2-Wavelength Fluorometer

Reference Designator
GA05MOAS-GL538-01-FLORDM000
Review Status
Review Complete
Note
Depth
0 to 1,000m
Class
FLORD (2-Wavelength Fluorometer)
Make / Model
WET Labs / ECO Puck FLBB-SLC

Dataset Reviews Last processed: 4/10/19, 6:38 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host 346 293 0 33 1 21 4,917,564 1 1000 / 979 Complete
Data Ranges Review Images

Test Notes

All good!

Data Coverage

Deployment: 1
85%

Deployment Map

System Annotations

Metadata Start Date End Date Comment
GA05MOAS-GL538
9/9/16, 1:50 AM 9/9/16, 1:50 AM

Deployment 1: Glider went adrift and stopped collecting science data on 9/9/2016. It was recovered covered in barnacles and missing its right wing.

Id: 435 By: lgarzio
Flag: not_operational Exclude: No

Review Notes

Metadata Start Date End Date Comment
GA05MOAS-GL538-01-FLORDM000

Comparison with the shipboard CTD cast was done for D0001. Chl-a recorded by the glider was variable from 0-50 dbar, generally matched the CTD cast data from 50-90 dbar, and was higher than the CTD cast from 90-110 dbar. Further analysis using the shipboard bottle data should be conducted. This is outside the scope of this project.

By Lori Garzio, on 3/8/19

GA05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/31/19

GA05MOAS-GL538-01-FLORDM000
Deployment: 1

Biofouling noted in annotation ID 435 is evident in the data.

By Lori Garzio, on 6/3/19

New Note