Dep. | Preferred Method | Stream | DD | FD | SG | EG | Gaps | GD | TS | Rate (s) | Pressure Comp. | Time Order | Valid Data | Missing Data | Data Comp. | Missing Coords. | Review |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1 | recovered_wfp | flort_sample | 119 | 11 | 1 | 107 | 0 | 0 | 20,546 | 6 | 536 / | Complete | |||||
2 | recovered_wfp | flort_sample | 180 | 179 | 0 | 0 | 0 | 0 | 399,268 | 5 | 542 / | Complete | |||||
3 | recovered_wfp | flort_sample | 195 | 194 | 0 | 0 | 0 | 0 | 364,245 | 6 | 542 / | Complete | |||||
4 | recovered_wfp | flort_sample | 205 | 204 | 0 | 0 | 0 | 0 | 468,341 | 5 | 542 / | Complete | |||||
5 | recovered_wfp | flort_sample | 133 | 133 | 0 | 0 | 0 | 0 | 245,083 | 6 | 542 / | Complete | |||||
6 | telemetered | flort_sample | 203 | 23 | 0 | 180 | 0 | 0 | 49,114 | 5 | 540 / | 1 | 1 | Complete | |||
7 | recovered_wfp | flort_sample | 175 | 175 | 0 | 0 | 0 | 0 | 350,230 | 5 | 533 / | Complete | |||||
8 | recovered_wfp | flort_sample | 174 | 174 | 0 | 0 | 0 | 0 | 522,425 | 5 | 533 / | Complete |
Deployment: | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 |
---|---|---|---|---|---|---|---|---|
flort_sample | 9% | 99% | 99% | 100% | 100% | 11% | 100% | 100% |
Deployment: | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 |
---|---|---|---|---|---|---|---|---|
1 | 0.00 | |||||||
2 | 0.41 | 0.00 | ||||||
3 | 0.20 | 0.60 | 0.00 | |||||
4 | 0.38 | 0.10 | 0.56 | 0.00 | ||||
5 | 0.84 | 1.23 | 0.69 | 1.22 | 0.00 | |||
6 | 0.84 | 1.23 | 0.70 | 1.22 | 0.04 | 0.00 | ||
7 | 0.47 | 0.84 | 0.38 | 0.84 | 0.41 | 0.39 | 0.00 | |
8 | 0.11 | 0.46 | 0.15 | 0.41 | 0.84 | 0.85 | 0.50 | 0.00 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CE09OSPM-WFP01 Method: telemetered |
4/29/14, 10:00 AM | 8/14/14, 8:00 PM |
The bulkhead connector for the PAR sensor failed and shorted the profiler power system to seawater, draining the profiler battery pack in fairly short order. No data is available for this time period.
Id: 1663
By: wingardc
|
CE09OSPM-WFP01 Method: recovered-wfp |
4/29/14, 10:00 AM | 8/14/14, 8:00 PM |
The bulkhead connector for the PAR sensor failed and shorted the profiler power system to seawater, draining the profiler battery pack in fairly short order. No data is available for this time period.
Id: 1664
By: wingardc
|
CE09OSPM Method: telemetered |
12/6/15, 1:50 AM | 5/6/16, 10:00 PM |
The surface expression for the profiler mooring broke free during a storm event on December 06, 2015 and ended up drifting north to Vancouver Island where it was recovered. No further telemetered data will be available from this mooring or profiler system for this deployment.
Id: 1665
By: wingardc
|
CE09OSPM-WFP01 Method: recovered-wfp |
9/21/16, 10:18 PM | 4/12/17, 7:12 PM |
The surface expression for the profiler mooring broke free during a storm event in October 2016. On March 13, 2017, the subsurface sphere came to the surface and drifted north towards Vancouver Island where it was recovered from the shore. The mooring wire was severed near the base and showed signs of abrasion and other damage consistent with a ship strike. The profiler was no longer on the wire and is presumed lost at sea. No recovered profiler data is available for this deployment.
Id: 1661
By: wingardc
|
CE09OSPM Method: telemetered |
10/14/16, 3:01 PM | 4/12/17, 7:12 PM |
UPDATED, 2017-03-13: The subsurface float came to the surface at 06:16 UTC and drifted north towards Vancouver Island where it was recovered. The mooring wire was severed and showed signs of abrasion and other damage consistent with a ship strike. The profiler was no longer attached to the wire and is presumed lost at sea. * The buoy of the offshore Washington profiler went adrift at sometime around noon local time October 14, 2016. The subsurface beacons have not reported, so it is likely that the stretch hose between the buoy and the subsurface float parted, just as it did last year. If this is the case, the profiler is continuing to collect data, but it is no longer able to telemeter data to shore.
Id: 925
By: michaesm
|
CE09OSPM-WFP01 Method: telemetered |
7/10/17, 8:00 AM | 10/3/17, 4:13 PM |
At noon on July 10, 2017, the IMM port of the profiler mooring was turned off due to the incorrect configuration of the VEL3D-K. The instrument's sample rate was set too high, resulting in unusually large data files which caused a significant increase in the IMM telemetry time and an unsustainable power draw from the profiler battery pack. To insure the profiler continues to collect data over the course of the deployment, the IMM will remain off for the duration of the Spring/Summer 2017 deployment.
Id: 96
By: m.vardaro
|
CE09OSPM-WFP01-04-FLORTK000 Method: recovered-wfp Stream: flort_sample |
7/11/17, 10:40 PM | 10/3/17, 8:00 AM |
The ECO Triplet (FLORT-K) sensor flooded. No valid data is available for this time period.
Id: 1674
By: wingardc
|
CE09OSPM-WFP01-04-FLORTK000 Stream: flort_sample |
3/4/18, 7:00 PM | 3/26/18, 6:42 AM |
The ECO Triplet (FLORT-K) sensor appeared to biofoul beginning on or about 2018-03-05. The problem got progressively worse during the remainder of the deployment with the optical backscatter channel maxing out and the chlorophyll and CDOM flourescence baseline signals increasing in a nonlinear fashion.
Id: 1675
By: wingardc
|
CE09OSPM-WFP01-04-FLORTK000 |
3/26/18, 4:20 AM | 9/23/18, 4:31 AM |
A replacement ECO Triplet (FLORT-K) sensor (for unit that flooded during deployment 7) was not available in time for deployment 9. Profiler was deployed without an ECO Triplet.
Id: 1676
By: wingardc
|
CE09OSPM-WFP01 Method: telemetered |
4/7/19, 10:23 PM | 5/4/19, 5:22 AM |
Battery pack on the profiler is getting low. To conserve power for profiling, data telemetry has been disabled.
Id: 1666
By: wingardc
|
CE09OSPM-WFP01 Method: recovered-wfp |
4/10/19, 4:00 AM | 5/4/19, 5:22 AM |
Battery pack on the profiler was getting low. Sensors did not have enough power to fully operate (most noticeable in the VEL3D-K data streams), and the profiler did not have enough power to descend to the full operating depth. Stopped operating altogether on 2019-04-26 after completing 1728 profiles over the course of the deployment.
Id: 1667
By: wingardc
|
CE09OSPM-WFP01 Method: telemetered |
10/11/19, 12:00 AM | 12/31/19, 7:00 PM |
Data indicates that profiler travel up the wire is periodically blocked near 150 m, keeping it from fully ascending during a profile. Suspect fouling on the wire is impeding travel. Id: 1830 By: wingardc |
CE09OSPM-WFP01 Method: telemetered |
12/31/19, 7:00 PM |
Due to a clock rollover issue, all WFPs ceased sampling on 2020-01-01. No further WFP data are being collected from this system. A resolution is being researched by the vendor.
Id: 1905
By: wingardc
|
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CE09OSPM-WFP01 |
Deployment 6 for all instruments: Are the gaps in the data as a result of a change in the sampling rate? By Leila Belabbassi, on 3/26/19 |
||
CE09OSPM-WFP01 |
Deployment 9 for all instruments except FLORT (instrument was not deployed): By Leila Belabbassi, on 3/26/19 |
||
CE09OSPM-WFP01-04-FLORTK000 |
missing annotation to show that FLORT data are not expected because the instrument was not deployed during deployment 9 By Leila Belabbassi, on 3/27/19 |
||
CE09OSPM-WFP01-04-FLORTK000 |
The pressure coordinate (pressure_depth) contain non-valid pressure data. There is another variable in the files called in_ctd_pressure that contains what appears to be valid pressure data. This is the case for every FLORTK on every WFP. https://opendap.oceanobservatories.org/thredds/dodsC/ooi/lgarzio@marine.rutgers.edu/20181213T021222-CE09OSPM-WFP01-04-FLORTK000-recovered_wfp-flort_sample/deployment0008_CE09OSPM-WFP01-04-FLORTK000-recovered_wfp-flort_sample_20171004T200204-20180326T195110.nc.ascii?pressure_depth[0:1:1000],int_ctd_pressure[0:1:1000] By Lori Garzio, on 3/12/19 |
||
CE09OSPM-WFP01 |
Deployment 1 for all instruments: By Leila Belabbassi, on 3/28/19 |
||
CE09OSPM-WFP01-04-FLORTK000 Deployment: 5 |
|||
CE09OSPM-WFP01-04-FLORTK000 Deployment: 7 |
data are suspect: except for the seawater_scattering_coeff parameter that does not have a global-ranges in the system, 50 % of Chla, CDOM, and Backscatter data were outside of global ranges. By Leila Belabbassi, on 4/10/19 |
||
CE09OSPM-WFP01-04-FLORTK000 Deployment: 7 |
4/12/17, 7:16 PM | 10/3/17, 8:13 PM | |
CE09OSPM-WFP01-04-FLORTK000 Deployment: 8 |
3/1/18, 12:00 AM | 3/27/18, 12:00 AM |