3-Wavelength Fluorometer

Reference Designator
CP05MOAS-GL375-02-FLORTM000
Start Depth
0
End Depth
1,000
Location
Current Status
Review Complete
Class
FLORT (3-Wavelength Fluorometer)
Series
FLORT-M
Science Discipline
Biological
Make
WET Labs
Model
ECO Puck FLBBCD-SLK
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/CP05MOAS/GL375/02-FLORTM000/metadata
Method Data Stream Content Type
recovered_host flort_m_sample Data Products Report M2M Stats Science
telemetered flort_m_sample Data Products Report M2M Stats Science
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review SI20150110 01/11/2015 02/13/2015 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00375 CGINS-FLORTM-02817 40.0199 -70.3929 1000 1000
Metadata Start Date End Date Comment
CP05MOAS

Bad altimeter pings are apparent in many deployments of the coastal gliders. A few example plots are attached - m_water_depth plotted as semi-transparent bathymetry.

By Lori Garzio, on 5/10/19

CP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/15/19

CP05MOAS-GL375-02-FLORTM000
Deployment: 1

Suspect data - chlorophyll-a values are very high for the middle of the winter (up to 35 ug/L). CP05MOAS-GL336 was sampling the same area at the same time (Jan-Feb 2015) and chl-a values recorded by GL336 are more reasonable (up to ~4 ug/L).

By Lori Garzio, on 9/4/19

New Note

Metadata Start Date End Date Comment
CP05MOAS-GL375
1/11/15, 4:47 PM 2/12/15, 7:00 PM

Glider lost. No recovered data are available.

Id: 673 By: leila