Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 telemetered flort_m_sample 34 33 0 0 1 1 26,349 60 1000 / 957 1 1 Complete
Data Ranges Review Images

Test Notes

  1. no other streams for comparison

Data Coverage

Deployment: 1
flort_m_sample 97%

Deployment Map

System Annotations

Metadata Start Date End Date Comment
CP05MOAS-GL375
1/11/15, 4:47 PM 2/12/15, 7:00 PM

Glider lost. No recovered data are available.

Id: 673 By: leila

Review Notes

Metadata Start Date End Date Comment
CP05MOAS

Bad altimeter pings are apparent in many deployments of the coastal gliders. A few example plots are attached - m_water_depth plotted as semi-transparent bathymetry.

By Lori Garzio, on 5/10/19

CP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/15/19

CP05MOAS-GL375-02-FLORTM000
Deployment: 1

Suspect data - chlorophyll-a values are very high for the middle of the winter (up to 35 ug/L). CP05MOAS-GL336 was sampling the same area at the same time (Jan-Feb 2015) and chl-a values recorded by GL336 are more reasonable (up to ~4 ug/L).

By Lori Garzio, on 9/4/19