Velocity Profiler (150kHz)

Reference Designator
CP02PMCI-RII01-02-ADCPTG010
Start Depth
104
End Depth
104
Location
Sensor cage
Current Status
Review Complete
Class
ADCPT (Velocity Profiler (short range))
Series
ADCPT-G
Science Discipline
Physical
Make
Teledyne RDI
Model
WorkHorse Sentinel 150khz - inductive
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/CP02PMCI/RII01/02-ADCPTG010/metadata
Method Data Stream Content Type
recovered_host adcps_jln_stc_instrument_recovered Data Products Report M2M Stats Science
recovered_host adcps_jln_stc_metadata_recovered Metadata Report M2M Engineering
recovered_inst adcp_config Configuration Data Report M2M Engineering
recovered_inst adcp_engineering Engineering Data Report M2M Engineering
recovered_inst adcp_velocity_earth Earth Coordinates Data Products Report M2M Stats Science
telemetered adcps_jln_stc_instrument Data Products Report M2M Stats Science
telemetered adcps_jln_stc_metadata Metadata Report M2M Engineering
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review KN217 04/16/2014 10/10/2014 CGMCP-02PMCI-00001 CGINS-ADCPTG-18596 40.2267 -70.8886 105 127
2 Review KN222 10/10/2014 12/15/2014 CGMCP-02PMCI-00002 CGINS-ADCPTG-20495 40.2265 -70.878 105 127
3 Review AT27 05/01/2015 10/15/2015 CGMCP-02PMCI-00003 CGINS-ADCPTG-19151 40.2267 -70.8888 106 128
4 Review AT31 10/15/2015 05/20/2016 CGMCP-02PMCI-00004 CGINS-ADCPTG-20495 40.2267 -70.8779 105 127
5 Review AR4 05/21/2016 05/30/2016 CGMCP-02PMCI-00005 CGINS-ADCPTG-19151 40.2266 -70.889 106 128
6 Review AR7-01 07/01/2016 10/02/2016 CGMCP-02PMCI-00006 CGINS-ADCPTG-19151 40.2269 -70.8894 105 127
7 Review AR8 10/03/2016 06/17/2017 CGMCP-02PMCI-00007 CGINS-ADCPTG-20495 40.2267 -70.8782 104 126
8 Review AR18 06/16/2017 11/07/2017 CGMCP-02PMCI-00008 CGINS-ADCPTG-24543 40.227 -70.8889 105 127
9 Review AR24 11/07/2017 04/11/2018 CGMCP-02PMCI-00009 CGINS-ADCPTG-19151 40.2266 -70.8782 103 125
10 Review AR28 04/11/2018 11/05/2018 CGMCP-02PMCI-00010 CGINS-ADCPTG-24543 40.2265 -70.8889 105 127
11 Review AR31 11/05/2018 04/18/2019 CGMCP-02PMCI-00011 CGINS-ADCPTG-19151 40.2267 -70.8783 104 126
12 Review AR34 04/17/2019 10/07/2019 CGMCP-02PMCI-00012 CGINS-ADCPTG-24542 40.2267 -70.8886 104 126
13 Review AR39 10/07/2019 06/12/2020 CGMCP-02PMCI-00013 CGINS-ADCPTG-19151 40.227 -70.8779 104 126
14 Review AR44 06/12/2020 CGMCP-02PMCI-00014 CGINS-ADCPTG-24542 40.2267 -70.889 105 127
Metadata Start Date End Date Comment
CP02PMCI-RII01-02-ADCPTG010

The telemetered .nc files for deployments 2 and 5 don't have pressure as a coordinate. (recovered data are not available for these deployments). These files are also missing two key variables: percent_bad_beams and bin_depths

By Lori Garzio, on 8/20/19

CP02PMCI-RII01-02-ADCPTG010
Deployment: 2

Annotation IDs 392 and 393 should be updated. These annotations say the recovered data are missing, however according to the ingest csvs, the data aren't expected.

By Lori Garzio, on 8/20/19

New Note

Metadata Start Date End Date Comment
CP02PMCI-RII01-02-ADCPTG010
Method: telemetered
Stream: adcps_jln_stc_instrument
4/16/14, 7:28 PM

Telemetered inductive ADCP data is in pd12 format which does not include the depth-to-first-bin value which is needed for proper calculation of the data products. A nominal value is entered into the system so that telemetered data can be used when needed. For the best data the recovered data in pd0 format should be used.

Id: 1914 By: cdobson

CP02PMCI-RII01-02-ADCPTG010
10/11/14, 2:11 PM 12/14/14, 7:00 PM

Telemetry session associated with profile zero showed no communication with ADCP.

Id: 604 By: leila

CP02PMCI-RII01-02-ADCPTG010
Method: recovered_host
10/11/14, 2:11 PM 12/14/14, 7:00 PM

Missing.

Id: 392 By: leila
Flag: not_available Exclude: No

CP02PMCI-RII01-02-ADCPTG010
Method: recovered_inst
10/11/14, 2:11 PM 12/14/14, 7:00 PM

Missing.

Id: 393 By: leila
Flag: not_available Exclude: No

CP02PMCI
5/20/16, 8:18 PM 5/30/16, 8:05 AM

Recovered data are not available because the Wire-Following Profiler motor and PAR cables were swapped and the data are thus unreliable.

Id: 228 By: leila

CP02PMCI
10/2/16, 1:30 PM 10/2/16, 2:30 PM

Telemetered data are sparse because telemetry schedule was irregular during this deployment.

Id: 232 By: leila

CP02PMCI-RII01-02-ADCPTG010
10/3/16, 4:52 PM 6/17/17, 8:12 AM

Deployment 7: this instrument was misconfigured with a blanking distance of 0.12 m - smaller than both the intended 4.12 m and the manufacturer's recommended minimum. The resulting calculated distance to the first bin is 5 m rather than 9 m. All bin depths will be offset 4 m from their expected values.

Id: 2000 By: swhite
Flag: suspect Exclude: No

CP02PMCI-RII01-02-ADCPTG010
Method: telemetered
12/20/16, 9:30 PM 6/16/17, 7:00 PM

Instrument ceased transmitting data.

Id: 231 By: leila
Flag: not_operational Exclude: No