Dep. | Preferred Method | Stream | DD | FD | SG | EG | Gaps | GD | TS | Rate (s) | Pressure Comp. | Time Order | Valid Data | Missing Data | Data Comp. | Missing Coords. | Review |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1 | recovered_inst | / | Complete | ||||||||||||||
2 | recovered_inst | 365 | 365 | 0 | 0 | 0 | 0 | 104,754,961 | -5 / | 1 | 2 | 2 | Complete |
Deployment: | 1 | 2 |
---|---|---|
100% |
Deployment: | 1 | 2 |
---|---|---|
2 | 0.00 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
GS01SUMO-SBD12-08-FDCHPA000 |
2/18/15, 4:06 PM | 12/11/16, 7:00 PM |
Deployments 1 and 2: Sonic temperature bugs: general slow response problem leading to errors of +/- 6 degrees from ambient, and an additional nonlinear error in the slow response when the ambient temperature is between -5 and 5 degrees. The fast response (> 1 Hz) of these sensors is still accurate. Therefore, flux measurements using the detrended temperature data should still be valid. w bug: positive vertical velocities w output by the anemometer need to be multiplied by 1.166; negative vertical velocities need to be multiplied by 1.289 (Redmine 11645)
Id: 75
By: lgarzio
|
GS01SUMO-SBD12-08-FDCHPA000 Method: telemetered Stream: fdchp_a_dcl_instrument |
2/18/15, 4:06 PM | 1/20/20, 5:01 AM |
UPDATED, 2019-11-07: Per communication with the instrument subject matter expert, the onboard processing error applies only to the motion-corrected wind speed and flux calculations. Those parameters are marked here with a QC Flag = "fail". All other measurements are considered to be good. Users should only use the motion-corrected wind speed and flux data from the recovered instrument data stream. * The telemetered FDCHP data contains an onboard processing error that makes the motion-corrected wind speed and flux data unusable. The instrument subject matter expert is being consulted, and the algorithms are under review. Only the processed, recovered data should be used. For further questions please email help@oceanobservatories.org.
Id: 1376
By: swhite
|
GS01SUMO-SBD12-08-FDCHPA000 |
2/18/15, 4:06 PM | 12/27/15, 6:20 AM |
Deployment 1: Recovered_instrument data not available for this deployment. Id: 631 By: lgarzio |
GS01SUMO-SBD12 |
12/13/15, 7:00 PM | 12/11/16, 7:00 PM |
Upon recovery of deployment 2, damage to some components of the surface buoy were noted (e.g., FDCHP was bent, the tops of the METBK-PRC and FleetBroadBand antenna were missing). Id: 95 By: lgarzio |
GS01SUMO-SBD12-08-FDCHPA000 Method: recovered_inst Stream: fdchp_a_instrument_recovered |
12/14/15, 3:20 PM |
There are errors in the calculations of the recovered_instrument L1 and L2 data products (motion-corrected wind speeds, sonic temperature, and flux data products). This issue is under investigation (Redmine 13402). Id: 1395 By: lgarzio |
|
GS01SUMO-SBD12-08-FDCHPA000 |
12/14/15, 3:20 PM | 12/12/16, 2:58 AM |
Deployment 2: Instrument bent at recovery. Id: 755 By: lgarzio |
GS01SUMO-SBD12-08-FDCHPA000 |
6/11/17, 8:00 PM | 12/8/18, 11:42 PM |
Deployment 3: The FDCHP is known to have ceased sampling on 2017-06-12. No data are expected after this date.
Id: 2147
By: cdobson
|
GS01SUMO |
12/8/17, 7:00 PM | 12/8/18, 11:42 PM |
Deployment 3: *UPDATED 2020-04-28: This mooring was recovered 2018-12-09 which was a year later than originally intended. Some instruments were able to continue sampling for the extended deployment.* Recovery of the mooring was planned during the NBP 17-09 cruise (Nov - Dec 2017) but not completed due to adverse weather conditions. Recovery options are pending. Instruments that are still operational are likely reaching (or have surpassed) their deployment lifespan. Id: 834 By: lgarzio |
GS01SUMO-SBD12-08-FDCHPA000 |
12/13/18, 7:00 PM | 1/20/20, 5:01 AM |
UPDATED 2020-06-23: On recovery anemometer was found to have a bent shaft and missing one of the transducer tines. * Deployment 4: This instrument failed on deployment 12/14/2018. No telemetered or recovered data expected. Id: 1591 By: cdobson |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
GS01SUMO-SBD12-08-FDCHPA000 |
Re annotation ID 1376 for this instrument - the equivalent annotation on CE02SHSM-SBD12-08-FDCHPA000 (ID 1373) was updated by wingardc: "UPDATED, 2019-02-05: Per communication with the instrument subject matter expert, the onboard processing error applies only to the motion-corrected wind speed and flux calculations. Those parameters are marked here with a QC Flag = "fail". All other measurements are considered to be good. Users should only use the motion-corrected wind speed and flux data from the recovered instrument data stream." This information presumably also applies to this instrument and should be communicated to users. By Lori Garzio, on 6/7/19 |
||
GS01SUMO-SBD12-08-FDCHPA000 |
There are no global ranges in the system for this instrument for fdchp_a_fluxhot, fdchp_a_fluxmom_alongwind, and fdchp_a_fluxmom_crosswind By Lori Garzio, on 6/7/19 |
||
GS01SUMO-SBD12-08-FDCHPA000 |
Data arrays are all fill values for every deployment for fdchp_a_tmpatur, fdchp_a_windtur_north, fdchp_a_windtur_up, fdchp_a_windtur_west, fdchp_a_fluxhot, fdchp_a_fluxmom_alongwind, fdchp_a_fluxmom_crosswind (Redmine 13402). There are several errors listed in the computed_provenance section of the provenance.json files (one example file attached). When this issue is resolved annotation ID 1395 should be updated. By Lori Garzio, on 6/7/19 |
||
GS01SUMO Deployment: 3 |
Annotation ID 834 needs to be updated: "Deployment 3: Recovery of the mooring was planned during the NBP 17-09 cruise (Nov - Dec 2017) but not completed due to adverse weather conditions. Recovery options are pending. Instruments that are still operational are likely reaching (or have surpassed) their deployment lifespan." By Lori Garzio, on 1/14/19 |