CTD

Reference Designator
CP05MOAS-GL374-03-CTDGVM000
Start Depth
0
End Depth
1,000
Location
Current Status
Review Complete
Class
CTDGV (CTD Glider)
Series
CTDGV-M
Science Discipline
Physical
Make
Sea-Bird
Model
SBE Glider Payload CTD (GP-CTD)
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/CP05MOAS/GL374/03-CTDGVM000/metadata
Method Data Stream Content Type
recovered_host ctdgv_m_glider_instrument_recovered Data Products Report M2M Stats Science
telemetered ctdgv_m_glider_instrument Data Products Report M2M Stats Science
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review KN-217 04/16/2014 07/11/2014 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00374 CGINS-CTDGVM-09083 40.1115 -70.5555 200 200
2 Review AT-27 05/09/2015 07/24/2015 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00374 CGINS-CTDGVM-09083 39.9418 -70.8266 200 200
3 Review AR-04 05/29/2016 08/09/2016 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00374 CGINS-CTDGVM-09083 40.1195 -70.4483 200 200
4 Review AR-08 10/11/2016 12/20/2016 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00374 CGINS-CTDGVM-09361 40.0873 -70.3782 200 200
5 Review SI20170418 04/20/2017 04/20/2017 CGVCP-05MOAS-00000 CGVEH-GLDRCP-00374 CGINS-CTDGVM-09361 40.0288 -70.4857 200 200
Metadata Start Date End Date Comment
CP05MOAS

Bad altimeter pings are apparent in many deployments of the coastal gliders. A few example plots are attached - m_water_depth plotted as semi-transparent bathymetry.

By Lori Garzio, on 5/10/19

CP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/15/19

CP05MOAS-GL374-03-CTDGVM000

Comparisons with the shipboard CTD casts were done for D0001, D0002, D0003 and D0004. Salinity, conductivity, and temperature from the glider are higher for D0001 compared to the CTD cast. Comparison to the shipboard bottle data is outside the scope of this project.

By Lori Garzio, on 3/13/19

CP05MOAS-GL374-03-CTDGVM000
Deployment: 3

Suspect data towards the end of the deployment (e.g. high variability in salinity and density)

By Lori Garzio, on 9/3/19

CP05MOAS-GL374
Deployment: 4

Glider looks like it drifted at the end of the deployment - this should be annotated

By Lori Garzio, on 9/3/19

New Note

No annotations yet.