Dissolved Oxygen

Reference Designator
CE05MOAS-GL320-04-DOSTAM000
Review Status
Review Complete
Note
Depth
0 to 1,000m
Class
DOSTA (Dissolved Oxygen Stable Response)
Make / Model
Aanderaa / Optode 4831

Dataset Reviews Last processed: 8/23/19, 7:11 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host 84 79 0 4 0 0 3,082,772 2 1000 / 1007 Complete
2 recovered_host 104 95 0 8 0 0 2,163,165 2 1000 / 1000 Complete
3 recovered_host 43 42 0 1 0 0 971,073 2 1000 / 991 Complete
4 recovered_host 92 92 0 0 0 0 778,007 2 1000 / 993 Complete
Data Ranges Review Images

Test Notes

All good!

Data Coverage

Deployment: 1234
94%91%98%100%

Deployment Map

System Annotations

Metadata Start Date End Date Comment
CE05MOAS-GL320
10/5/14, 8:00 PM 12/28/14, 7:00 PM

Deployment 1 - Power consumption was greater than 1% per day and battery ran down to 10% before the end of the deployment. Glider was drifting before it was recovered.

Id: 890 By: michaesm
Flag: not_operational Exclude: No

Review Notes

Metadata Start Date End Date Comment
CE05MOAS

Bad altimeter pings are apparent in many deployments of the coastal gliders. A few example plots are attached - m_water_depth plotted as semi-transparent bathymetry.

By Lori Garzio, on 5/10/19

CE05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of 3 consecutive down and up casts for CE05MOAS-GL312-04-DOSTAM000 on 7/8/2014 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/14/19

New Note