QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host 17 16 0 0 0 0 414,233 2 1000 / 1017 Complete
2 recovered_host 53 52 0 0 0 0 1,233,757 2 1000 / 980 Complete
3 recovered_host 48 43 0 4 0 0 1,595,125 2 1000 / 667 Complete
4 recovered_host 116 102 0 13 0 0 2,587,142 2 1000 / 982 1 Complete
5 recovered_host 75 75 0 0 0 0 2,854,314 2 1000 / 993 Complete
6 recovered_host 68 68 0 0 0 0 2,646,487 2 1000 / 994 Complete
Data Ranges Review Images

Test Notes

  1. fail: check ['Seawater Pressure']

Data Coverage

Deployment: 123456
94%98%90%88%100%100%

Deployment Map

System Annotations

Metadata Start Date End Date Comment
CP05MOAS-GL387
7/13/16, 9:00 PM 7/28/16, 8:00 AM

The glider altimeter may have failed.

Id: 644 By: leila
Flag: not_available Exclude: No

CP05MOAS-GL387-03-CTDGVM000
Stream: ctdgv_m_glider_instrument
11/12/17, 2:00 PM 1/2/18, 9:00 PM

Conductivity sensor intermittent failure.

Id: 940 By: leila
Flag: fail Exclude: No

CP05MOAS-GL387
5/26/19, 8:00 PM

Deployment 7: Glider's altimeter failed on 5/27/2019 so it will be limited to sampling the bottom line of the FZ box and to 500m until the end of the deployment.

Id: 1620 By: cdobson

Review Notes

Metadata Start Date End Date Comment
CP05MOAS

Bad altimeter pings are apparent in many deployments of the coastal gliders. A few example plots are attached - m_water_depth plotted as semi-transparent bathymetry.

By Lori Garzio, on 5/10/19

CP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/15/19

CP05MOAS-GL387-03-CTDGVM000

Comparisons with the shipboard CTD casts were done for D0003 and D0006. Glider conductivity and salinity are unreasonable for the D0006 comparison. Further analysis using the shipboard bottle data should be conducted. This is outside the scope of this project.

By Lori Garzio, on 3/13/19

CP05MOAS-GL387
Deployment: 1

Glider doesn't look like it sampled according to plan. This should be annotated.

By Lori Garzio, on 9/5/19

CP05MOAS-GL387-03-CTDGVM000
Deployment: 6

30% of conductivity, salinity, and 40% of density values are nans or outside of global ranges. Annotation ID 940: conductivity sensor intermittently failed. It looks like it started to fail on/around 11/12/2017