CTD

Reference Designator
CP05MOAS-GL387-03-CTDGVM000
Start Depth
0
End Depth
1,000
Location
Current Status
Review Complete
Class
CTDGV (CTD Glider)
Series
CTDGV-M
Science Discipline
Physical
Make
Sea-Bird
Model
SBE Glider Payload CTD (GP-CTD)
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/CP05MOAS/GL387/03-CTDGVM000/metadata
Method Data Stream Content Type
recovered_host ctdgv_m_glider_instrument_recovered Data Products Report M2M Stats Science
telemetered ctdgv_m_glider_instrument Data Products Report M2M Stats Science
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review KN-217 04/16/2014 05/02/2014 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00387 CGINS-CTDGVM-09087 39.9617 -70.9058 1000 1000
2 Review SI20150110 01/11/2015 03/04/2015 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00387 CGINS-CTDGVM-09087 40.0098 -70.3986 1000 1000
3 Review AT-31 10/14/2015 11/30/2015 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00387 CGINS-CTDGVM-09087 40.07 -70.4152 1000 1000
4 Review AR1-07 04/04/2016 07/29/2016 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00387 CGINS-CTDGVM-09087 39.9685 -70.5384 1000 1000
5 Review SI20170418 04/19/2017 07/02/2017 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00387 CGINS-CTDGVM-09089 39.7836 -70.4657 1000 1000
6 Review AR-24 10/28/2017 01/03/2018 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00387 CGINS-CTDGVM-09089 40.0318 -70.6204 1000 1000
7 Review AR-34 04/08/2019 06/18/2019 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00387 CGINS-CTDGVM-09027 39.9681 -71 1000 1000
Metadata Start Date End Date Comment
CP05MOAS

Bad altimeter pings are apparent in many deployments of the coastal gliders. A few example plots are attached - m_water_depth plotted as semi-transparent bathymetry.

By Lori Garzio, on 5/10/19

CP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/15/19

CP05MOAS-GL387-03-CTDGVM000

Comparisons with the shipboard CTD casts were done for D0003 and D0006. Glider conductivity and salinity are unreasonable for the D0006 comparison. Further analysis using the shipboard bottle data should be conducted. This is outside the scope of this project.

By Lori Garzio, on 3/13/19

CP05MOAS-GL387
Deployment: 1

Glider doesn't look like it sampled according to plan. This should be annotated.

By Lori Garzio, on 9/5/19

CP05MOAS-GL387-03-CTDGVM000
Deployment: 6

30% of conductivity, salinity, and 40% of density values are nans or outside of global ranges. Annotation ID 940: conductivity sensor intermittently failed. It looks like it started to fail on/around 11/12/2017

By Lori Garzio, on 9/5/19

New Note

Metadata Start Date End Date Comment
CP05MOAS-GL387
7/13/16, 9:00 PM 7/28/16, 8:00 AM

The glider altimeter may have failed.

Id: 644 By: leila
Flag: not_available Exclude: No

CP05MOAS-GL387-03-CTDGVM000
Stream: ctdgv_m_glider_instrument
11/12/17, 2:00 PM 1/2/18, 9:00 PM

Conductivity sensor intermittent failure.

Id: 940 By: leila
Flag: fail Exclude: No

CP05MOAS-GL387
5/26/19, 8:00 PM

Deployment 7: Glider's altimeter failed on 5/27/2019 so it will be limited to sampling the bottom line of the FZ box and to 500m until the end of the deployment.

Id: 1620 By: cdobson