Bio-acoustic Sonar (Coastal)

Reference Designator
CE09OSSM-MFD37-07-ZPLSCC000
Review Status
Will Not Review
Note
Not Delivered Via uFrame
Depth
540m
Class
ZPLSC (Bio-acoustic Sonar (Coastal))
Make / Model
ASL Environmental Sciences / AZFP

Dataset Reviews Last processed:

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_inst zplsc_echogram_data / Todo
2 recovered_inst zplsc_echogram_data / Todo
3 recovered_inst zplsc_echogram_data / Todo
4 recovered_inst zplsc_echogram_data / Todo
5 recovered_inst zplsc_echogram_data / Todo
6 recovered_inst zplsc_echogram_data / Todo
7 recovered_inst zplsc_echogram_data / Todo
Data Ranges Review Images

Test Notes

All good!

System Annotations

Metadata Start Date End Date Comment
CE09OSSM-MFD37
1/21/17, 7:00 PM 4/14/17, 8:00 PM

The high voltage lines providing power to the MFN via the CVT for CE09OSSM-00004 shorted out on 2017-01-22 at 08:31 UTC. Sea state was high, but not unusual for this time of year. Disabled future power scheduling to the MFN and applied an override to disable the CVT. No further data will be logged from the instruments on the MFN. This affects:
On DCL35:
* ADCPS (had already failed upon deployment)
* PCO2W -- should continue to autosample
* PHSEN -- should continue to autosample
* PRESF -- should continue to autosample
* VEL3D -- should continue to autosample
On DCL37
* CTDBP (plus DOSTA) -- should continue to autosample
* OPTAA
* ZPLSC -- had stopped reporting data early on in deployment, may still be autosampling

Id: 930 By: michaesm
Flag: not_available Exclude: No

CE09OSSM-MFD37
Method: telemetered
6/26/19, 8:40 PM

The high voltage power supply providing power to the seafloor dataloggers and instrumentation failed for reasons unknown at this time. No further telemetered data will be available for this deployment.

Id: 1696 By: wingardc
Flag: not_operational Exclude: No

CE09OSSM-MFD37
Method: telemetered
11/3/19, 3:45 AM

Power to the seafloor sensors failed. With the exception of the VEL3D and OPTAA, the sensors on the seafloor platform should continue to autonomously sample for the remainder of deployment. Until we recover and redeploy the mooring in the spring of 2020, no further telemetered data will be available from the seafloor sensors.

Id: 1763 By: wingardc
Flag: not_operational Exclude: No

CE09OSSM
Method: telemetered
5/9/20, 8:40 PM 5/10/20, 8:00 PM

Due to a bug in the mooring telemetry system, data during this time range was not telemetered to shore. It was recorded on the mooring and instruments and will be available in the recovered data streams.

Id: 2182 By: wingardc
Flag: not_available Exclude: No

Review Notes

Metadata Start Date End Date Comment
CE09OSSM-MFD37-07-ZPLSCC000
Deployment: 4
9/26/16, 12:00 AM 5/15/17, 12:00 AM

The ZPLSC scheduled sample time has shifted out alignment with the DCL sampling window. Last sample was on 2016/09/26 17:36. Sent new mission file changing the sample window on 2016-10-13, but that did not correct the issue. Will send new mission file with broader sample window.

Have attempted to expand sample window. Will see if that corrects issue

By Mike Smith, on 10/20/16
Redmine Issue #11404

New Note