled, but the firmware is not storing the new value. This results in the original zero reference value being used for all measurements and an observable and consistent negative drift in both the air and water CO2 values. The vendor has indicated that the data may be correctable post-recovery, however at this time no such correction algorithm has been provided.

Id: 1680 By: wingardc
Flag: suspect Exclude: Yes

CE02SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_air
10/1/19, 9:30 PM 10/2/19, 12:30 PM

A HITL review of the data has found unusual, suspect data during this time range. The data during this time range is marked as suspect, and users should exercise caution using this data in their analysis.

Id: 1936 By: wingardc
Flag: suspect Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_air_recovered
10/1/19, 9:30 PM 10/2/19, 12:30 PM

A HITL review of the data has found unusual, suspect data during this time range. The data during this time range is marked as suspect, and users should exercise caution using this data in their analysis.

Id: 1944 By: wingardc
Flag: suspect Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: telemetered
11/15/19, 1:30 PM 11/18/19, 12:30 PM

The instrument suffered a firmware crash and had to be power cycled in order to restore operations. We are in consultation with the vendor to determine the cause of this failure.

Id: 2052 By: wingardc
Flag: not_operational Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_air
12/29/19, 2:30 PM 12/30/19, 6:30 AM

A HITL review of the data has found unusual, suspect data during this time range. The data during this time range is marked as suspect, and users should exercise caution using this data in their analysis.

Id: 1937 By: wingardc
Flag: suspect Exclude: No

CE02SHSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_air_recovered
12/29/19, 2:30 PM 12/30/19, 6:30 AM

A HITL review of the data has found unusual, suspect data during this time range. The data during this time range is marked as suspect, and users should exercise caution using this data in their analysis.

Id: 1945 By: wingardc
Flag: suspect Exclude: No

CE02SHSM-SBD12
Method: telemetered
1/13/20, 1:45 PM

During a large storm event on 2020-01-11, power from the wind turbines was lost. This is the primary means of recharging the battery system on the mooring. With the loss of recharging power, we have had to implement a conservative power model. These systems and instruments have been disabled for the foreseeable future, and no further data will collected from them.

Id: 1900 By: wingardc
Flag: not_operational Exclude: No

Review Notes

Metadata Start Date End Date Comment
CE02SHSM-SBD12-04-PCO2AA000
Deployment: 1
4/2/15, 8:15 PM 9/22/15, 7:48 AM

Suspect measured_air_co2 - values get very low for this deployment (close to 360 ppm)

By Lori Garzio, on 7/26/19

CE02SHSM-SBD12-04-PCO2AA000
Deployment: 2
4/1/16, 12:00 AM 5/16/16, 9:09 PM

measured_water_co2 is suspect towards the end of the deployment (large swing from very high to low values).

By Lori Garzio, on 7/26/19

CE02SHSM-SBD12-04-PCO2AA000
Deployment: 4
9/26/16, 9:25 PM 11/1/16, 12:00 AM

measured_water_co2 is suspect towards the beginning of the deployment (large swing from very high to low values).

By Lori Garzio, on 7/26/19

CE02SHSM-SBD12-04-PCO2AA000
Deployment: 6
2/1/18, 12:00 AM 4/4/18, 1:57 PM

Several unreasonable values for measured_air_co2 in the second half of the deployment.

By Lori Garzio, on 7/26/19

CE02SHSM-SBD12-04-PCO2AA000
Deployment: 6
4/1/18, 12:00 AM 4/4/18, 1:57 PM

Suspect measured_water_co2 values at the end of the deployment

By Lori Garzio, on 7/26/19

CE02SHSM-SBD12-04-PCO2AA000
Deployment: 7
4/3/18, 1:17 AM 9/25/18, 4:16 PM

Suspect data - values for CO2 in the atmosphere are too low/variable (annotation ID 1446).

By Lori Garzio, on 7/26/19

New Note