3-Wavelength Fluorometer

Reference Designator
GS05MOAS-PG565-04-FLORTO000
Review Status
Review Complete
Note
Depth
0 to 200m
Class
FLORT (3-Wavelength Fluorometer)
Make / Model
WET Labs / ECO BB3

Dataset Reviews Last processed: 1/28/20, 7:21 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host flort_o_glider_data 338 315 0 15 4 7 1,570,616 1 200 / 211 1 1 Complete
Data Ranges Review Images

Test Notes

  1. no other streams for comparison

Data Coverage

Deployment: 1
flort_o_glider_data 93%

Deployment Map

System Annotations

Metadata Start Date End Date Comment
GS05MOAS-PG565
12/20/15, 12:56 PM 11/5/16, 9:02 PM

Deployment 1: Deployed on 12/20/2015, recovered on 12/25/2015 20:40 UTC for ballast correction and was re-deployed. Starting around 5/1/2016, glider became increasingly difficult to steer. Glider stopped profiling on 11/6/2016 to conserve remaining battery power. At recovery, barnacles were present mostly on the underside of the glider, and were not in an area to hinder flight performance.

Id: 577 By: lgarzio

Review Notes

Metadata Start Date End Date Comment
GS05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/31/19

GS05MOAS-PG565-04-FLORTO000

The arrays for the following variables are all NaNs: sci_bb3slo_b470_units, flort_o_bback_b470_total, flort_o_bback_b532_total, sci_bb3slo_b660_units, flort_o_bback_b660_total. According to the provenance.json (attached), the following cal values are missing: CC_1_dark_counts_volume_scatter, CC_1_scale_factor_volume_scatter, CC_1_measurement_wavelength, CC_2_dark_counts_volume_scatter, CC_2_scale_factor_volume_scatter, CC_2_measurement_wavelength, CC_3_dark_counts_volume_scatter, CC_3_scale_factor_volume_scatter, and CC_3_measurement_wavelength.

By Lori Garzio, on 1/29/20

GS05MOAS-PG565-04-FLORTO000

There are no global ranges in the system for sci_bb3slo_b470_units, flort_o_bback_b470_total, flort_o_bback_b532_total, sci_bb3slo_b660_units, and flort_o_bback_b660_total.

By Lori Garzio, on 1/29/20

GS05MOAS-PG565-04-FLORTO000

Biofouling could be an issue in the second half of the deployment.

By Lori Garzio, on 1/29/20