Photosynthetically Available Radiation

Reference Designator
CE09OSPM-WFP01-05-PARADK000
Start Depth
15
End Depth
540
Location
Current Status
Review Complete
Class
PARAD (Photosynthetically Available Radiation)
Series
PARAD-K
Science Discipline
Biological
Make
Biospherical Instruments
Model
QSP-2200
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/CE09OSPM/WFP01/05-PARADK000/metadata
Method Data Stream Content Type
recovered_wfp parad_k__stc_imodem_instrument_recovered Data Products Report M2M Stats Science
telemetered parad_k__stc_imodem_instrument Data Products Report M2M Stats Science
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review OC1404B 04/19/2014 08/15/2014 CGMCE-09OSPM-00001 CGVEH-CWFPCE-99103 CGINS-PARADK-20438 46.8517 -124.971 536 536
2 Review OC1410A 10/08/2014 04/06/2015 CGMCE-09OSPM-00002 CGVEH-CWFPCE-99104 CGINS-PARADK-20448 46.853 -124.966 542 542
3 Review OC1503D 04/06/2015 10/16/2015 CGMCE-09OSPM-00003 CGVEH-CWFPCE-99103 CGINS-PARADK-20438 46.8517 -124.967 542 542
4 Review OC1510A 10/16/2015 05/07/2016 CGMCE-09OSPM-00004 CGVEH-CWFPCE-99104 CGINS-PARADK-20448 46.8521 -124.966 542 542
5 Review TN-342 05/12/2016 09/21/2016 CGMCE-09OSPM-00005 CGVEH-CWFPCE-99103 CGINS-PARADK-20438 46.8513 -124.982 542 542
6 Review AT37-03 09/22/2016 04/12/2017 CGMCE-09OSPM-00006 CGVEH-CWFPCE-99104 CGINS-PARADK-20448 46.8517 -124.982 540 540
7 Review SKQ201704S 04/12/2017 10/03/2017 CGMCE-09OSPM-00007 CGVEH-CWFPCE-99103 CGINS-PARADK-20438 46.8526 -124.977 533 533
8 Review SKQ201715S 10/04/2017 03/26/2018 CGMCE-09OSPM-00008 CGVEH-CWFPCE-37801 CGINS-PARADK-20555 46.8507 -124.971 533 533
9 Review SKQ201808S 03/26/2018 09/23/2018 CGMCE-09OSPM-00009 CGVEH-CWFPCE-99103 CGINS-PARADK-20438 46.8537 -124.967 542 542
10 Review SR1813 09/23/2018 05/04/2019 CGMCE-09OSPM-00010 CGVEH-CWFPCE-37801 CGINS-PARADK-20555 46.8507 -124.97 533 533
11 Review OC1905C 05/04/2019 10/13/2019 CGMCE-09OSPM-00011 CGVEH-CWFPCE-99103 CGINS-PARADK-20438 46.8528 -124.966 533 533
12 Review SKQ201921S 10/11/2019 CGMCE-09OSPM-00012 CGVEH-CWFPCE-37801 CGINS-PARADK-20555 46.8525 -124.974 533 533
Metadata Start Date End Date Comment
CE09OSPM-WFP01

Deployment 6 for all instruments:
Update annotation ID # 925 to indicate that the profiler was lost at sea and no recovered data are expected.
Two other issues with telemetered data need to be mentioned:
(1) the gaps in the data.
(2) the change in the sampling rate.

Are the gaps in the data as a result of a change in the sampling rate?

By Leila Belabbassi, on 3/26/19

CE09OSPM-WFP01

Deployment 9 for all instruments except FLORT (instrument was not deployed):
The data missing at the end of the deployment in recovered_host are found in the telemetered stream. Are recovered_host data incomplete?

By Leila Belabbassi, on 3/26/19

CE09OSPM-WFP01

Deployment 1 for all instruments:
The data record ended long before the deployment end date. There is no annotation to explain what happened.

By Leila Belabbassi, on 3/28/19

CE09OSPM-WFP01-05-PARADK000

For all deployments, the increase in PAR at the bottom is unrealistic and also in general, data below 100 dbar are suspect (example plot attached)

By Leila Belabbassi, on 3/28/19

CE09OSPM-WFP01-05-PARADK000
Deployment: 1
4/18/14, 12:00 AM 5/16/14, 12:00 AM

suspect data profiles

By Leila Belabbassi, on 3/28/19

New Note

Metadata Start Date End Date Comment
CE09OSPM-WFP01
Method: telemetered
4/29/14, 10:00 AM 8/14/14, 8:00 PM

The bulkhead connector for the PAR sensor failed and shorted the profiler power system to seawater, draining the profiler battery pack in fairly short order. No data is available for this time period.

Id: 1663 By: wingardc
Flag: not_available Exclude: No

CE09OSPM-WFP01
Method: recovered-wfp
4/29/14, 10:00 AM 8/14/14, 8:00 PM

The bulkhead connector for the PAR sensor failed and shorted the profiler power system to seawater, draining the profiler battery pack in fairly short order. No data is available for this time period.

Id: 1664 By: wingardc
Flag: not_available Exclude: No

CE09OSPM
Method: telemetered
12/6/15, 1:50 AM 5/6/16, 10:00 PM

The surface expression for the profiler mooring broke free during a storm event on December 06, 2015 and ended up drifting north to Vancouver Island where it was recovered. No further telemetered data will be available from this mooring or profiler system for this deployment.

Id: 1665 By: wingardc
Flag: not_available Exclude: No

CE09OSPM-WFP01
Method: recovered-wfp
9/21/16, 10:18 PM 4/12/17, 7:12 PM

The surface expression for the profiler mooring broke free during a storm event in October 2016. On March 13, 2017, the subsurface sphere came to the surface and drifted north towards Vancouver Island where it was recovered from the shore. The mooring wire was severed near the base and showed signs of abrasion and other damage consistent with a ship strike. The profiler was no longer on the wire and is presumed lost at sea. No recovered profiler data is available for this deployment.

Id: 1661 By: wingardc
Flag: not_available Exclude: No

CE09OSPM
Method: telemetered
10/14/16, 3:01 PM 4/12/17, 7:12 PM

UPDATED, 2017-03-13: The subsurface float came to the surface at 06:16 UTC and drifted north towards Vancouver Island where it was recovered. The mooring wire was severed and showed signs of abrasion and other damage consistent with a ship strike. The profiler was no longer attached to the wire and is presumed lost at sea. * The buoy of the offshore Washington profiler went adrift at sometime around noon local time October 14, 2016. The subsurface beacons have not reported, so it is likely that the stretch hose between the buoy and the subsurface float parted, just as it did last year. If this is the case, the profiler is continuing to collect data, but it is no longer able to telemeter data to shore.

Id: 925 By: michaesm
Flag: not_available Exclude: No

CE09OSPM-WFP01
Method: telemetered
7/10/17, 8:00 AM 10/3/17, 4:13 PM

At noon on July 10, 2017, the IMM port of the profiler mooring was turned off due to the incorrect configuration of the VEL3D-K. The instrument's sample rate was set too high, resulting in unusually large data files which caused a significant increase in the IMM telemetry time and an unsustainable power draw from the profiler battery pack. To insure the profiler continues to collect data over the course of the deployment, the IMM will remain off for the duration of the Spring/Summer 2017 deployment.

Id: 96 By: m.vardaro
Flag: not_available Exclude: No

CE09OSPM-WFP01-05-PARADK000
9/23/18, 3:48 PM 5/4/19, 5:22 PM

It is possible that at some point during the deployment, the diffuser on the PAR sensor was knocked off. We know the diffuser was missing when the profiler was recovered, but there is not

Id: 1677 By: wingardc
Flag: suspect Exclude: No

CE09OSPM-WFP01
Method: telemetered
4/7/19, 10:23 PM 5/4/19, 5:22 AM

Battery pack on the profiler is getting low. To conserve power for profiling, data telemetry has been disabled.

Id: 1666 By: wingardc
Flag: not_available Exclude: No

CE09OSPM-WFP01
Method: recovered-wfp
4/10/19, 4:00 AM 5/4/19, 5:22 AM

Battery pack on the profiler was getting low. Sensors did not have enough power to fully operate (most noticeable in the VEL3D-K data streams), and the profiler did not have enough power to descend to the full operating depth. Stopped operating altogether on 2019-04-26 after completing 1728 profiles over the course of the deployment.

Id: 1667 By: wingardc
Flag: not_available Exclude: No

CE09OSPM-WFP01
Method: telemetered
10/11/19, 12:00 AM 12/31/19, 7:00 PM

Data indicates that profiler travel up the wire is periodically blocked near 150 m, keeping it from fully ascending during a profile. Suspect fouling on the wire is impeding travel.

Id: 1830 By: wingardc

CE09OSPM-WFP01
Method: telemetered
12/31/19, 7:00 PM

Due to a clock rollover issue, all WFPs ceased sampling on 2020-01-01. No further WFP data are being collected from this system. A resolution is being researched by the vendor.

Id: 1905 By: wingardc
Flag: not_operational Exclude: No