Bio-acoustic Sonar (Coastal)

Reference Designator
CP04OSSM-MFD37-07-ZPLSCC000
Start Depth
450
End Depth
450
Location
Multi-Function Node
Current Status
Will Not Review
Class
ZPLSC (Bio-acoustic Sonar (Coastal))
Series
ZPLSC-C
Science Discipline
Biological
Make
ASL Environmental Sciences
Model
AZFP
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/CP04OSSM/MFD37/07-ZPLSCC000/metadata
Method Data Stream Content Type
recovered_host zplsc_c_instrument Data Products Report M2M Stats Science
recovered_inst zplsc_echogram_data Data Products Report M2M Stats Science
telemetered zplsc_c_instrument Data Products Report M2M Stats Science
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
2 Review AT-27 05/10/2015 10/23/2015 CGMCP-04OSSM-00002 CGINS-ZPLSCC-55077 39.9406 -70.8785 441 442
3 Review AT-31 10/22/2015 05/13/2016 CGMCP-04OSSM-00003 CGINS-ZPLSCC-55101 39.9373 -70.8867 453 454
4 Review AR-04 05/27/2016 10/12/2016 CGMCP-04OSSM-00004 CGINS-ZPLSCC-55080 39.94 -70.8774 445 446
5 Review AR-08 10/12/2016 06/08/2017 CGMCP-04OSSM-00005 CGINS-ZPLSCC-55101 39.9371 -70.887 449 450
6 Review AR-18 06/06/2017 11/02/2017 CGMCP-04OSSM-00006 CGINS-ZPLSCC-55080 39.9394 -70.8768 442 443
7 Review AR-24 10/28/2017 03/29/2018 CGMCP-04OSSM-00007 CGINS-ZPLSCC-55103 39.9376 -70.8863 451 452
8 Review AR-28 03/24/2018 10/30/2018 CGMCP-04OSSM-00008 CGINS-ZPLSCC-55074 39.9391 -70.8764 444 445
9 Review AR-31 10/25/2018 04/08/2019 CGMCP-04OSSM-00009 CGINS-ZPLSCC-55103 39.9361 -70.885 448 449
10 Review AR-34 04/05/2019 09/28/2019 CGMCP-04OSSM-00010 CGINS-ZPLSCC-55077 39.9399 -70.8774 450 451
11 Review AR-39 09/29/2019 CGMCP-04OSSM-00011 CGINS-ZPLSCC-55102 39.9368 -70.8865 446 447

No notes yet.

New Note

Metadata Start Date End Date Comment
CP04OSSM-MFD37-07-ZPLSCC000
Stream: zplsc_c_instrument
5/8/15, 9:04 PM

ZPLSC data may contain acoustic interference from the ADCP instrument subject to real-time clock drift of either instrument.

Id: 2006 By: swhite

CP04OSSM
Method: telemetered
10/22/15, 3:38 PM 5/13/16, 1:18 PM

Fuel cell impacted available instrument power, so telemetered data are not available or reliable. Recovered data are available for instruments that could record internally.

Id: 1362 By: leila

CP04OSSM-MFD37-07-ZPLSCC000
5/27/16, 11:50 AM 10/12/16, 1:12 PM

Deployment 4: 125 kHz transducer failure; Post-recovery evaluation of the instrument revealed failures or defects affecting the 125 kHz transducer. The precise time of failure cannot be determined and it is unknown to what extent data are or may be affected.

Id: 2022 By: swhite
Flag: suspect Exclude: No

CP04OSSM-MFD37
10/16/16, 8:00 PM 4/2/17, 9:00 PM

No telemetered data are expected because of ground faults on the seafloor Multi-Function Node. Recovered data are available from instruments recording internally.

Id: 356 By: leila

CP04OSSM-MFD37
6/28/17, 5:00 AM 10/29/17, 3:00 AM

No telemetered data expected because the DSL line has failed.

Id: 358 By: leila

CP04OSSM-MFD37
Method: telemetered
2/9/18, 1:02 PM 3/29/18, 8:42 AM

Deployment 7: Communication with the MFN instruments was disabled due to intermittent communications over the DSL line.

Id: 938 By: leila
Flag: not_available Exclude: No

CP04OSSM-MFD37-07-ZPLSCC000
3/24/18, 10:37 AM 10/30/18, 10:30 AM

Deployment 8: 125 kHz transducer failure; Post-recovery evaluation of the instrument revealed failures or defects affecting the 125 kHz transducer. The precise time of failure cannot be determined and it is unknown to what extent data are or may be affected.

Id: 2026 By: swhite
Flag: suspect Exclude: No

CP04OSSM-MFD37
Method: telemetered
5/9/18, 11:00 PM 10/30/18, 10:30 AM

Instruments are presently disabled. Note: Instruments may continue to sample using internal batteries and storage cards.

Id: 1518 By: leila

CP04OSSM
Method: telemetered
1/30/19, 10:00 AM 4/8/19, 7:39 AM

CP04OSSM D00009 buoy broke free and went adrift on 1/30/2019, 1500UTC. All instruments on the MFN ceased telemetry on 1/30/2019 and instruments on the buoy/NSIF continued to sample through 2/1/2019. No more telemetered data expected for this deployment.

Id: 1557 By: cdobson

CP04OSSM-MFD37-07-ZPLSCC000
4/5/19, 10:55 AM 9/28/19, 8:38 AM

Deployment 10: 125 kHz transducer failure; Post-recovery evaluation of the instrument revealed failures or defects affecting the 125 kHz transducer. The precise time of failure cannot be determined and it is unknown to what extent data are or may be affected.

Id: 2031 By: swhite
Flag: suspect Exclude: No

CP04OSSM-MFD37
5/11/19, 8:00 PM

Deployment 10: Telemetered data from all MFN instruments could be delayed up to 3 days due to a limitation in communications with the MFN.

Id: 1629 By: cdobson

CP04OSSM-MFD37
11/14/19, 7:00 PM

Deployment 11: MFN offline due to high voltage short on 2019-11-15. No telemetered data expected from instruments on the MFN; instruments with external batteries will continue to log internally.

Id: 1820 By: cdobson
Flag: not_operational Exclude: No