3-Wavelength Fluorometer

Reference Designator
CE05MOAS-GL384-02-FLORTM000
Review Status
Review Complete
Note
Depth
0 to 1,000m
Class
FLORT (3-Wavelength Fluorometer)
Make / Model
WET Labs / ECO Puck FLBBCD-SLK

Dataset Reviews Last processed: 8/24/19, 10:44 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host flort_m_sample 89 88 0 0 0 0 3,177,364 1 1000 / 1018 Complete
2 recovered_host flort_m_sample 94 81 0 12 0 0 3,252,253 1 1000 / 1020 1 Complete
3 recovered_host flort_m_sample 74 72 0 0 0 0 3,025,468 1 1000 / 1016 Complete
4 recovered_host flort_m_sample 91 91 0 0 0 0 3,748,100 1 1000 / 1016 2 Complete
5 recovered_host flort_m_sample 89 89 0 0 0 0 3,651,334 1 1000 / 1014 1 Complete
Data Ranges Review Images

Test Notes

  1. fail: data found in another stream (gaps: [2] days: [2])
  2. fail: data found in another stream (gaps: [1] days: [1])

Data Coverage

Deployment: 12345
flort_m_sample 99%86%97%100%100%

Deployment Map

System Annotations

No annotations yet.

Review Notes

Metadata Start Date End Date Comment
CE05MOAS

Bad altimeter pings are apparent in many deployments of the coastal gliders. A few example plots are attached - m_water_depth plotted as semi-transparent bathymetry.

By Lori Garzio, on 5/10/19

CE05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of 3 consecutive down and up casts for CE05MOAS-GL312-04-DOSTAM000 on 7/8/2014 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/14/19

New Note