Dep. | Preferred Method | Stream | DD | FD | SG | EG | Gaps | GD | TS | Rate (s) | Pressure Comp. | Time Order | Valid Data | Missing Data | Data Comp. | Missing Coords. | Review |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1 | recovered_inst | 150 | 149 | 0 | 0 | 0 | 0 | 259,351 | 10 | 7 / 7 | 1 | Complete | |||||
2 | recovered_inst | 168 | 167 | 0 | 0 | 0 | 0 | 288,741 | 10 | 7 / 7 | 1 | Complete | |||||
3 | recovered_inst | 205 | 186 | 0 | 0 | 3 | 20 | 316,800 | 10 | 7 / 7 | 1 | Complete | |||||
4 | recovered_inst | 139 | 139 | 0 | 0 | 0 | 0 | 238,260 | 10 | 7 / 7 | 2 | Complete | |||||
5 | recovered_inst | 240 | 227 | 0 | 0 | 13 | 19 | 37,789 | 7 / 7 | 3 | 1 | Complete | |||||
6 | recovered_inst | 150 | 150 | 0 | 0 | 0 | 0 | 14,301 | 900 | 7 / 7 | 1 | Complete | |||||
7 | recovered_inst | 153 | 153 | 0 | 0 | 0 | 0 | 275,873 | 10 | 7 / 7 | 1 | Complete |
Deployment: | 1 | 2 | 3 | 4 | 5 | 6 | 7 |
---|---|---|---|---|---|---|---|
99% | 99% | 91% | 100% | 95% | 100% | 100% |
Deployment: | 1 | 2 | 3 | 4 | 5 | 6 | 7 |
---|---|---|---|---|---|---|---|
1 | 0.00 | ||||||
2 | 0.51 | 0.00 | |||||
3 | 0.56 | 0.79 | 0.00 | ||||
4 | 0.49 | 0.12 | 0.85 | 0.00 | |||
5 | 0.58 | 0.82 | 0.03 | 0.88 | 0.00 | ||
6 | 0.46 | 0.20 | 0.88 | 0.08 | 0.91 | 0.00 | |
7 | 0.54 | 0.75 | 0.05 | 0.81 | 0.08 | 0.84 | 0.00 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP04OSSM-RID27-03-CTDBPC000 |
12/12/14, 4:47 PM | 5/10/15, 3:57 PM |
Deployment 1: The CTDBP instrument deployed was a Series E which has 600m strain gauge rather than a Series C which has a 100m strain gauge pressure sensor. Data resolution is reduced. Id: 1908 By: cdobson |
CP04OSSM Method: telemetered |
10/22/15, 3:38 PM | 5/13/16, 1:18 PM |
Fuel cell impacted available instrument power, so telemetered data are not available or reliable. Recovered data are available for instruments that could record internally. Id: 1362 By: leila |
CP04OSSM-RID27-03-CTDBPC000 Method: telemetered Stream: ctdbp_cdef_dcl_instrument |
2/22/16, 5:30 AM | 3/29/16, 8:03 AM |
Deployment 3: The instrument deployment depth is at 7 m. Pressure sensor is at fault. The sensor values are above the sea surface.
Id: 764
By: leila
|
CP04OSSM Method: telemetered |
1/30/19, 10:00 AM | 4/8/19, 7:39 AM |
CP04OSSM D00009 buoy broke free and went adrift on 1/30/2019, 1500UTC. All instruments on the MFN ceased telemetry on 1/30/2019 and instruments on the buoy/NSIF continued to sample through 2/1/2019. No more telemetered data expected for this deployment. Id: 1557 By: cdobson |
CP04OSSM-RID27 |
11/1/19, 8:00 PM | 11/10/19, 7:00 PM |
Deployment 11: Telemetered data from instruments on DCL27 not expected for date range 11/2/2019-11/11/2019 due to CE being locked up. Id: 1807 By: cdobson |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP04OSSM-RID27-03-CTDBPC000 Deployment: 3 |
|||
CP04OSSM-RID27-03-CTDBPC000 Deployment: 5 |
Recovered_inst data don't match with the recovered_host and telemetered data. It looks like the timestamps are substantially offset. This should be investigated and resolved and/or annotated. By Lori Garzio, on 3/26/19 |
||
CP04OSSM-RID27-03-CTDBPC000 Deployment: 5 |
Deployment 5 data gaps (13 gaps spanning 19 days) should be annotated: [['2017-04-11T08:55:14', '2017-04-13T17:06:00'], ['2017-04-14T22:53:53', '2017-04-16T08:15:53'], ['2017-04-16T11:54:23', '2017-04-17T15:17:43'], ['2017-04-17T18:54:52', '2017-04-20T00:01:40'], ['2017-04-20T09:07:23', '2017-04-22T00:15:30'], ['2017-04-23T00:36:19', '2017-04-24T06:20:39'], ['2017-04-29T22:52:40', '2017-05-02T07:43:13'], ['2017-05-03T13:53:47', '2017-05-06T23:44:38'], ['2017-05-08T13:52:48', '2017-05-10T01:26:22'], ['2017-05-11T21:55:27', '2017-05-14T20:54:00'], ['2017-05-15T20:52:59', '2017-05-17T12:41:30'], ['2017-05-21T11:53:43', '2017-05-22T17:33:56'], ['2017-05-24T18:52:46', '2017-05-25T21:57:21']] By Lori Garzio, on 3/26/19 |
||
CP04OSSM-RID27-03-CTDBPC000 Deployment: 6 |
The sampling rate for deployment 6 is 900 seconds (for every other deployment the sampling rate is 10 seconds). This deviation from the sampling plan should be annotated. By Lori Garzio, on 3/26/19 |