Single Point Velocity Meter

Reference Designator
CE07SHSP-SP001-02-VELPTJ000
Review Status
Review Complete
Note
Velocity data from profilers were run through the automated QC tests, but were not plotted or reviewed visually.
Depth
0 to 87m
Class
VELPT (Single Point Velocity Meter)
Make / Model
Nortek / Aquadopp HR

Dataset Reviews Last processed: 7/1/19, 6:30 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_cspp 91 7 0 83 2 2 16,693 1 87 / 80 1 1 Complete
2 recovered_cspp 53 2 0 50 0 0 20,609 1 87 / 81 1 1 Complete
3 recovered_cspp 33 23 0 10 0 0 13,369 1 87 / 85 1 1 Complete
4 recovered_cspp 69 20 0 49 0 0 11,364 1 87 / 81 1 1 Complete
5 recovered_cspp 64 40 0 0 5 27 19,372 1 87 / 83 1 1 Complete
6 recovered_cspp 31 7 0 24 0 0 4,931 1 87 / 83 1 1 Complete
Data Ranges Review Images

Test Notes

  1. no other streams for comparison

Data Coverage

Deployment: 123456
8%4%70%29%62%23%

Lat/Lon Differences (km)

Deployment: 123456
1 0.00
2 0.290.00
3 0.500.210.00
4 0.420.130.090.00
5 0.380.150.190.120.00
6 0.470.330.390.380.470.00

System Annotations

Metadata Start Date End Date Comment
CE07SHSP-SP001-02-VELPTJ000
4/9/15, 7:25 PM

According to the manufacturer, data are suspect when the instrument is tilted more than 20 degrees (measured by pitch and roll).

Id: 810 By: lgarzio

Review Notes

Metadata Start Date End Date Comment
CE07SHSP-SP001

During deployments 2, and 3 there are several pressure (dbar) data points that are outliers (much deeper than the max deployment depth of 90m).

By Lori Garzio, on 7/2/19

CE07SHSP-SP001

Missing data for all deployments should be annotated

By Lori Garzio, on 7/2/19

New Note