pCO2 Air-Sea

Reference Designator
CE09OSSM-SBD12-04-PCO2AA000
Review Status
Review Complete
Note
Depth
0m
Class
PCO2A (pCO2 Air-Sea)
Make / Model
Pro-Oceanus / pCO2-pro

Dataset Reviews Last processed: 7/23/19, 7:28 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host 170 170 0 0 0 0 36,513 1 / 1 Complete
1 recovered_host 170 170 0 0 0 0 36,512 1 1 / Complete
2 recovered_host 206 206 0 0 0 0 44,208 1 / Complete
2 recovered_host 206 206 0 0 0 0 44,177 1 1 / Complete
3 recovered_host 134 134 0 0 0 0 28,701 1 / Complete
3 recovered_host 134 134 0 0 0 0 28,699 1 1 / Complete
4 recovered_host 206 206 0 0 0 0 44,406 1 / 2 Complete
4 recovered_host 206 206 0 0 0 0 44,292 1 1 / 2 Complete
5 recovered_host 178 178 0 0 0 0 38,436 1 / Complete
5 recovered_host 178 178 0 0 0 0 38,436 1 1 / Complete
6 recovered_host 178 177 0 0 0 0 38,178 1 / 3 Complete
6 recovered_host 178 177 0 0 0 0 38,134 1 1 / 3 Complete
7 telemetered 178 132 0 0 2 48 17,730 1 / 4 4 Complete
7 telemetered 178 132 0 0 2 48 17,729 1 1 / 4 4 Complete
Data Ranges Review Images

Test Notes

  1. fail: check ['CO2 Mole Fraction in Atmosphere', 'pCO2 Atmosphere']
  2. fail: data found in another stream (gaps: [1] days: [1])
  3. fail: data found in another stream (gaps: [11] days: [11])
  4. no other streams for comparison

Data Coverage

Deployment: 1234567
74%
100%100%100%100%100%99%
74%
100%100%100%100%100%99%

Lat/Lon Differences (km)

Deployment: 1234567
1 0.00
2 0.320.00
3 0.230.540.00
4 1.121.400.900.00
5 0.130.380.181.020.00
6 2.102.361.870.971.990.00
7 0.100.310.241.100.082.070.00

Example Composite Data Plot

System Annotations

Metadata Start Date End Date Comment
CE09OSSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_air
4/9/15, 1:30 PM 9/25/15, 2:30 PM

A vendor error in assembly pinched closed an internal tube carrying the surface air stream into the sample volume, resulting in an erroneous measurement. Users should not use the PCO2A air measurement for the Spring 2015 deployment (deployment 01), nor the associated sea-air pCO2 flux estimate.

Id: 1981 By: wingardc
Flag: fail Exclude: No

CE09OSSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_air_recovered
4/9/15, 1:30 PM 9/25/15, 2:30 PM

A vendor error in assembly pinched closed an internal tube carrying the surface air stream into the sample volume, resulting in an erroneous measurement. Users should not use the PCO2A air measurement for the Spring 2015 deployment (deployment 01), nor the associated sea-air pCO2 flux estimate.

Id: 1983 By: wingardc
Flag: fail Exclude: No

CE09OSSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_water_recovered
4/9/15, 1:30 PM 9/25/15, 2:30 PM

A vendor error in assembly pinched closed an internal tube carrying the surface air stream into the sample volume, resulting in an erroneous measurement. Users should not use the PCO2A air measurement for the Spring 2015 deployment (deployment 01), nor the associated sea-air pCO2 flux estimate.

Id: 1996 By: wingardc
Flag: fail Exclude: No

CE09OSSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_water
4/9/15, 1:30 PM 9/25/15, 2:30 PM

A vendor error in assembly pinched closed an internal tube carrying the surface air stream into the sample volume, resulting in an erroneous measurement. Users should not use the PCO2A air measurement for the Spring 2015 deployment (deployment 01), nor the associated sea-air pCO2 flux estimate.

Id: 1998 By: wingardc
Flag: fail Exclude: No

CE09OSSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_air
6/18/15, 5:30 PM 7/15/15, 4:30 PM

A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.

Id: 1982 By: wingardc
Flag: fail Exclude: No

CE09OSSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_air_recovered
6/18/15, 5:30 PM 7/15/15, 4:30 PM

A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.

Id: 1984 By: wingardc
Flag: fail Exclude: No

CE09OSSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_water
6/18/15, 5:30 PM 7/15/15, 4:30 PM

A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.

Id: 1986 By: wingardc
Flag: fail Exclude: No

CE09OSSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_water_recovered
6/18/15, 5:30 PM 7/15/15, 4:30 PM

A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.

Id: 1987 By: wingardc
Flag: fail Exclude: No

CE09OSSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_air
3/25/18, 9:30 PM 9/19/18, 10:30 AM

UPDATED, 2020-01-28: The QC flag for the air measurement has been reset to fail. Users should not use the PCO2A air measurement for the Spring 2018 deployment (deployment 0), nor the associated sea-air pCO2 flux estimate. *PCO2A air measurement for the Spring 2018 deployment is considered suspect. Recent investigations have indicated a vendor error in assembly may have pinched closed an internal tube carrying the surface air stream into the sample volume resulting in an erroneous measurement.

Id: 1448 By: wingardc
Flag: fail Exclude: No

CE09OSSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_air_recovered
3/25/18, 9:30 PM 9/19/18, 10:30 AM

A vendor error in assembly pinched closed an internal tube carrying the surface air stream into the sample volume, resulting in an erroneous measurement. Users should not use the PCO2A air measurement for the Spring 2018 deployment (deployment 07), nor the associated sea-air pCO2 flux estimate.

Id: 1985 By: wingardc
Flag: fail Exclude: No

CE09OSSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_water_recovered
3/25/18, 9:30 PM 9/19/18, 10:30 AM

A vendor error in assembly pinched closed an internal tube carrying the surface air stream into the sample volume, resulting in an erroneous measurement. Users should not use the PCO2A air measurement for the Spring 2018 deployment (deployment 07), nor the associated sea-air pCO2 flux estimate.

Id: 1997 By: wingardc
Flag: fail Exclude: No

CE09OSSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_water
3/25/18, 9:30 PM 9/19/18, 10:30 AM

A vendor error in assembly pinched closed an internal tube carrying the surface air stream into the sample volume, resulting in an erroneous measurement. Users should not use the PCO2A air measurement for the Spring 2018 deployment (deployment 07), nor the associated sea-air pCO2 flux estimate.

Id: 1999 By: wingardc
Flag: fail Exclude: No

CE09OSSM-SBD12-04-PCO2AA000
Method: telemetered
4/24/19, 6:25 PM 10/13/19, 10:31 AM

Due to a firmware bug, the sensor zeroing function for correcting the baseline drift over time is not currently saving the updated reference value used to compute the raw CO2 value outputs during the measurements. The sensors are zeroing at the end of the sample cycle when scheduled, but the firmware is not storing the new value. This results in the original zero reference value being used for all measurements and an observable and consistent negative drift in both the air and water CO2 values. The vendor has indicated that the data may be correctable post-recovery, however at this time no such correction algorithm has been provided.

Id: 1692 By: wingardc
Flag: fail Exclude: Yes

CE09OSSM-SBD12
Method: telemetered
5/25/19, 6:05 PM 5/27/19, 4:55 PM

The datalogger computer crashed and had to be power cycled to restore operation. Telemetered data was lost for this time period.

Id: 1693 By: wingardc
Flag: not_available Exclude: No

CE09OSSM-SBD12
Method: recovered-host
5/25/19, 6:05 PM 5/27/19, 4:55 PM

The datalogger computer crashed and had to be power cycled to restore operation. Telemetered data was lost for this time period.

Id: 1751 By: wingardc
Flag: not_available Exclude: No

CE09OSSM-SBD12-04-PCO2AA000
Method: telemetered
11/10/19, 7:30 PM 11/12/19, 7:30 PM

The instrument suffered a firmware crash and had to be power cycled in order to restore operations. We are in consultation with the vendor to determine the cause of this failure.

Id: 2055 By: wingardc
Flag: not_operational Exclude: No

CE09OSSM-SBD12-04-PCO2AA000
Method: telemetered
3/7/20, 12:30 PM 3/11/20, 8:30 AM

The instrument suffered a firmware crash and had to be power cycled in order to restore operations. We are in consultation with the vendor to determine the cause of this failure.

Id: 2060 By: wingardc
Flag: not_operational Exclude: No

CE09OSSM
Method: telemetered
5/9/20, 8:40 PM 5/10/20, 8:00 PM

Due to a bug in the mooring telemetry system, data during this time range was not telemetered to shore. It was recorded on the mooring and instruments and will be available in the recovered data streams.

Id: 2182 By: wingardc
Flag: not_available Exclude: No

Review Notes

Metadata Start Date End Date Comment
CE09OSSM-SBD12-04-PCO2AA000
Deployment: 1
4/9/15, 5:30 PM 9/25/15, 6:00 PM

Suspect data - values for CO2 in the atmosphere are too low and variable.

By Lori Garzio, on 7/26/19

CE09OSSM-SBD12-04-PCO2AA000
Deployment: 2
10/15/15, 11:06 PM 5/7/16, 3:28 PM

Suspect data - measured_water_co2 decreases gradually from Jan 1, 2016 to the end of the deployment, then drops off substantially in March 2016.

By Lori Garzio, on 7/26/19

CE09OSSM-SBD12-04-PCO2AA000
Deployment: 4
2/12/17, 12:00 AM 2/15/17, 12:00 AM

Suspect measured_water_co2 values

By Lori Garzio, on 7/26/19

CE09OSSM-SBD12-04-PCO2AA000
Deployment: 6
2/10/18, 12:00 AM 3/30/18, 7:15 PM

measured_water_co2 is suspect towards the end of the deployment (large swing from very high to low values).

By Lori Garzio, on 7/26/19

CE09OSSM-SBD12-04-PCO2AA000
Deployment: 7

There are no recovered_host data available for download for this deployment. According to the ingest csv, the recovered_host raw data files are available. These files should be ingested or the datasets should be annotated to explain why the recovered_host data aren't available.

By Lori Garzio, on 7/26/19

CE09OSSM-SBD12-04-PCO2AA000
Deployment: 7
3/26/18, 1:39 AM 9/19/18, 2:15 PM

Suspect data - values for CO2 in the atmosphere are too low and variable (annotation ID 1448).

By Lori Garzio, on 7/26/19

New Note