Velocity Profiler (600kHz)

Reference Designator
CE07SHSM-RID26-01-ADCPTA000
Start Depth
7
End Depth
7
Location
Near Surface Instrument Frame
Current Status
Review Complete
Class
ADCPT (Velocity Profiler (short range))
Series
ADCPT-A
Science Discipline
Physical
Make
Teledyne RDI
Model
WorkHorse Sentinel 600khz
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/CE07SHSM/RID26/01-ADCPTA000/metadata
Method Data Stream Content Type
recovered_host adcp_config Configuration Data Report M2M Engineering
recovered_host adcp_engineering Engineering Data Report M2M Engineering
recovered_host adcp_velocity_earth Earth Coordinates Data Products Report M2M Stats Science
recovered_inst adcp_config Configuration Data Report M2M Engineering
recovered_inst adcp_engineering Engineering Data Report M2M Engineering
recovered_inst adcp_velocity_earth Earth Coordinates Data Products Report M2M Stats Science
telemetered adcp_config Configuration Data Report M2M Engineering
telemetered adcp_engineering Engineering Data Report M2M Engineering
telemetered adcp_velocity_earth Earth Coordinates Data Products Report M2M Stats Science
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review OC1503D 04/05/2015 09/23/2015 CGMCE-07SHSM-00001 CGINS-ADCPTA-18222 46.9861 -124.566 7 87
2 Review OC1510A 10/13/2015 05/13/2016 CGMCE-07SHSM-00002 CGINS-ADCPTC-22809 46.9875 -124.569 7 87
3 Review TN-342 05/05/2016 09/27/2016 CGMCE-07SHSM-00003 CGINS-ADCPTA-19016 46.9895 -124.569 7 87
4 Review AT37-03 09/20/2016 04/13/2017 CGMCE-07SHSM-00004 CGINS-ADCPTA-22816 46.9859 -124.566 7 87
5 Review SKQ201704S 04/11/2017 10/06/2017 CGMCE-07SHSM-00005 CGINS-ADCPTA-19016 46.9882 -124.567 7 89
6 Review SKQ201715S 10/03/2017 03/29/2018 CGMCE-07SHSM-00006 CGINS-ADCPTA-22816 46.9852 -124.564 7 87
7 Review SKQ201808S 03/25/2018 09/20/2018 CGMCE-07SHSM-00007 CGINS-ADCPTA-19016 46.9881 -124.568 7 87
8 Review SR1813 09/18/2018 04/26/2019 CGMCE-07SHSM-00008 CGINS-ADCPTA-22816 46.9835 -124.564 7 87
9 Review SKQ201910S 04/24/2019 10/12/2019 CGMCE-07SHSM-00009 CGINS-ADCPTA-22814 46.9878 -124.568 7 87
10 Review SKQ201921S 10/10/2019 CGMCE-07SHSM-00010 CGINS-ADCPTA-19024 46.9841 -124.566 7 87
Metadata Start Date End Date Comment
CE07SHSM-RID26-01-ADCPTA000

The pressure variable (daPa) is an array of fill values for every deployment.

By Lori Garzio, on 8/20/19

CE07SHSM-RID26-01-ADCPTA000
Deployment: 7

There are no recovered_inst adcp_velocity_earth data available for download for this deployment. According to the ingest csv, raw recovered_inst data files are available. These should be ingested or the dataset should be annotated to explain why recovered_inst data aren't available.

By Lori Garzio, on 8/20/19

New Note

Metadata Start Date End Date Comment
CE07SHSM-RID26-01-ADCPTA000
6/23/16, 8:00 PM 9/30/16, 8:00 PM

ADCP battery has been depleted. There is probably a leak or short somewhere.

Id: 923 By: michaesm
Flag: not_operational Exclude: No

CE07SHSM
12/1/17, 9:35 AM 3/25/18, 3:09 PM

Data loss due to a power system failure. The mooring stopped receiving power from photo voltaic panel 1 of 4 on Nov-2. By Nov-20, the mooring stopped receiving power from two more solar panels (PV2 & PV4) and one of two wind turbines (WT2). This indicates the Power System Controller (PSC) they connect to is failing. Mooring components will be shut down as the mooring loses the ability to recharge its batteries.

Id: 1237 By: crisien
Flag: not_operational Exclude: No

CE07SHSM-RID26-01-ADCPTA000
Method: telemetered
Stream: adcp_velocity_earth
1/24/19, 7:15 PM

Generated serial line noise has disrupted the ADCPs sampling schedule resulting in a shift in the instruments sampling time. To protect against further disruptions and possible damage to the ADCP, we have disabled the data logger port for the duration of the deployment.

Id: 1550 By: wingardc
Flag: not_operational Exclude: No

CE07SHSM-RID26-01-ADCPTA000
Method: telemetered
Stream: adcp_engineering
1/24/19, 7:15 PM

Generated serial line noise has disrupted the ADCPs sampling schedule resulting in a shift in the instruments sampling time. To protect against further disruptions and possible damage to the ADCP, we have disabled the data logger port for the duration of the deployment.

Id: 1551 By: wingardc
Flag: not_operational Exclude: No

CE07SHSM-RID26-01-ADCPTA000
Method: telemetered
Stream: adcp_config
1/24/19, 7:15 PM

Generated serial line noise has disrupted the ADCPs sampling schedule resulting in a shift in the instruments sampling time. To protect against further disruptions and possible damage to the ADCP, we have disabled the data logger port for the duration of the deployment.

Id: 1552 By: wingardc
Flag: not_operational Exclude: No

CE07SHSM
Method: telemetered
5/13/20, 6:40 PM

Due to a bug in the mooring telemetry system, data during this time range was not telemetered to shore. It was recorded on the mooring and instruments and will be available in the recovered data streams.

Id: 2185 By: wingardc
Flag: not_available Exclude: No