Velocity Profiler (75 kHz)

Reference Designator
CP04OSPM-RII01-02-ADCPSL010
Review Status
Review Complete
Note
Depth
425m
Class
ADCPS (Velocity Profiler (long range))
Make / Model
Teledyne RDI / WorkHorse LongRanger Sentinel 75khz - inductive

Dataset Reviews Last processed: 8/15/19, 6:47 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
8 recovered_inst adcp_velocity_earth 152 152 0 0 0 0 3,618 3,600 427 / 416 1 Complete
Data Ranges Review Images

Test Notes

  1. no comparison: timestamps do not match

Data Coverage

Deployment: 8
adcp_velocity_earth 100%

Lat/Lon Differences (km)

Deployment: 8
8 0.00

System Annotations

Metadata Start Date End Date Comment
CP04OSPM
10/12/16, 2:00 AM 6/1/17, 7:22 PM

Failed

Id: 354 By: leila
Flag: not_operational Exclude: No

CP04OSPM-RII01-02-ADCPSL010
Method: telemetered
Stream: adcps_jln_stc_instrument
11/9/17, 5:08 PM

Telemetered inductive ADCP data is in pd12 format which does not include the depth-to-first-bin value which is needed for proper calculation of the data products. A nominal value is entered into the system so that telemetered data can be used when needed. For the best data the recovered data in pd0 format should be used.

Id: 1919 By: cdobson

Review Notes

No notes yet.

New Note