Method | Data Stream | Content | Type | |
---|---|---|---|---|
recovered_host | flort_o_glider_data | Data Products | Report M2M Stats | Science |
telemetered | flort_o_glider_data | Data Products | Report M2M Stats | Science |
Deployment | Cruise | Start Date | Stop Date | Mooring Asset | Node Asset | Sensor Asset | Latitude | Longitude | Deployment Depth | Water Depth |
---|---|---|---|---|---|---|---|---|---|---|
1 Review | NBP-15-11 | 12/20/2015 | 11/21/2016 | CGVGS-05MOAS-00000 | CGVEH-GLDRGP-00565 | CGINS-FLORTO-01332 | -54.1252 | -89.7625 | 200 | 200 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
GS05MOAS |
Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue. By Lori Garzio, on 5/31/19 |
||
GS05MOAS-PG565-04-FLORTO000 |
The arrays for the following variables are all NaNs: sci_bb3slo_b470_units, flort_o_bback_b470_total, flort_o_bback_b532_total, sci_bb3slo_b660_units, flort_o_bback_b660_total. According to the provenance.json (attached), the following cal values are missing: CC_1_dark_counts_volume_scatter, CC_1_scale_factor_volume_scatter, CC_1_measurement_wavelength, CC_2_dark_counts_volume_scatter, CC_2_scale_factor_volume_scatter, CC_2_measurement_wavelength, CC_3_dark_counts_volume_scatter, CC_3_scale_factor_volume_scatter, and CC_3_measurement_wavelength. By Lori Garzio, on 1/29/20 |
||
GS05MOAS-PG565-04-FLORTO000 |
There are no global ranges in the system for sci_bb3slo_b470_units, flort_o_bback_b470_total, flort_o_bback_b532_total, sci_bb3slo_b660_units, and flort_o_bback_b660_total. By Lori Garzio, on 1/29/20 |
||
GS05MOAS-PG565-04-FLORTO000 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
GS05MOAS-PG565 |
12/20/15, 12:56 PM | 11/5/16, 9:02 PM |
Deployment 1: Deployed on 12/20/2015, recovered on 12/25/2015 20:40 UTC for ballast correction and was re-deployed. Starting around 5/1/2016, glider became increasingly difficult to steer. Glider stopped profiling on 11/6/2016 to conserve remaining battery power. At recovery, barnacles were present mostly on the underside of the glider, and were not in an area to hinder flight performance. Id: 577 By: lgarzio |