Dep. | Preferred Method | Stream | DD | FD | SG | EG | Gaps | GD | TS | Rate (s) | Pressure Comp. | Time Order | Valid Data | Missing Data | Data Comp. | Missing Coords. | Review |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1 | recovered_host | 149 | 89 | 0 | 59 | 0 | 0 | 382,922 | 7 / | 1 | 2 | 3 | Complete | ||||
2 | telemetered | 3 | 3 | 0 | 0 | 0 | 0 | 8,329 | 7 / | 4 | 4 | 3 | Complete | ||||
3 | recovered_host | 170 | 170 | 0 | 0 | 0 | 0 | 1,678,154 | 7 / | 1 | 2 | 3 | Complete | ||||
4 | recovered_host | 165 | 151 | 0 | 2 | 2 | 13 | 524,020 | 7 / | 1 | 4 | 4 | 3 | Complete | |||
5 | recovered_host | 154 | 139 | 0 | 0 | 5 | 19 | 550,857 | 7 / | 2 | 3 | Complete | |||||
6 | recovered_host | 240 | 240 | 0 | 0 | 0 | 0 | 2,380,521 | 7 / | 1 | 2 | 3 | Complete | ||||
7 | recovered_host | 146 | 132 | 0 | 1 | 1 | 14 | 1,289,720 | 7 / | 1 | 2 | 3 | Complete | ||||
8 | recovered_host | 152 | 152 | 0 | 0 | 0 | 0 | 1,503,023 | 7 / | 1 | 2 | 3 | Complete |
Deployment: | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 |
---|---|---|---|---|---|---|---|---|
100% | ||||||||
60% | 100% | 92% | 90% | 100% | 90% | 100% |
Deployment: | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 |
---|---|---|---|---|---|---|---|---|
1 | 0.00 | |||||||
2 | 0.84 | 0.00 | ||||||
3 | 0.40 | 0.95 | 0.00 | |||||
4 | 0.83 | 0.09 | 0.98 | 0.00 | ||||
5 | 0.42 | 0.97 | 0.02 | 1.00 | 0.00 | |||
6 | 0.83 | 0.07 | 0.97 | 0.02 | 0.99 | 0.00 | ||
7 | 0.35 | 0.93 | 0.05 | 0.96 | 0.07 | 0.95 | 0.00 | |
8 | 0.82 | 0.08 | 0.96 | 0.02 | 0.98 | 0.02 | 0.94 | 0.00 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP01CNSM-RID27-01-OPTAAD000 Stream: optaa_dj_dcl_instrument |
11/21/13, 1:15 PM |
Per vendor feedback, the first 1-2 min of each AC-S instrument sampling period should be considered suspect. Signal drift can occur as the instrument warms up, flushes with seawater, and the filter wheel spins up. Id: 1783 By: swhite |
|
CP01CNSM-RID27-01-OPTAAD000 Stream: optaa_dj_dcl_instrument_recovered |
11/21/13, 1:15 PM |
Per vendor feedback, the first 1-2 min of each AC-S instrument sampling period should be considered suspect. Signal drift can occur as the instrument warms up, flushes with seawater, and the filter wheel spins up. Id: 1784 By: swhite |
|
CP01CNSM |
2/16/14, 7:00 PM | 4/17/14, 8:00 PM |
Surface buoy exploded due to hydrogen build-up in well.
Id: 103
By: swhite
|
CP01CNSM-RID27-01-OPTAAD000 |
12/13/14, 1:47 PM | 12/15/14, 3:58 PM |
Data not recoverable. Id: 521 By: leila |
CP01CNSM |
12/13/14, 1:47 PM | 12/15/14, 3:58 PM |
Deployment 2: Short deployment due to the fuel cell not functioning correctly. No data expected. Id: 709 By: cdobson |
CP01CNSM-RID27-01-OPTAAD000 Method: telemetered |
10/23/15, 2:49 PM | 4/4/16, 8:03 AM |
Data not telemetered.
Id: 364
By: leila
|
CP01CNSM-RID27 |
4/23/18, 1:22 PM | 10/29/18, 9:48 PM |
Deployent 9: Power down - Instrument Disabled. No telemetered or recovered data expected for this time period.
Id: 1380
By: leila
|
CP01CNSM-RID27-01-OPTAAD000 Stream: optaa_dj_dcl_instrument |
9/27/19, 2:30 PM |
The OPTAA burst duration was changed from 2 minutes to 4 minutes as of the start date/time of this annotation (start of Deployment 12). Id: 1843 By: swhite |
|
CP01CNSM-RID27-01-OPTAAD000 Stream: optaa_dj_dcl_instrument_recovered |
9/27/19, 2:30 PM |
The OPTAA burst duration was changed from 2 minutes to 4 minutes as of the start date/time of this annotation (start of Deployment 12). Id: 1844 By: swhite |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP01CNSM-RID27-01-OPTAAD000 |
According to Roesler and Barnard (Optical proxy for phytoplankton biomass in the absence of photophysiology: Rethinking the absorption line height, Methods in Oceanography, 2013), “absorption meters are highly prone to biofouling, particularly biofilms which not only attenuate the collimated beam but also impact the scattering properties of the optical surfaces and tubes. In productive coastal waters biofouling can have significant impacts (i.e. 10% of the signal) within one to two weeks”. By Lori Garzio, on 6/28/19 |
||
CP01CNSM-RID27-01-OPTAAD000 |
There is no pressure variable in the .nc files. This should be a coordinate. By Lori Garzio, on 6/28/19 |
||
CP01CNSM-RID27-01-OPTAAD000 |
wavelength_a and wavelength_c are arrays of fill values for deployments 7 and 8. These variables should contain the measurement wavelengths. By Lori Garzio, on 6/28/19 |
||
CP01CNSM-RID27-01-OPTAAD000 |
beam_attenuation and optical_absorption are arrays of fill values for deployments 2, 3 and 6. By Lori Garzio, on 6/28/19 |
||
CP01CNSM Deployment: 2 |
Annotation ID 709 "Short deployment due to the fuel cell not functioning correctly" should be updated to indicate that this refers to deployment 2, and data aren't expected from most instruments for this time period. By Lori Garzio, on 3/22/19 |