Dep. | Preferred Method | Stream | DD | FD | SG | EG | Gaps | GD | TS | Rate (s) | Pressure Comp. | Time Order | Valid Data | Missing Data | Data Comp. | Missing Coords. | Review |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1 | recovered_host | flort_sample | 399 | 399 | 0 | 0 | 0 | 0 | 5,680,072 | 1 | 7 / | Complete | |||||
2 | recovered_host | flort_sample | 139 | 139 | 0 | 0 | 0 | 0 | 1,977,297 | 1 | 7 / | Complete | |||||
3 | recovered_host | flort_sample | 203 | 203 | 0 | 0 | 0 | 0 | 2,894,933 | 1 | 7 / | Complete | |||||
4 | recovered_host | flort_sample | 176 | 176 | 0 | 0 | 0 | 0 | 2,398,283 | 1 | 7 / | Complete | |||||
5 | recovered_host | flort_sample | 175 | 173 | 0 | 2 | 0 | 0 | 2,290,524 | 1 | 7 / | 1 | Complete | ||||
6 | recovered_host | flort_sample | 178 | 167 | 0 | 11 | 0 | 0 | 1,896,341 | 1 | 7 / | Complete |
Deployment: | 1 | 2 | 3 | 4 | 5 | 6 |
---|---|---|---|---|---|---|
flort_sample | 100% | 100% | 100% | 100% | 99% | 94% |
Deployment: | 1 | 2 | 3 | 4 | 5 | 6 |
---|---|---|---|---|---|---|
1 | 0.00 | |||||
2 | 0.26 | 0.00 | ||||
3 | 1.82 | 2.09 | 0.00 | |||
4 | 0.34 | 0.18 | 2.10 | 0.00 | ||
5 | 1.72 | 1.97 | 0.48 | 1.94 | 0.00 | |
6 | 0.40 | 0.28 | 2.10 | 0.10 | 1.92 | 0.00 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CE04OSSM-RID27-02-FLORTD000 Method: telemetered |
3/19/20, 8:00 PM |
The optical backscatter and chlorophyll fluoresence values are unreasonable and highly variable. The data would seem to indicate that the sensor is fouled. Users should treat the data during this time range as suspect.
Id: 2061
By: wingardc
|
|
CE04OSSM Method: telemetered |
5/6/20, 10:40 PM | 5/11/20, 8:00 PM |
Due to a bug in the mooring telemetry system, data during this time range was not telemetered to shore. It was recorded on the mooring and instruments and will be available in the recovered data streams.
Id: 2181
By: wingardc
|
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CE04OSSM-RID27-02-FLORTD000 |
The variable pressure_depth (dbar) is an array of all fill values for every deployment. This should contain valid pressure data. By Lori Garzio, on 7/3/19 |
||
CE04OSSM-RID27-02-FLORTD000 Deployment: 1 |
27% of optical_backscatter data for this deployment are fill values. By Lori Garzio, on 7/3/19 |
||
CE04OSSM-RID27-02-FLORTD000 Deployment: 1 |
6/20/15, 12:00 AM | 5/9/16, 2:53 PM | |
CE04OSSM-RID27-02-FLORTD000 Deployment: 4 |
5/25/17, 12:00 AM | 10/13/17, 4:28 PM |