3-Wavelength Fluorometer

Reference Designator
CP05MOAS-GL340-02-FLORTM000
Review Status
Review Complete
Note
Depth
0 to 1,000m
Class
FLORT (3-Wavelength Fluorometer)
Make / Model
WET Labs / ECO Puck FLBBCD-SLK

Dataset Reviews Last processed: 8/29/19, 3:24 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host flort_m_sample 68 67 0 0 0 0 3,566,569 1 200 / 193 1 Complete
2 recovered_host flort_m_sample 58 55 1 1 0 0 3,364,413 1 200 / 197 1 Complete
3 recovered_host flort_m_sample 1 1 0 0 0 0 3,961 1 200 / 155 Complete
4 telemetered flort_m_sample 1 1 0 0 0 0 47 60 200 / 101 2 2 Complete
5 recovered_host flort_m_sample / Complete
6 telemetered flort_m_sample 3 2 0 0 0 0 2,633 60 200 / 193 2 2 Complete
7 recovered_host flort_m_sample 73 73 0 0 0 0 5,038,923 1 200 / 194 3 Complete
Data Ranges Review Images

Test Notes

  1. fail: data found in another stream (gaps: [3] days: [3])
  2. no other streams for comparison
  3. fail: data found in another stream (gaps: [4] days: [4])

Data Coverage

Deployment: 1234567
flort_m_sample 99%95%100%100%67%100%

Deployment Map

System Annotations

Metadata Start Date End Date Comment
CP05MOAS-GL340
1/21/16, 12:08 PM 2/2/16, 8:32 PM

Deployment 3: Glider experienced an aft leak shortly after deployment and was drifted until it was recovered; no data are expected from this deployment.

Id: 2037 By: cdobson

CP05MOAS-GL340
8/9/16, 2:21 AM 8/9/16, 7:40 PM

Deployment 4: Glider's altimeter failed upon deployment so it was recovered. No data are expected from this deployment.

Id: 2038 By: cdobson

CP05MOAS-GL340
10/10/16, 9:12 PM 10/13/16, 7:11 AM

Deployment 5: Glider leaked after one day. No data are expected except from the ADCP.

Id: 672 By: cdobson

CP05MOAS-GL340
4/19/17, 8:50 PM 4/21/17, 8:00 PM

Deployment 6: This is a test deployment and was recovered the same day due to leak. No data are expected.

Id: 640 By: cdobson
Flag: not_available Exclude: No

CP05MOAS-GL340
Method: telemetered
5/1/19, 8:00 PM 5/4/19, 10:00 AM

Deployment 8: Glider stopped sampling on 5/3/2019 15:00 UTC due to a failed altimeter. Glider recovered 5/4/2019 at 14:00 UTC.

Id: 1583 By: cdobson

CP05MOAS-GL340
10/25/19, 8:00 AM 11/4/19, 7:00 PM

Deployment 9: Glider experienced GPS failure and forced to drift on 10/26/2019. No telemetered data expected after this date. Glider recovered 11/5/2019.

Id: 1744 By: cdobson

CP05MOAS-GL340
2/22/20, 7:00 AM 3/3/20, 7:00 AM

Deployment 10: Glider experienced a digifin leak on 2/22/2020 and was put into surface drift mode. No data are expected after this date.

Id: 2014 By: cdobson

Review Notes

Metadata Start Date End Date Comment
CP05MOAS

Bad altimeter pings are apparent in many deployments of the coastal gliders. A few example plots are attached - m_water_depth plotted as semi-transparent bathymetry.

By Lori Garzio, on 5/10/19

CP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/15/19

CP05MOAS-GL340-02-FLORTM000

Comparisons with the shipboard CTD casts were done for D0001, D0002, and R0007. Chl-a recorded by the glider was generally comparable to the CTD cast. Further analysis using the shipboard bottle data should be conducted. This is outside the scope of this project.

By Lori Garzio, on 3/13/19

CP05MOAS-GL340
Deployment: 1

Glider looks like it drifted at the end of the deployment - this should be annotated

By Lori Garzio, on 9/3/19

CP05MOAS-GL340-02-FLORTM000
Deployment: 1

Suspect data towards the end of the deployment.

By Lori Garzio, on 9/3/19

CP05MOAS-GL340
Deployment: 3

Short deployment should be annotated.

By Lori Garzio, on 9/3/19

CP05MOAS-GL340-02-FLORTM000
Deployment: 3

Values for all parameters are close to zero. This looks similar to CE05MOAS-GL386-02-FLORTM000 deployment 5 where the electrical tape was left on the sensor at deployment (Redmine 12049) - this should be investigated and annotated.

By Lori Garzio, on 9/3/19

CP05MOAS-GL340
Deployment: 4

Short deployment should be annotated.

By Lori Garzio, on 9/3/19

CP05MOAS-GL340-02-FLORTM000
Deployment: 4

There are no recovered_host data available for download for this deployment. According to the ingest csv, the raw recovered_host data files are available. These data should be ingested or annotated to indicate why they are missing.

By Lori Garzio, on 9/3/19

CP05MOAS-GL340-02-FLORTM000
Deployment: 7

Suspect data towards the end of the deployment.

By Lori Garzio, on 9/3/19

New Note