Spectral Irradiance

Reference Designator
CE04OSSM-RID26-08-SPKIRB000
Review Status
Review Complete
Note
Depth
7m
Class
SPKIR (Spectral Irradiance)
Make / Model
Satlantic / OCR507 ICSW

Dataset Reviews Last processed: 7/9/19, 8:01 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host 399 399 0 0 0 0 6,416,051 1 7 / 1 2 3 Complete
2 recovered_host 139 137 0 0 2 2 2,192,816 1 7 / 2 3 Complete
3 recovered_host 203 203 0 0 0 0 3,275,127 1 7 / 2 3 Complete
4 recovered_host 176 176 0 0 0 0 2,716,207 1 7 / 2 3 Complete
5 recovered_host 175 175 0 0 0 0 2,632,730 1 7 / 2 3 Complete
6 recovered_host 178 167 0 11 0 0 2,155,770 1 7 / 2 3 Complete
Data Ranges Review Images

Test Notes

  1. pass unique test - ascending not tested
  2. no comparison: timestamps do not match
  3. missing: ['pressure']

Data Coverage

Deployment: 123456
100%99%100%100%100%94%

Lat/Lon Differences (km)

Deployment: 123456
1 0.00
2 0.260.00
3 1.822.090.00
4 0.340.182.100.00
5 1.721.970.481.940.00
6 0.400.282.100.101.920.00

Example Composite Data Plot

System Annotations

Metadata Start Date End Date Comment
CE04OSSM-RID26
Method: telemetered
6/15/19, 6:35 AM 6/17/19, 3:25 PM

The datalogger computer crashed and had to be power cycled to restore operation. Telemetered data was lost for this time period.

Id: 1684 By: wingardc
Flag: not_operational Exclude: No

CE04OSSM-RID26-08-SPKIRB000
10/21/19, 1:39 PM

The use of UV lamps has been implemented for biofouling mitigation on this instrument as of 2019-10-21T17:39:00Z.

Id: 2122 By: wingardc

CE04OSSM
Method: telemetered
5/6/20, 10:40 PM 5/11/20, 8:00 PM

Due to a bug in the mooring telemetry system, data during this time range was not telemetered to shore. It was recorded on the mooring and instruments and will be available in the recovered data streams.

Id: 2181 By: wingardc
Flag: not_available Exclude: No

Review Notes

Metadata Start Date End Date Comment
CE04OSSM-RID26-08-SPKIRB000

There is no pressure variable in the .nc files. This should be a coordinate.

By Lori Garzio, on 7/12/19

CE04OSSM-RID26-08-SPKIRB000

A substantial percentage (between 18-47%) of data are outside of global ranges for every deployment. Zooming in on the x-axis in the plots shows that all of the values that are outside of global ranges are negative.

By Lori Garzio, on 7/12/19

New Note