Spectral Irradiance

Reference Designator
CE07SHSM-RID26-08-SPKIRB000
Review Status
Review Complete
Note
Depth
7m
Class
SPKIR (Spectral Irradiance)
Make / Model
Satlantic / OCR507 ICSW

Dataset Reviews Last processed: 7/9/19, 8:04 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host 162 162 0 0 0 0 2,099,295 1 7 / 1 2 Complete
2 recovered_host 214 202 0 0 2 12 3,245,928 1 7 / 1 2 Complete
3 recovered_host 146 146 0 0 0 0 2,340,548 1 7 / 1 2 Complete
4 recovered_host 206 96 0 110 0 0 1,527,724 1 7 / 1 2 Complete
5 recovered_host 179 111 0 17 5 51 1,536,704 1 7 / 1 2 Complete
6 recovered_host 178 105 0 72 0 0 1,541,368 1 7 / 1 2 Complete
7 recovered_host 180 171 0 0 2 9 2,029,939 1 7 / 1 2 Complete
Data Ranges Review Images

Test Notes

  1. no comparison: timestamps do not match
  2. missing: ['pressure']

Data Coverage

Deployment: 1234567
100%94%100%47%62%59%95%

Lat/Lon Differences (km)

Deployment: 1234567
1 0.00
2 0.280.00
3 0.440.220.00
4 0.020.290.460.00
5 0.250.170.210.270.00
6 0.180.460.610.170.400.00
7 0.270.100.170.290.080.440.00

Example Composite Data Plot

System Annotations

Metadata Start Date End Date Comment
CE07SHSM
12/1/17, 9:35 AM 3/25/18, 3:09 PM

Data loss due to a power system failure. The mooring stopped receiving power from photo voltaic panel 1 of 4 on Nov-2. By Nov-20, the mooring stopped receiving power from two more solar panels (PV2 & PV4) and one of two wind turbines (WT2). This indicates the Power System Controller (PSC) they connect to is failing. Mooring components will be shut down as the mooring loses the ability to recharge its batteries.

Id: 1237 By: crisien
Flag: not_operational Exclude: No

CE07SHSM-RID26-08-SPKIRB000
Method: telemetered
Stream: spkir_abj_dcl_instrument
11/30/18, 7:00 PM 2/21/19, 1:33 PM

Signal appears to decreasing in a manner that would suggest biofouling. Data from this time-period should be considered as suspect.

Id: 1567 By: wingardc
Flag: suspect Exclude: No

CE07SHSM-RID26-08-SPKIRB000
Method: telemetered
Stream: spkir_abj_dcl_instrument
2/21/19, 1:33 PM

Data suggests the sensor has become fouled. Readings essentially flat-lined beginning around December 31, 2018. Sensor was powered off February 21, 2019. No further data will be collected for the remainder of the deployment.

Id: 1566 By: wingardc
Flag: not_operational Exclude: No

CE07SHSM-RID26-08-SPKIRB000
10/10/19, 5:07 PM

The use of UV lamps has been implemented for biofouling mitigation on this instrument as of 2019-10-10T21:07:00Z.

Id: 2124 By: wingardc

CE07SHSM
Method: telemetered
5/13/20, 6:40 PM

Due to a bug in the mooring telemetry system, data during this time range was not telemetered to shore. It was recorded on the mooring and instruments and will be available in the recovered data streams.

Id: 2185 By: wingardc
Flag: not_available Exclude: No

Review Notes

Metadata Start Date End Date Comment
CE07SHSM-RID26-08-SPKIRB000

A substantial percentage (between 19-43%) of data are outside of global ranges for every deployment. Zooming in on the x-axis in the plots shows that all of the values that are outside of global ranges are negative.

By Lori Garzio, on 7/12/19

CE07SHSM-RID26-08-SPKIRB000

There is no pressure variable in the .nc files. This should be a coordinate.

By Lori Garzio, on 7/12/19

CE07SHSM-RID26-08-SPKIRB000
Deployment: 4
11/15/16, 12:00 AM 4/13/17, 11:22 PM

Suspect data - signal decreased starting in Nov 2016. This should be investigated and annotated.

By Lori Garzio, on 7/12/19

CE07SHSM-RID26-08-SPKIRB000
Deployment: 7
7/1/18, 12:00 AM 9/20/18, 8:04 PM

Suspect data - signal decreased to almost zero starting in July 2018. This should be investigated and annotated.

By Lori Garzio, on 7/12/19

New Note