Mobile Asset Controller

Reference Designator
GI05MOAS-GL484-00-ENG000000
Review Status
Engineering
Note
Depth
0 to 1,000m
Class
ENG00 (Mobile Asset Controller)
Make / Model
Teledyne Webb / G2 Slocum Gliders

Dataset Reviews

No reviews found.

Deployment Map

System Annotations

Metadata Start Date End Date Comment
GI05MOAS-GL484
8/17/15, 6:45 AM 8/26/15, 8:00 PM

Deployment 2: Deployed on 8/17/2015, recovered on 8/23/2015 at 12:26 UTC for ballast adjustment. Re-deployed at 12:40 UTC. Was recovered after 10 days due to rudder issues.

Id: 575 By: lgarzio

GI05MOAS-GL484
7/9/16, 6:32 AM 10/18/16, 6:06 PM

Glider went adrift and was not recovered. Only telemetered data are available.

Id: 164 By: lgarzio

Review Notes

Metadata Start Date End Date Comment
GI05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/31/19

GI05MOAS-GL484
Deployment: 2

Recovered data for deployment 2 should be ingested or annotated to explain why no recovered data are available for download. The ingest csv has a note that the glider was lost, but the glider was lost on deployment 3 not deployment 2.

By Lori Garzio, on 6/5/19

New Note