Seawater pH (20 m)

Reference Designator
GS01SUMO-RII11-02-PHSENE041
Review Status
Review Complete
Note
Depth
20m
Class
PHSEN (Seawater pH)
Make / Model
Sunburst / SAMI-pH- inductive

Dataset Reviews Last processed: 1/12/19, 2:07 AM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_inst 313 136 0 177 0 0 3,221 3,600 20 / 1 Complete
2 recovered_inst 365 73 0 291 0 0 1,723 3,600 20 / 1 Complete
Data Ranges Review Images

Test Notes

  1. missing: ['pressure']

Data Coverage

Deployment: 12
43%20%

Lat/Lon Differences (km)

Deployment: 12
1 0.00
2 9.790.00

System Annotations

Metadata Start Date End Date Comment
GS01SUMO-RII11-02-PHSENE041
Method: recovered_inst
11/25/16, 6:00 AM 12/8/18, 11:42 PM

Deployment 3: The PHSEN041 was not recovered as it was not attached to the inductive line upon recovery. No recovered instrument data expected for this deployment.

Id: 2156 By: cdobson
Flag: not_operational Exclude: No

GS01SUMO-RII11-02-PHSENE041
3/9/17, 7:00 PM 12/8/18, 11:42 PM

Deployment 3: The PHSEN041 is known to ceased sampling 2017-03-10. No data are expected after this date. Root cause unknown.

Id: 2155 By: cdobson
Flag: not_operational Exclude: No

GS01SUMO
12/8/17, 7:00 PM 12/8/18, 11:42 PM

Deployment 3: *UPDATED 2020-04-28: This mooring was recovered 2018-12-09 which was a year later than originally intended. Some instruments were able to continue sampling for the extended deployment.* Recovery of the mooring was planned during the NBP 17-09 cruise (Nov - Dec 2017) but not completed due to adverse weather conditions. Recovery options are pending. Instruments that are still operational are likely reaching (or have surpassed) their deployment lifespan.

Id: 834 By: lgarzio

GS01SUMO-RII11-02-PHSENE041
12/4/18, 7:00 PM 1/20/20, 5:01 AM

UPDATED 2020-04-15: The instrument failed one day after deployment and was not recovered; no telemetered or recovered data.* Deployment 4: Telemetered data for this instrument ceased on 12/6/2018. Root cause still to be determined.

Id: 1596 By: cdobson

Review Notes

Metadata Start Date End Date Comment
GS01SUMO-RII11-02-PHSENE041

pressure coordinate is missing in the netCDF files

By Leila Belabbassi, on 2/19/19

GS01SUMO-RII11-02-PHSENE041

No annotation in the system for pH records that ended long before the deployment end dates:

Deployment 1: 177 out of 313 are missing
Deployment 2: 291 out of 365 are missing

By Leila Belabbassi, on 2/19/19

GS01SUMO-RII11-02-PHSENE041
7/1/15, 12:00 AM 12/15/16, 12:00 AM

suspect data

By Leila Belabbassi, on 2/19/19

GS01SUMO
Deployment: 3

Annotation ID 834 needs to be updated: "Deployment 3: Recovery of the mooring was planned during the NBP 17-09 cruise (Nov - Dec 2017) but not completed due to adverse weather conditions. Recovery options are pending. Instruments that are still operational are likely reaching (or have surpassed) their deployment lifespan."

By Lori Garzio, on 1/14/19

GS01SUMO-RII11-02-PHSENE041
Deployment: 3
3/10/17, 12:00 AM

Instrument is not sending valid data records via telemetry.

By Lori Garzio, on 3/13/17
Redmine Issue #11857

New Note