Method | Data Stream | Content | Type | |
---|---|---|---|---|
recovered_host | zplsc_c_instrument | Data Products | Report M2M Stats | Science |
recovered_inst | zplsc_echogram_data | Data Products | Report M2M Stats | Science |
telemetered | zplsc_c_instrument | Data Products | Report M2M Stats | Science |
Deployment | Cruise | Start Date | Stop Date | Mooring Asset | Node Asset | Sensor Asset | Latitude | Longitude | Deployment Depth | Water Depth |
---|---|---|---|---|---|---|---|---|---|---|
2 Review | AT-27 | 05/10/2015 | 10/23/2015 | CGMCP-04OSSM-00002 | CGINS-ZPLSCC-55077 | 39.9406 | -70.8785 | 441 | 442 | |
3 Review | AT-31 | 10/22/2015 | 05/13/2016 | CGMCP-04OSSM-00003 | CGINS-ZPLSCC-55101 | 39.9373 | -70.8867 | 453 | 454 | |
4 Review | AR-04 | 05/27/2016 | 10/12/2016 | CGMCP-04OSSM-00004 | CGINS-ZPLSCC-55080 | 39.94 | -70.8774 | 445 | 446 | |
5 Review | AR-08 | 10/12/2016 | 06/08/2017 | CGMCP-04OSSM-00005 | CGINS-ZPLSCC-55101 | 39.9371 | -70.887 | 449 | 450 | |
6 Review | AR-18 | 06/06/2017 | 11/02/2017 | CGMCP-04OSSM-00006 | CGINS-ZPLSCC-55080 | 39.9394 | -70.8768 | 442 | 443 | |
7 Review | AR-24 | 10/28/2017 | 03/29/2018 | CGMCP-04OSSM-00007 | CGINS-ZPLSCC-55103 | 39.9376 | -70.8863 | 451 | 452 | |
8 Review | AR-28 | 03/24/2018 | 10/30/2018 | CGMCP-04OSSM-00008 | CGINS-ZPLSCC-55074 | 39.9391 | -70.8764 | 444 | 445 | |
9 Review | AR-31 | 10/25/2018 | 04/08/2019 | CGMCP-04OSSM-00009 | CGINS-ZPLSCC-55103 | 39.9361 | -70.885 | 448 | 449 | |
10 Review | AR-34 | 04/05/2019 | 09/28/2019 | CGMCP-04OSSM-00010 | CGINS-ZPLSCC-55077 | 39.9399 | -70.8774 | 450 | 451 | |
11 Review | AR-39 | 09/29/2019 | CGMCP-04OSSM-00011 | CGINS-ZPLSCC-55102 | 39.9368 | -70.8865 | 446 | 447 |
No notes yet.
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP04OSSM-MFD37-07-ZPLSCC000 Stream: zplsc_c_instrument |
5/8/15, 9:04 PM |
ZPLSC data may contain acoustic interference from the ADCP instrument subject to real-time clock drift of either instrument. Id: 2006 By: swhite |
|
CP04OSSM Method: telemetered |
10/22/15, 3:38 PM | 5/13/16, 1:18 PM |
Fuel cell impacted available instrument power, so telemetered data are not available or reliable. Recovered data are available for instruments that could record internally. Id: 1362 By: leila |
CP04OSSM-MFD37-07-ZPLSCC000 |
5/27/16, 11:50 AM | 10/12/16, 1:12 PM |
Deployment 4: 125 kHz transducer failure; Post-recovery evaluation of the instrument revealed failures or defects affecting the 125 kHz transducer. The precise time of failure cannot be determined and it is unknown to what extent data are or may be affected.
Id: 2022
By: swhite
|
CP04OSSM-MFD37 |
10/16/16, 8:00 PM | 4/2/17, 9:00 PM |
No telemetered data are expected because of ground faults on the seafloor Multi-Function Node. Recovered data are available from instruments recording internally. Id: 356 By: leila |
CP04OSSM-MFD37 |
6/28/17, 5:00 AM | 10/29/17, 3:00 AM |
No telemetered data expected because the DSL line has failed. Id: 358 By: leila |
CP04OSSM-MFD37 Method: telemetered |
2/9/18, 1:02 PM | 3/29/18, 8:42 AM |
Deployment 7: Communication with the MFN instruments was disabled due to intermittent communications over the DSL line.
Id: 938
By: leila
|
CP04OSSM-MFD37-07-ZPLSCC000 |
3/24/18, 10:37 AM | 10/30/18, 10:30 AM |
Deployment 8: 125 kHz transducer failure; Post-recovery evaluation of the instrument revealed failures or defects affecting the 125 kHz transducer. The precise time of failure cannot be determined and it is unknown to what extent data are or may be affected.
Id: 2026
By: swhite
|
CP04OSSM-MFD37 Method: telemetered |
5/9/18, 11:00 PM | 10/30/18, 10:30 AM |
Instruments are presently disabled. Note: Instruments may continue to sample using internal batteries and storage cards. Id: 1518 By: leila |
CP04OSSM Method: telemetered |
1/30/19, 10:00 AM | 4/8/19, 7:39 AM |
CP04OSSM D00009 buoy broke free and went adrift on 1/30/2019, 1500UTC. All instruments on the MFN ceased telemetry on 1/30/2019 and instruments on the buoy/NSIF continued to sample through 2/1/2019. No more telemetered data expected for this deployment. Id: 1557 By: cdobson |
CP04OSSM-MFD37-07-ZPLSCC000 |
4/5/19, 10:55 AM | 9/28/19, 8:38 AM |
Deployment 10: 125 kHz transducer failure; Post-recovery evaluation of the instrument revealed failures or defects affecting the 125 kHz transducer. The precise time of failure cannot be determined and it is unknown to what extent data are or may be affected.
Id: 2031
By: swhite
|
CP04OSSM-MFD37 |
5/11/19, 8:00 PM |
Deployment 10: Telemetered data from all MFN instruments could be delayed up to 3 days due to a limitation in communications with the MFN. Id: 1629 By: cdobson |
|
CP04OSSM-MFD37 |
11/14/19, 7:00 PM |
Deployment 11: MFN offline due to high voltage short on 2019-11-15. No telemetered data expected from instruments on the MFN; instruments with external batteries will continue to log internally.
Id: 1820
By: cdobson
|