Single Point Velocity Meter

Reference Designator
CE02SHSP-SP001-02-VELPTJ000
Review Status
Review Complete
Note
Velocity data from profilers were run through the automated QC tests, but were not plotted or reviewed visually.
Depth
0 to 80m
Class
VELPT (Single Point Velocity Meter)
Make / Model
Nortek / Aquadopp HR

Dataset Reviews Last processed: 6/28/19, 2:16 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_cspp 16 15 0 0 3 3 15,274 1 80 / 73 1 1 Complete
3 recovered_cspp 41 41 0 0 0 0 31,333 1 80 / 70 1 1 Complete
4 recovered_cspp 88 44 0 0 13 47 18,917 1 80 / 73 1 1 Complete
5 recovered_cspp 32 18 0 9 3 6 11,348 1 80 / 73 1 1 Complete
6 recovered_cspp 5 4 0 1 0 0 3,304 1 80 / 78 1 1 Complete
7 recovered_cspp 105 55 0 40 3 11 34,470 1 80 / 78 1 1 Complete
8 recovered_cspp 27 22 0 3 1 2 11,074 1 80 / 78 1 1 Complete
9 recovered_cspp 20 20 0 0 0 0 11,392 1 80 / 78 1 1 Complete
10 recovered_cspp / Complete
Data Ranges Review Images

Test Notes

  1. no other streams for comparison

Data Coverage

Deployment: 1345678910
94%100%50%56%80%52%81%100%

Lat/Lon Differences (km)

Deployment: 1345678910
1 0.00
3 0.880.00
4 0.080.960.00
5 0.230.950.190.00
6 0.360.560.420.400.00
7 0.470.680.490.370.250.00
8 0.260.800.330.480.410.620.00
9 0.660.790.670.510.450.210.830.00

System Annotations

Metadata Start Date End Date Comment
CE02SHSP-SP001-02-VELPTJ000
3/18/15, 3:36 PM

According to the manufacturer, data are suspect when the instrument is tilted more than 20 degrees (measured by pitch and roll).

Id: 800 By: lgarzio

CE02SHSP
3/8/17, 7:00 PM 4/17/17, 8:00 PM

The main batteries on CE02SHSP-00005 drained unexpectedly. CSPPs are deployed at 33.4 V. Voltage drops roughly linearly with each profile. Batteries were at 31 V at the beginning of this week. Yesterday (2017-03-09), it reported via acoustic modem that it was in a LOW POWER state, which it does automatically at 28 V to protect the main batteries. In the past, similar drains have been due to faulty connections between the main batteries and the profilers at the battery adapter plates. This has been the CSPP's main failure behavior.

Id: 882 By: michaesm
Flag: not_operational Exclude: No

CE02SHSP
4/24/17, 8:00 PM 10/4/17, 8:00 PM

CE02SHSP-00006 reported that it had surfaced unexpectedly April 25 at 5:10 pm local time. Within a few hours it was determined that it was still tethered to its anchor. It's beacon wasn't reporting either. The profiler and its anchor were recovered the following day. Data were sent to the profiler vendor. By April 28 we confirmed there was a firmware bug that had never been seen before. While moving between two positions, the control canister thought the profiler was 20 cm below its destination, but the profiler's winch thought the profiler was 20 cm above its destination, so the profiler was winched up to the surface, instead of retreating 20 cm to its parking depth. The bug will be corrected in the next firmware release. It also turns out that the beacon was programmed improperly, so it had run out of batteries. We switched to a new beacon model this year and confirmed with the vendor which settings to use.

Id: 883 By: michaesm

CE02SHSP
5/29/18, 8:00 PM 8/1/18, 4:30 PM

Profiler was profiling twice per day, but it has not surfaced as planned since May 30 ~1am PDT. It is deployed too far from CE02SHSM for us to have good acoustic communications (1150m vs. the planned 150-400m), so we cannot debug it remotely. It will need to be recovered, but it is not at the surface or adrift, so it does not need immediate recovery.

Id: 1458 By: michaesm

Review Notes

Metadata Start Date End Date Comment
CE02SHSP-SP001

During deployments 3 - 5, there are several pressure (dbar) data points that are outliers (much deeper than the max deployment depth of 80m).

By Lori Garzio, on 7/1/19

CE02SHSP-SP001-02-VELPTJ000
Deployment: 10

There are no data available for download for this deployment. According to the ingest csv, recovered raw data files are available. These files should be ingested or the dataset should be annotated.

By Lori Garzio, on 7/1/19

New Note