Velocity Profiler (600kHz)

Reference Designator
CE07SHSM-RID26-01-ADCPTA000
Review Status
Review Complete
Note
Depth
7m
Class
ADCPT (Velocity Profiler (short range))
Make / Model
Teledyne RDI / WorkHorse Sentinel 600khz

Dataset Reviews Last processed: 8/14/19, 7:30 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_inst adcp_velocity_earth 162 67 0 95 0 0 6,310 900 7 / 1 2 Complete
2 recovered_inst adcp_velocity_earth 214 214 0 0 0 0 20,463 900 7 / 3 3 Complete
3 recovered_inst adcp_velocity_earth 146 51 0 95 0 0 4,765 900 7 / 2 Complete
4 recovered_inst adcp_velocity_earth 206 206 0 0 0 0 19,684 900 7 / 2 Complete
5 recovered_inst adcp_velocity_earth 179 36 0 143 0 0 3,210 900 7 / 2 Complete
6 recovered_inst adcp_velocity_earth 178 178 0 0 0 0 16,990 900 7 / 2 Complete
7 recovered_host adcp_velocity_earth 180 6 0 174 0 0 377 7 / Complete
Data Ranges Review Images

Test Notes

  1. fail: data found in another stream (gaps: [1] days: [0])
  2. no comparison: timestamps do not match
  3. no other streams for comparison

Data Coverage

Deployment: 1234567
adcp_velocity_earth 41%100%35%100%20%100%3%

Lat/Lon Differences (km)

Deployment: 1234567
1 0.00
2 0.280.00
3 0.440.220.00
4 0.020.290.460.00
5 0.250.170.210.270.00
6 0.180.460.610.170.400.00
7 0.270.100.170.290.080.440.00

System Annotations

Metadata Start Date End Date Comment
CE07SHSM-RID26-01-ADCPTA000
6/23/16, 8:00 PM 9/30/16, 8:00 PM

ADCP battery has been depleted. There is probably a leak or short somewhere.

Id: 923 By: michaesm
Flag: not_operational Exclude: No

CE07SHSM
12/1/17, 9:35 AM 3/25/18, 3:09 PM

Data loss due to a power system failure. The mooring stopped receiving power from photo voltaic panel 1 of 4 on Nov-2. By Nov-20, the mooring stopped receiving power from two more solar panels (PV2 & PV4) and one of two wind turbines (WT2). This indicates the Power System Controller (PSC) they connect to is failing. Mooring components will be shut down as the mooring loses the ability to recharge its batteries.

Id: 1237 By: crisien
Flag: not_operational Exclude: No

CE07SHSM-RID26-01-ADCPTA000
Method: telemetered
Stream: adcp_velocity_earth
1/24/19, 7:15 PM

Generated serial line noise has disrupted the ADCPs sampling schedule resulting in a shift in the instruments sampling time. To protect against further disruptions and possible damage to the ADCP, we have disabled the data logger port for the duration of the deployment.

Id: 1550 By: wingardc
Flag: not_operational Exclude: No

CE07SHSM-RID26-01-ADCPTA000
Method: telemetered
Stream: adcp_engineering
1/24/19, 7:15 PM

Generated serial line noise has disrupted the ADCPs sampling schedule resulting in a shift in the instruments sampling time. To protect against further disruptions and possible damage to the ADCP, we have disabled the data logger port for the duration of the deployment.

Id: 1551 By: wingardc
Flag: not_operational Exclude: No

CE07SHSM-RID26-01-ADCPTA000
Method: telemetered
Stream: adcp_config
1/24/19, 7:15 PM

Generated serial line noise has disrupted the ADCPs sampling schedule resulting in a shift in the instruments sampling time. To protect against further disruptions and possible damage to the ADCP, we have disabled the data logger port for the duration of the deployment.

Id: 1552 By: wingardc
Flag: not_operational Exclude: No

CE07SHSM
Method: telemetered
5/13/20, 6:40 PM

Due to a bug in the mooring telemetry system, data during this time range was not telemetered to shore. It was recorded on the mooring and instruments and will be available in the recovered data streams.

Id: 2185 By: wingardc
Flag: not_available Exclude: No

Review Notes

Metadata Start Date End Date Comment
CE07SHSM-RID26-01-ADCPTA000

The pressure variable (daPa) is an array of fill values for every deployment.

By Lori Garzio, on 8/20/19

CE07SHSM-RID26-01-ADCPTA000
Deployment: 7

There are no recovered_inst adcp_velocity_earth data available for download for this deployment. According to the ingest csv, raw recovered_inst data files are available. These should be ingested or the dataset should be annotated to explain why recovered_inst data aren't available.

By Lori Garzio, on 8/20/19

New Note