3-Wavelength Fluorometer

Reference Designator
CP05MOAS-GL388-02-FLORTM000
Review Status
Review Complete
Note
Depth
0 to 1,000m
Class
FLORT (3-Wavelength Fluorometer)
Make / Model
WET Labs / ECO Puck FLBBCD-SLK

Dataset Reviews Last processed: 8/30/19, 5:19 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host flort_m_sample 69 68 0 0 0 0 2,393,694 1 1000 / 984 1 Complete
2 recovered_host flort_m_sample / Complete
3 recovered_host flort_m_sample 92 84 0 7 0 0 2,847,549 1 1000 / 987 2 Complete
4 recovered_host flort_m_sample 51 51 0 0 0 0 1,878,909 1 1000 / 1000 3 Complete
5 recovered_host flort_m_sample 94 94 0 0 0 0 3,281,090 1 1000 / 994 4 Complete
6 recovered_host flort_m_sample 88 88 0 0 0 0 3,368,721 1 1000 / 1000 3 Complete
Data Ranges Review Images

Test Notes

  1. fail: data found in another stream (gaps: [3] days: [3])
  2. fail: data found in another stream (gaps: [2] days: [2])
  3. fail: data found in another stream (gaps: [1] days: [1])
  4. fail: data found in another stream (gaps: [4] days: [4])

Data Coverage

Deployment: 123456
flort_m_sample 99%91%100%100%100%

Deployment Map

System Annotations

Metadata Start Date End Date Comment
CP05MOAS-GL388
7/23/15, 4:19 PM 7/24/15, 3:00 AM

Failed.

Id: 674 By: leila

CP05MOAS-GL388
1/13/16, 1:00 AM 1/20/16, 7:00 PM

Battery power depleted. All science sensors were disabled to conserve power for communications.

Id: 645 By: leila
Flag: not_available Exclude: No

CP05MOAS-GL388
2/15/20, 7:00 AM 2/27/20, 7:00 AM

Deployment 9: Glider was trapped in a warm core ring and was blown off of the SS-1 line and out of the array.

Id: 2013 By: cdobson

CP05MOAS-GL388
2/27/20, 7:00 AM 3/2/20, 1:49 AM

Deployment 9: Science sampling was disabled on this glider on 2/27/2020 to preserve battery power for recovery. No science data are expected after this date.

Id: 2012 By: cdobson

Review Notes

Metadata Start Date End Date Comment
CP05MOAS

Bad altimeter pings are apparent in many deployments of the coastal gliders. A few example plots are attached - m_water_depth plotted as semi-transparent bathymetry.

By Lori Garzio, on 5/10/19

CP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/15/19

CP05MOAS-GL388-02-FLORTM000

Comparisons with the shipboard CTD casts were done for D0003 and D0006. Chl-a recorded by the glider was higher compared to the CTD cast. Further analysis using the shipboard bottle data should be conducted. This is outside the scope of this project.

By Lori Garzio, on 3/13/19

CP05MOAS-GL388-02-FLORTM000
Deployment: 1

Suspect backscatter data near the end of Nov 2014

By Lori Garzio, on 9/5/19

CP05MOAS-GL388
Deployment: 2

No data expected for deployment 2. Annotation ID 674 should be updated to provide details of the glider failure and that no data are available.

By Lori Garzio, on 9/5/19

CP05MOAS-GL388
Deployment: 3

Glider looks like it drifted - this should be annotated

By Lori Garzio, on 9/5/19

New Note