3-Wavelength Fluorometer

Reference Designator
GI05MOAS-PG528-03-FLORTM000
Review Status
Review Complete
Note
Depth
0 to 200m
Class
FLORT (3-Wavelength Fluorometer)
Make / Model
WET Labs / ECO Puck FLBBCD-SLK

Dataset Reviews Last processed: 1/9/20, 5:01 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host flort_m_sample 307 213 0 18 2 76 1,116,787 1 200 / 214 1 1 Complete
Data Ranges Review Images

Test Notes

  1. no other streams for comparison

Data Coverage

Deployment: 1
flort_m_sample 69%

Deployment Map

System Annotations

Metadata Start Date End Date Comment
GI05MOAS-PG528
8/17/15, 10:45 AM 5/29/16, 8:00 PM

Deployment 1: Glider was damaged in Jan. 2016, likely during a period of bad weather. It was able to dive and collect data, but not maintain position. Last science data received on 3/16/2016. Glider was recovered in the Davis Strait.

Id: 563 By: lgarzio

Review Notes

Metadata Start Date End Date Comment
GI05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/31/19

GI05MOAS-PG528-03-FLORTM000

Comparison with the shipboard CTD cast was done for D0001. Chl-a recorded by the glider was comparable to the CTD cast near the surface, and slightly higher compared to the CTD cast at depth (>100 dbar). Further analysis using the shipboard bottle data should be conducted. This is outside the scope of this project.

By Lori Garzio, on 3/8/19

New Note