Method | Data Stream | Content | Type | |
---|---|---|---|---|
recovered_wfp | ctdpf_ckl_wfp_instrument_recovered | Data Products | Report M2M Stats | Science |
recovered_wfp | ctdpf_ckl_wfp_metadata_recovered | Metadata | Report M2M | Engineering |
telemetered | ctdpf_ckl_wfp_instrument | Data Products | Report M2M Stats | Science |
telemetered | ctdpf_ckl_wfp_metadata | Metadata | Report M2M | Engineering |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP04OSPM-WFP01-03-CTDPFK000 |
Comparisons with the shipboard CTD casts were done for D0001, R0001, D0002, R0002, D0003, R0003, D0004, R0004, D0005, R0005, D0006, D0007, R0007, D0008. The available science parameter values and patterns are comparable between the profiler and CTD cast (with the exception of R0005). Comparison to the shipboard bottle data is outside the scope of this project. By Lori Garzio, on 3/13/19 |
||
CP04OSPM-WFP01 |
9/26/16, 12:00 AM | 9/28/16, 1:25 PM | |
CP04OSPM-WFP01-03-CTDPFK000 Deployment: 4 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP04OSPM-WFP01 |
11/2/14, 10:12 AM | 4/28/15, 11:13 AM |
Wire Following Profiler is stuck around 55 dbar. Id: 1393 By: leila |
CP04OSPM-WFP01 |
8/5/15, 8:14 AM | 10/12/15, 5:05 AM |
Wire Following Profiler is stuck at the bottom of its range. Id: 1394 By: leila |
CP04OSPM-WFP01 |
6/2/16, 11:00 AM | 9/28/16, 9:25 AM |
Wire Following Profiler is stuck at the top of its range. Id: 613 By: leila |
CP04OSPM |
10/12/16, 2:00 AM | 6/1/17, 7:22 PM |
Failed
Id: 354
By: leila
|
CP04OSPM-WFP01 |
1/2/18, 7:00 PM | 4/9/18, 12:25 PM |
Deployment 8: Wire-Following Profiler stopped transmitting data. Issue identified as a software problem.
Id: 227
By: leila
|
CP04OSPM-WFP01 |
12/31/19, 7:00 PM | 6/8/20, 12:19 PM |
Deployment 12: Due to a clock rollover issue, all WFPs have ceased sampling on 1/1/2020. No WFP data are being collected after this date. A resolution is being researched by the vendor. Id: 1885 By: cdobson |