QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host 97 90 0 6 0 0 661,043 1 1000 / 208 Complete
2 recovered_host 85 84 0 0 0 0 2,652,540 1 1000 / 215 Complete
3 recovered_host 98 89 0 9 0 0 1,416,157 1 1000 / 208 Complete
4 recovered_host 72 68 0 0 1 4 1,203,625 1 1000 / 212 Complete
5 recovered_host 24 21 0 2 0 0 329,473 1 1000 / 214 1 Complete
6 recovered_host 8 7 0 0 1 1 319,798 1 1000 / 208 Complete
Data Ranges Review Images

Test Notes

  1. fail: data found in another stream (gaps: [1] days: [1])

Data Coverage

Deployment: 123456
93%99%91%94%88%88%

Deployment Map

System Annotations

Metadata Start Date End Date Comment
CP05MOAS-GL380
Method: telemetered
10/25/18, 4:26 PM 10/26/18, 10:47 PM

Glider was deployed for two days before it expereienced a leak and needed to be recovered. No further data expected.

Id: 1570 By: cdobson

CP05MOAS-GL380
Method: recovered_host
10/25/18, 4:26 PM 10/26/18, 10:47 PM

Glider was deployed for two days before it expereienced a leak and needed to be recovered. No further data expected.

Id: 1571 By: cdobson

CP05MOAS-GL380-05-PARADM000
Method: recovered
10/25/18, 4:26 PM 10/26/18, 10:47 PM

Glider was deployed for two days before it expereienced a leak and needed to be recovered. No further data expected.

Id: 1572 By: cdobson

CP05MOAS-GL380-05-PARADM000
Method: recovered_host
10/25/18, 4:26 PM 10/26/18, 10:47 PM

Glider was deployed for two days before it expereienced a leak and needed to be recovered. No further data expected.

Id: 1573 By: cdobson

CP05MOAS-GL380
2/21/20, 7:00 AM 3/2/20, 2:50 AM

Deployment 8: Science sampling was disabled on this glider on 2/21/2020 to preserve battery power for recovery. No science data are expected after this date.

Id: 2011 By: cdobson

Review Notes

Metadata Start Date End Date Comment
CP05MOAS

Bad altimeter pings are apparent in many deployments of the coastal gliders. A few example plots are attached - m_water_depth plotted as semi-transparent bathymetry.

By Lori Garzio, on 5/10/19

CP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/15/19

CP05MOAS-GL380-05-PARADM000

Annotation IDs 1572 and 1573 are duplicates of 1570 and 1571

By Lori Garzio, on 9/4/19

CP05MOAS-GL380-05-PARADM000

99% of the sci_bsipar_par values for deployments 1 and 3 are outside of global ranges. sci_bsipar_par is the unscaled L0 PAR data product. The global ranges for this parameter are the same as for the scaled L1 PAR data product [0.0, 2500.0]. The global range for sci_bsipar_par should be reviewed.

By Lori Garzio, on 9/4/19

CP05MOAS-GL380-05-PARADM000
Deployment: 1

It looks like the sensor failed on 10/17/2014 (the CTD data record goes to 1/4/2015). If that is correct, this should be annotated.

By Lori Garzio, on 9/4/19

CP05MOAS-GL380
Deployment: 2

Glider looks like it had steering issues - this should be annotated.

By Lori Garzio, on 9/4/19

CP05MOAS-GL380-05-PARADM000
Deployment: 4

58% of the parad_m_par data for this deployment are outside of global ranges [0.0, 2500.0]. The majority of the values outside of global ranges are slightly negative

By Lori Garzio, on 9/4/19

CP05MOAS-GL380-05-PARADM000
Deployment: 5

Data are available in the telemetered data stream but are not in the recovered data stream from 2017-11-09T01:12:30 to 2017-11-09T08:07:26. These data should be in the recovered stream.

By Lori Garzio, on 9/4/19

CP05MOAS-GL380
Deployment: 6

Short deployment should be annotated

By Lori Garzio, on 9/4/19

New Note